journal article·2021-10-15Open access

Design and Development of Magnesium-Based Suture Anchor for Rotator Cuff Repair Using Finite Element Analysis and In Vitro Testing

Ting-Yu Su, Hao-Yuan Tang, Jason Shian-Ching Jang, Chih-Hwa Chen, Hsiang-Ho Chen

Applied Sciences· Volume 11· Issue 20

Abstract

A suture anchor is a medical device commonly used in rotator cuff repair surgery to attach tendons to the greater tuberosity of the humerus. Patient- and device-related factors, such as structural designs and poor bone density, can cause unsatisfactory clinical results. In this study, a new suture anchor design with four distinctive parameters was proposed, and the structural performance was optimized in a full factorial experimental design using finite element analysis. Two types of bone blocks—normal and osteoporotic bone—which received screw implants, were simulated to investigate the parametrical effects on various bone qualities. The prescribed motion at a constant removal velocity was used to evaluate the pullout strength. The von Mises criterion was employed in a force control simulation for topology optimization. Moreover, mechanical tests guided by ASTM-F543-17 were conducted for validation. This paper demonstrates the comprehensive process for developing a suture anchor with sufficient mechanical integrity for clinical use and clarifies the contributions of each distinctive design parameter in this application.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Shoulder Injury and Treatment · Shoulder and Clavicle Injuries · Nerve Injury and Rehabilitation

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

School of Biomedical Engineering, College of Biomedical Engineering, Taipei Medical University, Taipei 11031, Taiwan

Department of Mechanical Engineering, Institute of Materials Science and Engineering, National Central University, Taoyuan 32001, Taiwan

Department of Orthopedics, Shuang Ho Hospital, Taipei Medical University, Taipei 23561, Taiwan

Bone and Joint Research Center, Department of Orthopedics, Taipei Medical University Hospital, School of Medicine, College of Medicine, Taipei Medical University, Taipei 11031, Taiwan

Center for Biomedical Engineering, Graduate Institute of Biomedical Engineering, College of Engineering, Chang Gung University, Taoyuan 33302, Taiwan

Bone and Joint Research Center, Department of Orthopedic Surgery, Chang Gung Memorial Hospital Linkou Branch, Taoyuan 333, Taiwan

Funding

Ministry of Science and Technology, Taiwan

References (39)

