Development of an Evaluation System for Magnetic Resonance Imaging Based Three-Dimensional Modeling of a Transfemoral Prosthetic Socket Using Finite Elements
Source abstract
Abstract
Recent technologies have suggested the utilization of three-dimensional (3D) printing technology to enhance the fabrication accuracy of prosthetics. Accordingly, simulations are used to obtain precise parameters for subject-specified prosthetic socket. This study proposes an evaluation system to measure the accuracy of a subject-specific 3D transfemoral residuum model during the interaction with the socket in conjunction with the application of finite element methods. The proposed system can be used in future validations of socket fabrication. The evaluation is based on the measurement of the residuum’s soft tissue deformation inside two types of prosthetic sockets. In comparison with other studies, the 3D models were constructed with magnetic resonance images (MRI) with the aid of computer-aided design (CAD) software. The measurement of soft tissue deformation was conducted based on the measurement of the volumetric value of fat, muscle and skin in the pre- and post-donning phases. The result yielded a promising correlation coefficient value between the simulation and the experiment in the soft tissue deformation evaluation. The relation of the muscle–fat ratio in the residuum is extremely important in the determination of the ability of the prosthetic to deform. The environment during the socket fitting session was similar to that defined by the set boundary conditions in simulations. In view of the promising results of this study, the evaluation system proposed herein is considered reliable and is envisaged to be used in future research.
Source: crossref · Quality: baseline
Research context
OpenAlex topics: Prosthetics and Rehabilitation Robotics · Muscle activation and electromyography studies · Orthopaedic implants and arthroplasty
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
Affiliations
Department of Bio-Science and Engineering, Shibaura Institute of Technology, Tokyo 135-8548, Japan
Department of Supporting Prosthetic Orthotics, Niigata University of Health and Welfare, Niigata 950-3102, Japan
Department of Prosthetic and Orthotics, University of Human Arts and Science, Saitama 339-8539, Japan
Department of Science and Engineering, Tokyo Denki University, Tokyo 120-8551, Japan
Funding
JSPS KAKENHI
References (20)
Show source-provided references
1. Sugimoto, A. (2013). Charactheristic of Stump Shape and Stress during Gait Cycle for Transfemoral Prosthesis Socket. [Master’s Thesis, Shibaura Institute of Technology].
2. Zhang, L., Zhu, M., Shen, L., and Zheng, F. (2013, January 3–7). Finite element analysis of the contact interface between trans-femoral stump and prosthetic socket. Proceedings of the 2013 35th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC), Osaka, Japan.
3. Zuniga, J.M. (2018). 3D printed antibacterial Prostheses. Appl. Sci., 8.10.3390/app8091651 ↗
4. Nguyen, K.-T., Benabou, L., and Alfayad, S. (2018, January 7–11). Systematic Review of Prosthetic Socket Fabrication using 3D printing. Proceedings of the 2018 4th International Conference on Mechatronics and Robotics Engineering—ICMRE 2018, Valenciennes, France.10.1145/3191477.3191506 ↗
5. Finite element analysis of donning procedure of a prosthetic transfemoral socket
6. The finite element implementation of 3D fractional viscoelastic constitutive models10.1016/j.finel.2018.04.003 ↗
7. Direct medical image-based Finite Element modelling for patient-specific simulation of future implants10.1016/j.finel.2017.07.010 ↗
8. Multi-material 3-D viscoelastic model of a transtibial residuum from in-vivo indentation and MRI data10.1016/j.jmbbm.2016.02.020 ↗
9. Finite element modeling of the contact interface between trans-tibial residual limb and prosthetic socket10.1016/j.medengphy.2004.04.010 ↗
10. Effects of sitting postures on risks for deep tissue injury in the residuum of a transtibial prosthetic-user: A biomechanical case study10.1080/10255842.2010.504719 ↗
11. Comparison of image processing techniques (magnetic resonance imaging, computed tomography scan and ultrasound) for 3D modeling and analysis of the human bones10.1007/BF03168701 ↗
12. Udai, A.D., and Sinha, A.N. (2008, January 8–10). Processing magnetic resonance images for CAD model development of prosthetic limbs socket. Proceedings of the 2008 IEEE Region 10 and the Third International Conference on Industrial and Information Systems, Kharagpur, India.10.1109/ICIINFS.2008.4798343 ↗
13. Jamaludin, M.S., Hanafusa, A., Yamamoto, S.-I., Agarie, Y., Otsuka, H., and Onishi, K. (2018, January 3–6). Evaluation of the effect of the geometrical changes in prosthetic socket towards transfemoral residuum via finite element method. Proceedings of the IEEE–EMBS Conference on Biomedical Engineering and Science (IECBES), Sarawak, Malaysia.10.1109/IECBES.2018.8626703 ↗
14. Accuracy evaluation of 3D reconstruction of transfemoral residual limb model using basic spline interpolation
15. Real-time patient-specific finite element analysis of internal stresses in the soft tissues of a residual limb: A new tool for prosthetic fitting10.1007/s10439-006-9208-3 ↗
16. Finite element analysis for quantitative evaluation of a trans-femoral prosthesis socket for standing posture
17. Untaroiu, C., Darvish, K., Crandall, J.R., Deng, B., and Wang, J.-R. (2005, January 6–9). Characterization of the lower limb soft tissues in pedestrian finite element models. Proceedings of the 19th International technical conference on the Enhanced Safety of Vehicles, Washington, DC, USA.
18. Regional differences in pain threshold and tolerance of the transtibial residual limb: Including the effects of age and interface material10.1016/j.apmr.2004.08.005 ↗
19. Residual Limb Volume Change: Systematic Review of Measurement and Management10.1682/JRRD.2010.09.0189 ↗
20. Shape and volume change in the transtibial residuum over the short term: Preliminary investigation of six subjects10.1682/JRRD.2003.10.0153 ↗