journal article·2019-10-29Open access

Physiological Control Law for Rotary Blood Pumps with Full-State Feedback Method

Mohsen Bakouri

Applied Sciences· Volume 9· Issue 21

Abstract

One concern about pulsatile rotary blood pumps is their physiological controller reactions when “venous return” changes. When a patient rises from a supine to a standing position, the blood volume in the leg veins is raised, owing to vasodilation, thus venous returns to the right atrium, and consequently, the left atrium is reduced. In this work, a physiological control law using a full-state feedback control method was developed in order to drive mechanical circulatory support. This strategy was used as a validated state-space pump model, to implement the controller and regulate the desired reference flow. The control law was assessed using a software model of the hemodynamical cardiovascular system interacting with the left ventricular assist device in different physiological conditions ranging from rest to exercise scenarios. Under these scenarios, heart failure disease was simulated by changing the hemodynamic parameters of the total blood volume, heart rate, cardiac contractility, and systemic peripheral resistance. The results were numerically observed during the postural changes. The rate of change in the physiological variables showed that the proposed control law can regulate the desired reference pump flow with minimal error within the acceptable clinical range in order to prevent suction and over perfusion.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Mechanical Circulatory Support Devices · Cardiac Arrest and Resuscitation · Cardiac Structural Anomalies and Repair

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

Department of Medical Equipment Technology, College of Applied Medical Science, Majmaah University, Majmaah City 11952, Saudi Arabia

Department of Physics, College of Arts, Sebha University, Traghen City 71340, Libya

Funding

Majmaah University

References (36)

Show source-provided references
1. The heart failure epidemic: A UK perspective10.1530/ERP-16-0043 ↗
2. Incidence and aetiology of heart failure; a population-based study10.1053/euhj.1998.1280 ↗
3. Coronary artery disease as the cause of incident heart failure in the population10.1053/euhj.2000.2289 ↗
4. The evolution of mechanical circulatory support (MCS): A new wave of developments in MCS and heart failure treatment10.21037/jtd.2018.05.205 ↗
5. Mechanical cardiac assistance: Historical perspectives10.1053/stcs.2000.18455 ↗
6. Assessment of Aortic Valve Opening During Rotary Blood Pump Support Using Pump Signals10.1111/aor.12167 ↗
7. Left ventricular assist device therapy in patients with restrictive and hypertrophic cardiomyopathy10.1161/CIRCHEARTFAILURE.110.959288 ↗
8. Developments in control systems for rotary left ventricular assist devices for heart failure patients: A review10.1088/0967-3334/34/1/R1 ↗
9. Physiologic outcome of varying speed rotary blood pump support algorithms: A review study10.1007/s13246-015-0405-y ↗
10. Numerical Optimization Studies of Cardiovascular-Rotary Blood Pump Interaction10.1111/j.1525-1594.2012.01449.x ↗
11. A Sliding Mode-Based Starling-Like Controller for Implantable Rotary Blood Pumps10.1111/aor.12223 ↗
12. Development of a Reliable Automatic Speed Control System for Rotary Blood Pumps10.1016/j.healun.2005.02.004 ↗
13. Hierarchical control of heart-assist devices10.1109/MRA.2003.1191711 ↗
14. ACC/AHA/ESC 2006 guidelines for the management of patients with atrial fibrillation–executive summary. A report of the American College of Cardiology/American Heart Association Task Force on practice guidelines and the European Society of Cardiology Committee for Practice Guidelines (Writing Committee to Revise the 2001 Guidelines for the Management of Patients with Atrial Fibrillation)
15. Patients who Have Dilated Cardiomyopathy Must Have a Trial of Bridge to Recovery (Pro)10.1016/j.hfc.2007.05.006 ↗
16. Control Strategy for Maintaining Physiological Perfusion with Rotary Blood Pumps10.1046/j.1525-1594.2003.07089.x ↗
17. Modeling, Estimation, and Control of Human Circulatory System with a Left Ventricular Assist Device10.1109/TCST.2006.890288 ↗
18. Response of Rotary Blood Pumps to Changes in Preload and Afterload at a Fixed Speed Setting Are Unphysiological When Compared with the Natural Heart10.1111/j.1525-1594.2010.01168.x ↗
19. Controller for an Axial Flow Blood Pump10.1111/j.1525-1594.1996.tb04491.x ↗
20. Detection of Total Assist and Sucking Points Based on Pulsatility of a Continuous Flow Artificial Heart: In Vitro Evaluation10.1097/00002480-199809000-00082 ↗
21. Sensorless Controlling Method for a Continuous Flow Left Ventricular Assist Device10.1046/j.1525-1594.2000.06585.x ↗
22. Nonlinear modelling and control of left ventricular assist device10.1049/el.2014.4330 ↗
23. Evaluation of conventional nonpulsatile and novel pulsatile extracorporeal life support systems in a simulated pediatric extracorporeal life support model
24. Rotary pump speed modulation for generating pulsatile flow and phasic left ventricular volume unloading in a bovine model of chronic ischemic heart failure10.1016/j.healun.2014.09.017 ↗
25. Hemodynamic controller for left ventricular assist device based on pulsatility ratio10.1111/j.1525-1594.2007.00350.x ↗
26. Ng, B.C., Kleinheyer, M., Smith, P.A., Timms, D., Cohn, W.E., and Lim, E. (2018). Pulsatile operation of a continuous-flow right ventricular assist device (RVAD) to improve vascular pulsatility. PLoS ONE, 13.10.1371/journal.pone.0195975 ↗
27. Theoretical foundations of a Starling-like controller for rotary blood pumps10.1111/j.1525-1594.2012.01457.x ↗
28. Franklin, G.F., Powell, J.D., and Emami-Naeini, A. (2014). Feedback Control of Dynamic Systems, Prentice Hall Press.
29. Evaluation of an advanced model reference sliding mode control method for cardiac assist device using a numerical model10.1049/iet-syb.2017.0052 ↗
30. Bakouri, M. (2019, January 25–28). Physiological Control Approach for Heart Pump. Proceedings of the 2019 18th European Control Conference (ECC), Naples, Italy.10.23919/ECC.2019.8795740 ↗
31. Mechanical circulatory support in the new era: An overview10.1186/s13054-016-1235-3 ↗
32. Sensor-based physiologic control strategy for biventricular support with rotary blood pumps10.1097/MAT.0000000000000671 ↗
33. Evaluation of physiological control systems for rotary left ventricular assist devices: An in-vitro study10.1007/s10439-016-1552-3 ↗
34. Development of a Suction Detection System for Axial Blood Pumps10.1111/j.1525-1594.2004.00011.x ↗
35. Bakouri, M.A., Savkin, A.V., and Alomari, A.H. (June, January 31). A method for physiological control of a cardiac assist device. Proceedings of the 2015 10th Asian Control Conference (ASCC), Kota Kinabalu, Malaysia.10.1109/ASCC.2015.7244808 ↗
36. The Index of Motor Current Amplitude Has Feasibility in Control for Continuous Flow Pumps and Evaluation of Left Ventricular Function10.1046/j.1525-1594.2001.06859.x ↗