journal article·2025-02-07Open access

Preliminary Investigation of a Cd0.9Zn0.1Te Detector for Small-Field Dosimetry Applications Using Therapeutic MV Beams

Sangsu Kim, Ju-Young Song, Yong-Hyub Kim, Jae-Uk Jeong, Mee Sun Yoon, Taek-Keun Nam, Sung-Ja Ahn, Shinhaeng Cho

Applied Sciences· Volume 15· Issue 4

Abstract

Stereotactic radiosurgery (SRS) and stereotactic body radiation therapy (SBRT) require precise small-field dosimetry, verified through patient-specific quality assurance (PSQA). This study evaluated the feasibility of using a single-crystal cadmium–zinc–telluride (Cd0.9Zn0.1Te, CZT) detector for PSQA in SRS and SBRT. We fabricated a CZT detector with Au electrodes and examined its fundamental characteristics, including dose linearity, dose rate dependence, energy dependence, angular dependence, source-to-surface distance (SSD) dependence, field size dependence, depth dependence, and reproducibility, under 6 and 10 MV LINAC beam irradiation and compared the results with those from a standard ionization chamber. The results revealed that the CZT detector demonstrated excellent linearity across 0–1000 cGy with minimal deviation in the low-dose region, negligible dose rate dependence, and minimal energy dependence, exhibiting a 2.2% drop at 15 MV relative to 6 MV. Its angular and SSD dependencies deviated slightly from the ionization chamber, consistent with the expected physical behaviors and correctable in clinical practice. The detector also revealed consistent performance over time with excellent reproducibility, and its depth dependence results were consistent with those of the ionization chamber. Thus, the CZT detector provides consistent performance in small-field measurements under varying conditions, satisfying the requirements for SRS and SBRT.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Advanced Radiotherapy Techniques · Advanced Semiconductor Detectors and Materials · Advanced X-ray and CT Imaging

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

Global Health Technology Research Center, Korea University, Seoul 02841, Republic of Korea

Department of Radiation Oncology, Chonnam National University Medical School, Gwangju 61469, Republic of Korea

Funding

Chonnam National University Hospital Biomedical Research Institute

References (39)