Show source-provided references
1. Tendon-to-Bone Pressure Distributions at a Repaired Rotator Cuff Footprint Using Transosseous Suture and Suture Anchor Fixation Techniques10.1177/0363546504273053 ↗
2. Rotator Cuff Repair: An Ex Vivo Analysis of Suture Anchor Repair Techniques on Initial Load to Failure10.1016/j.arthro.2005.06.022 ↗
3. The Mechanical Effects of Suture Anchor Insertion Angle for Rotator Cuff Repair10.3928/0147-7447-20020401-14 ↗
4. Operative treatment of chronic ruptures of the rotator cuff of the shoulder10.1302/0301-620X.65B2.6826624 ↗
5. Arthroscopic Repair of Full-Thickness Tears of the Rotator Cuff10.2106/00004623-199806000-00007 ↗
6. Mode of failure for rotator cuff repair with suture anchors identified at revision surgery10.1067/mse.2003.21 ↗
7. The effect of suture anchor design and orientation on suture abrasion: An in vitro study10.1053/jars.2003.50032 ↗
8. The Outcome and Repair Integrity of Completely Arthroscopically Repaired Large and Massive Rotator Cuff Tears10.2106/00004623-200402000-00002 ↗
9. Functional and Anatomical Results After Rotator Cuff Repair10.1097/00003086-199407000-00009 ↗
10. Repairs of the rotator cuff. Correlation of functional results with integrity of the cuff10.2106/00004623-199173070-00004 ↗
11. Effect of structural design on the pullout strength of suture anchors for rotator cuff repair10.1002/jor.24135 ↗
12. Stress distribution inside bone after suture anchor insertion: Simulation using a three-dimensional finite element method10.1007/s00167-012-2060-0 ↗
13. Interference screw versus Endoscrew fixation for anterior cruciate ligament reconstruction: A biomechanical comparative study in sawbones and porcine knees
14. A Comparison of Screw Insertion Torque and Pullout Strength10.1097/BOT.0b013e3181c4a655 ↗
15. Trabecular deformations during screw pull-out: A micro-CT study of lapine bone10.1007/s10237-017-0891-9 ↗
16. Use of robotic technology to study the biomechanics of ligaments and their replacements10.1016/S1048-6666(00)80048-8 ↗
17. Contact Area, Contact Pressure, and Pressure Patterns of the Tendon-Bone Interface after Rotator Cuff Repair10.1177/0363546505278256 ↗
18. Does Change in Thread Shape Influence the Pull Out Strength of Mini Implants? An In vitro Study
19. Insertion torque in different bone models with different screw pitch: An in vitro study
20. Suture Anchor Materials, Eyelets, and Designs: Update 200810.1016/j.arthro.2008.03.006 ↗
21. The history of suture anchors in orthopaedic surgery10.1016/j.clinbiomech.2018.11.008 ↗
22. A Comparison of Open-Construct PEEK Suture Anchor and Non-Vented Biocomposite Suture Anchor in Arthroscopic Rotator Cuff Repair: A Prospective Randomized Clinical Trial10.1016/j.arthro.2019.08.049 ↗
23. In vivo comparison of a metal versus a biodegradable suture anchor10.1016/j.arthro.2004.03.008 ↗
24. Wang, J., Xu, J., Hopkins, C., Chow, D.H., and Qin, L. (2020). Biodegradable Magnesium-Based Implants in Orthopedics—A General Review and Perspectives. Adv. Sci., 7.10.1002/advs.201902443 ↗
25. Effects of magnesium coating on bone-implant interfaces with and without polyether-ether-ketone particle interference: A rabbit model based on porous Ti6Al4V implants10.1002/jbm.b.34332 ↗
26. Inflammatory response to magnesium-based biodegradable implant materials10.1016/j.actbio.2019.10.014 ↗
27. Er, M.S., Altinel, L., Eroglu, M., Verim, O., Demir, T., and Atmaca, H. (2014). Suture anchor fixation strength with or without augmentation in osteopenic and severely osteoporotic bones in rotator cuff repair: A biomechanical study on polyurethane foam model. J. Orthop. Surg. Res., 9.10.1186/1749-799X-9-48 ↗
28. Hibbeler, R.C. (2003). Mechanics of Materials, Pearson Education.
29. Liu, S., Qi, W., Zhang, Y., Wu, Z.-X., Yan, Y.-B., and Lei, W. (2014). Effect of bone material properties on effective region in screw-bone model: An experimental and finite element study. Biomed. Eng. Online, 13.10.1186/1475-925X-13-83 ↗
30. ASTM International (2017). ASTM F 543. Standard Specification and Test Methods for Metallic Medical Bone Screws, ASTM International.
31. Degradation behavior and mechanical strength of Mg-Zn-Ca bulk metallic glass composites with Ti particles as biodegradable materials10.1016/j.jallcom.2017.01.010 ↗
32. Mechanical properties of cancellous bone of the distal humerus10.1016/j.clinbiomech.2005.05.014 ↗
33. Finite Element Study On Arthroscopic Anchor Design Aspects10.3889/oamjms.2019.164 ↗
34. Optimized thread pitch design and stress analysis of the cylinder screwed dental implant
35. Kim, D.-G., Kim, K.-H., Jo, Y., Lee, J.Y., Park, Y.J., Chung, C.P., Seol, Y.-J., and Han, J.-S. (2020). Bone regeneration into side openings and hollow inner channel of a dental implant. J. Mech. Behav. Biomed. Mater., 101.10.1016/j.jmbbm.2019.103416 ↗
36. Finite element analysis of four thread-form configurations in a stepped screw implant10.1046/j.0305-182X.2003.01213.x ↗
37. Assessment of magnesium-based biomaterials: From bench to clinic10.1039/C9BM00289H ↗
38. Prasadh, S., Ratheesh, V., Manakari, V., Parande, G., Gupta, M., and Wong, R. (2019). The Potential of Magnesium Based Materials in Mandibular Reconstruction. Metals, 9.10.3390/met9030302 ↗
39. Osteocyte function under compressive mechanical force10.1016/j.jdsr.2013.10.004 ↗