journal article·2020-10-28Open access

Combined Use of a Transmission Detector and an EPID-Based In Vivo Dose Monitoring System in External Beam Whole Breast Irradiation: A Study with an Anthropomorphic Female Phantom

Chiara Arilli, Yannik Wandael, Chiara Galeotti, Livia Marrazzo, Silvia Calusi, Mattia Grusio, Isacco Desideri, Franco Fusi, Angelo Piermattei, Stefania Pallotta, Cinzia Talamonti

Applied Sciences· Volume 10· Issue 21

Abstract

We evaluate the combined usage of two systems, the Integral Quality Monitor (IQM) transmission detector and SoftDiso software, for in vivo dose monitoring by simultaneous detection of delivery and patient setup errors in whole breast irradiation. An Alderson RANDO phantom was adapted with silicon breast prostheses to mimic the female anatomy. Plans with simulated delivery errors were created from a reference left breast plan, and patient setup errors were simulated by moving the phantom. Deviations from reference values recorded by both monitoring systems were measured for all plans and phantom positions. A 2D global gamma analysis was performed in SoftDiso for all phantom displacements. Both IQM signals and SoftDiso R-values are sensitive to small MU variations. However, only IQM is sensitive to jaw position variations. Conversely, IQM is unable to detect patient positioning errors, and the R-value has good sensitivity to phantom displacements. A gamma comparison analysis allows one to determine alert thresholds to detect phantom shifts or relatively large rotations. The combined use of the IQM and SoftDiso allows for fast identification of both delivery and setup errors and substantially reduces the impact of error identification and correction on the treatment workflow.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Advanced Radiotherapy Techniques · Breast Cancer Treatment Studies · Digital Radiography and Breast Imaging

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

AOU Careggi, Medical Physics Unit, 50134 Florence, Italy

AOU Careggi, Radiotherapy Unit, 50134 Florence, Italy

Department of Experimental and Clinical Biomedical Sciences “Mario Serio”, University of Florence, 50134 Florence, Italy

UOC Fisica Sanitaria, Fondazione Policlinico Universitario Agostino Gemelli, 00168 Rome, Italy

References (50)