Show source-provided references
1. Stereotactic body radiation therapy: A novel treatment modality10.1038/nrclinonc.2009.188 ↗
2. Radiosurgery for brain metastases: Changing practice patterns and disparities in the United States10.6004/jnccn.2017.7003 ↗
3. SBRT dose and survival in non-small cell lung cancer: In regard to Koshy et al10.1016/j.ijrobp.2015.03.032 ↗
4. Lung stereotactic body radiation therapy (SBRT) dose gradient and PTV volume: A retrospective multicenter analysis10.1186/s13014-019-1334-9 ↗
5. A dose falloff gradient study in RapidArc planning of lung stereotactic body radiation therapy10.4103/jmp.JMP_38_18 ↗
6. Characterisation of a synthetic diamond detector for end-to-end dosimetry in stereotactic body radiotherapy and radiosurgery10.1016/j.phro.2021.10.002 ↗
7. AAPM-RSS medical physics practice Guideline 9.a. for SRS-SBRT10.1002/acm2.12146 ↗
8. Stereotactic body radiation therapy: The report of AAPM Task Group 10110.1118/1.3438081 ↗
9. Agency, I.A.E. (2017). Dosimetry of Small Static Fields Used in External Beam Radiotherapy: An International Code of Practice for Reference and Relative Dose Determination, International Atomic Energy Agency (IAEA).
10. Guidelines for safe practice of stereotactic body (ablative) radiation therapy10.1111/1754-9485.12336 ↗
11. The Canadian association of radiation oncology scope of practice guidelines for lung, liver and spine stereotactic body radiotherapy10.1016/j.clon.2012.04.006 ↗
12. Enhancing safety and quality through preplanning peer review for patients undergoing stereotactic body radiation therapy10.1016/j.prro.2015.09.009 ↗
13. A literature review of electronic portal imaging for radiotherapy dosimetry10.1016/j.radonc.2008.07.008 ↗
14. Doses on the central axes of narrow 6-MV x-ray beams10.1118/1.596475 ↗
15. Radiochromic film dosimetry: Recommendations of AAPM radiation therapy committee task group 55. American Association of Physicists in Medicine10.1118/1.598407 ↗
16. Scanning orientation effects on Gagchromic EBT film dosimetry10.1007/BF03178579 ↗
17. Precautions and strategies in using a commercial flatbed scanner for radiochromic film dosimetry10.1088/0031-9155/52/1/015 ↗
18. Characterization of small-field stereotactic radiosurgery beams with modern detectors10.1088/0031-9155/58/21/7595 ↗
19. The origin of the flatbed scanner artifacts in radiochromic film dosimetry—Key experiments and theoretical descriptions10.1088/0031-9155/61/21/7704 ↗
20. Darafsheh, A., Zhao, T., and Khan, R. (2020). Spectroscopic analysis of irradiated radiochromic EBT-XD films in proton and photon beams. Phys. Med. Biol., 65.10.1088/1361-6560/aba28e ↗
21. Experimental analysis of general ion recombination in a liquid-filled ionization chamber in high energy photon beams10.1118/1.4805109 ↗
22. Characterization of a two-dimensional liquid-filled ion chamber detector array used for verification of the treatments in radiotherapy10.1118/1.4870439 ↗
23. Detector comparison for small field output factor measurements in flattening filter free photon beams10.1016/j.radonc.2013.10.022 ↗
24. Energy and field size dependence of a silicon diode designed for small-field dosimetry10.1002/mp.12195 ↗
25. Comparative study of SRS end-to-end QA processes of a diode array device and an anthropomorphic phantom loaded with GafChromic XD film10.1002/acm2.13747 ↗
26. 2D EPID dose calibration for pretreatment quality control of conformal and IMRT fields: A simple and fast convolution approach10.1016/j.ejmp.2015.10.094 ↗
27. On flattening filter-free portal dosimetry10.1120/jacmp.v17i4.6147 ↗
28. Time dependent pre-treatment EPID dosimetry for standard and FFF VMAT10.1088/0031-9155/59/16/4749 ↗
29. Investigating the electronic portal imaging device for small radiation field measurements10.4103/jmp.JMP_131_16 ↗
30. Arlt, R., Brutscher, J., Gunnink, R., Ivanov, V., Parnham, K., Soldner, S., and Stein, J. (2000, January 15–20). Use of CdZnTe detectors in hand-held and portable isotope identifiers to detect illicit trafficking of nuclear material and radioactive sources. Proceedings of the 2000 IEEE Nuclear Science Symposium Conference Record (cat. No. 00CH37149), Lyon, France.
31. Progress in the development of CdTe and CdZnTe semiconductor radiation detectors for astrophysical and medical applications10.3390/s90503491 ↗
32. Investigation of room temperature nuclear radiation CdZnTe pixel array detector
33. Haijun, X. (2013). Data Acquisition System of CZT Nuclear Detector Based on FPGA. [Master’s Thesis, Chongqing University].
34. Growth of CdTexSe1-x from a Te-rich solution for applications in radiation detection10.1016/j.jcrysgro.2013.09.039 ↗
35. An ASIC for the study of charge sharing effects in small pixel CdZnTe X-ray detectors10.1109/TNS.2011.2162746 ↗
36. Intensity modulated radiation therapy plans verification using a Gaussian convolution kernel to correct the single chamber response Function of the I’mRT MatriXX Array10.3923/jas.2015.483.491 ↗
37. Simulation of energy spectrum characteristics of planar CdZnTe detector
38. A CdZnTe coplanar-grid detector array for environmental remediation10.1016/S0168-9002(00)00876-7 ↗
39. High efficiency pixellated CdTe detector10.1016/S0168-9002(97)00238-6 ↗