Show source-provided references
1. Twenty-Five-Year Follow-up of a Randomized Trial Comparing Radical Mastectomy, Total Mastectomy, and Total Mastectomy Followed by Irradiation10.1056/NEJMoa020128 ↗
2. Twenty-Year Follow-up of a Randomized Trial Comparing Total Mastectomy, Lumpectomy, and Lumpectomy plus Irradiation for the Treatment of Invasive Breast Cancer10.1056/NEJMoa022152 ↗
3. Twenty-Year Follow-up of a Randomized Study Comparing Breast-Conserving Surgery with Radical Mastectomy for Early Breast Cancer10.1056/NEJMoa020989 ↗
4. A planning comparison of dose patterns in organs at risk and predicted risk for radiation induced malignancy in the contralateral breast following radiation therapy of primary breast using conventional, IMRT and Volumetric modulated arc treatment techniques10.1080/02841860802657227 ↗
5. Comparison of conventional inserts and an add-on electron MLC for chest wall irradiation of left-sided breast cancer10.1080/02841860802477907 ↗
6. Clinical experience in breast irradiation with intensity modulated photon beams10.1080/02841860510029879 ↗
7. Dosimetric comparison of intensity modulated radiotherapy techniques and standard wedged tangents for whole breast radiotherapy10.1111/j.1754-9485.2009.02043.x ↗
8. Dosimetric comparison of left-sided whole-breast irradiation with 3DCRT, forward-planned IMRT, inverse-planned IMRT, helical tomotherapy, and volumetric arc therapy10.1016/j.ejmp.2015.02.005 ↗
9. Catching errors with in vivo EPID dosimetry10.1118/1.3397807 ↗
10. Quality Control Quantification (QCQ): A Tool to Measure the Value of Quality Control Checks in Radiation Oncology10.1016/j.ijrobp.2012.04.036 ↗
11. Using a Novel Dose QA Tool to Quantify the Impact of Systematic Errors Otherwise Undetected by Conventional QA Methods: Clinical Head and Neck Case Studies10.7785/tcrt.2012.500353 ↗
12. Evaluating IMRT and VMAT dose accuracy: Practical examples of failure to detect systematic errors when applying a commonly used metric and action levels10.1118/1.4826166 ↗
13. Sensitivity of volumetric modulated arc therapy patient specific QA results to multileaf collimator errors and correlation to dose volume histogram based metrics10.1118/1.4824433 ↗
14. In vivo dosimetry during external photon beam radiotherapy10.1016/S0360-3016(98)00341-1 ↗
15. Quality assurance by systematic in vivo dosimetry: Results on a large cohort of patients10.1016/S0167-8140(00)00195-X ↗
16. Practical Guidelines for the Implementation of In Vivo Dosimetry with Diodes in External Radiotherapy with Photon Beams (Entrance Dose)
17. AAPM Task Group 62 of the Radiation Therapy Committee (2005). Diode in vivo dosimetry for patients receiving external beam radiation therapy. AAPM Rep., 1, 1–84.
18. World Health Organization (2020, September 18). Radiotherapy Risk Profile 2008. WHO/IER/PSP/2008.12. Available online: https://www.who.int/patientsafety/activities/technical/radiotherapy_risk_profile.pdf?ua=1.
19. (2013). Development of Procedures for In Vivo Dosimetry in Radiotherapy, International Atomic Energy Agency. Number 8 in Human Health Reports.
20. The Glasgow Incident—A Physicist’s Reflections10.1016/j.clon.2006.12.003 ↗
21. Radiotherapy Near Misses, Incidents and Errors: Radiotherapy Incident at Glasgow10.1016/j.clon.2006.12.004 ↗
22. ICRP publication 112. A report of preventing accidental exposures from new external beam radiation therapy technologies10.1016/j.icrp.2009.12.001 ↗
23. Lessons from recent accidents in radiation therapy in France10.1093/rpd/ncn235 ↗
24. (2014). Council Directive 2013/59/EURATOM of 5 December 2013. OJ L, 13, 1–73. Available online: https://eur-lex.europa.eu/LexUriServ/LexUriServ.do?uri=OJ:L:2014:013:0001:0073:EN:PDF.
25. (2014). Radiation Protection and Safety of Radiation Sources: International Basic Safety Standards, International Atomic Energy Agency. Number GSR Part 3 in General Safety Requirements.
26. The Royal College of Radiologists, Society and College of Radiographers, Institute of Physics and Engineering in Medicine, British Institute of Radiology (2008). Implementing In Vivo Dosimetry, The Royal College of Radiologists.
27. (2000). The Swedish Radiation Protection Institute’s Regulations on Radiation Therapy, Swedish Radiation Protection Institute. Number SSI-FS–2000-4.
28. (2008). Critères D’Agrément Pour la Pratique de la Radiothérapie Externe, Institut National du Cancer (INCa).
29. (2014). Patient Safety. Paving the Way for Progress. In Vivo Dosimetry. Newsletters for Radiotherapy Professionals-n.5 2014. The French Nuclear Safety Authority.
30. In vivo dosimetry in external beam radiotherapy10.1118/1.4811216 ↗
31. A literature review of electronic portal imaging for radiotherapy dosimetry10.1016/j.radonc.2008.07.008 ↗
32. In vivo dosimetry by an aSi-based EPID10.1118/1.2360014 ↗
33. Application of a practical method for the isocenter point in vivo dosimetry by a transit signal10.1088/0031-9155/52/16/026 ↗
34. Held, M., Cheung, J., Andujar, A.P., and Morn, O. (2008). Commissioning and Evaluation of an Electronic Portal Imaging Device-Based In-Vivo Dosimetry Software. Cureus, 10.
35. A simple backprojection algorithm for 3D in vivo EPID dosimetry of IMRT treatments10.1118/1.3148482 ↗
36. 3D Dosimetric verification of volumetric-modulated arc therapy by portal dosimetry10.1016/j.radonc.2009.12.020 ↗
37. DAVID—A translucent multi-wire transmission ionization chamber forin vivoverification of IMRT and conformal irradiation techniques10.1088/0031-9155/51/5/013 ↗
38. Clinical performance of a transmission detector array for the permanent supervision of IMRT deliveries10.1016/j.radonc.2009.12.041 ↗
39. The influence of a novel transmission detector on 6 MV x-ray beam characteristics10.1088/0031-9155/54/10/014 ↗
40. An integral quality monitoring system for real-time verification of intensity modulated radiation therapy10.1118/1.3250859 ↗
41. A method for online verification of adapted fields using an independent dose monitor10.1118/1.4811204 ↗
42. Characterization of a novel two dimensional diode array the “magic plate” as a radiation detector for radiation therapy treatment10.1118/1.3700234 ↗
43. Real-time beam monitoring for error detection in IMRT plans and impact on dose-volume histograms10.1007/s00066-017-1245-3 ↗
44. Influence of the Integral Quality Monitor transmission detector on high energy photon beams: A multi-centre study10.1016/j.zemedi.2016.10.001 ↗
45. Sensitivity of the IQM transmission detector to errors of VMAT plans10.1002/mp.13228 ↗
46. Error detection capability of a novel transmission detector: A validation study for online VMAT monitoring10.1088/1361-6560/aa7dc7 ↗
47. A validation study of a dedicated software for an automated in vivo dosimetry control in radiotherapy10.1007/s11517-018-1822-3 ↗
48. Falco, M., Giancaterino, S., De Nicola, A., Adorante, N., De Lorenzo, R.G., Di Tommaso, M., Vinciguerra, A., Trignani, M., Perrotti, F., and Allajbej, A. (2018). A Feasibility Study for in vivo Dosimetry Procedure in Routine Clinical Practice. Technol. Cancer Res. Treat., 17.10.1177/1533033818779201 ↗
49. Epid-based in vivo dose verification for lung stereotactic treatments delivered with multiple breath-hold segmented volumetric modulated arc therapy10.1002/acm2.12538 ↗
50. Evaluation of interfraction setup variations for postmastectomy radiation therapy using EPID-based in vivo dosimetry10.1002/acm2.12712 ↗