Retrievability of Fractured Abutment Screws and Damage to Implant–Abutment Connections: An In Vitro Comparative Analysis of Conventional vs. Drilling Techniques
Source abstract
Abstract
This study aimed to evaluate the retrievability and potential damage to implant–abutment connections caused by fractured abutment screw removal using conventional and drilling techniques. A total of forty abutment screws were randomly inserted into forty dental implants, and then they were fractured and extracted using different removal methods: Group A employed a conventional approach utilizing an exploration probe and an ultrasonic device without irrigation (n = 10) (conventional); Group B used the Phibo drilling removal system without irrigation (n = 10) (Phibo); Group C utilized the Rhein83® drilling removal system without irrigation (n = 10) (Rhein83); and Group D implemented the Sanhigia® drilling removal system without irrigation (n = 10) (Sanhigia). Pre- and postoperative micro-computed tomography (micro-CT) scans were performed on the dental implants, and Standard Tessellation Language (STL) digital files were generated for morphometric analysis to measure the wear volume. ANOVA was used to assess the volumetric differences (mm3) and percentage ratios of the internal thread volumes of the implant–abutment connections before and after the procedures. Results: This study found no statistically significant differences in the volumetric and percentage ratios of internal threads among the implant groups (Phibo, Rhein83, Sanhigia, and conventional). However, the success rate for retrieving fractured abutment screws was higher (90%) with the drilling systems compared to the conventional technique (50%). These results suggest that drilling systems are more effective for the retrieval of damaged screws. Although drilling techniques without irrigation demonstrated higher removal efficiency compared to the conventional method, both approaches resulted in similar wear volumes at the implant–abutment connections when used to extract fractured screws.
Source: crossref · Quality: baseline
Research context
OpenAlex topics: Drilling and Well Engineering · Hydraulic Fracturing and Reservoir Analysis · Orthopaedic implants and arthroplasty
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
Affiliations
Faculty of Dentistry, Alfonso X El Sabio University, 28691 Madrid, Spain
Department of Surgery, Faculty of Medicine and Dentistry, University of Salamanca, 37008 Salamanca, Spain
Department of Stomatology, Faculty of Medicine and Dentistry, Universiat of València, 46010 Valencia, Spain
References (32)
Show source-provided references
1. A risk-based decision-making tree for managing fractured abutment and prosthetic screws: A systematic review10.1016/j.prosdent.2017.05.016 ↗
2. Stability of Implant-Abutment Connection After Using the Rescue Kit: An In Vitro Study10.11607/jomi.6386 ↗
3. Evaluation of torque maintenance of abutment and cylinder screws with Morse taper implants10.1097/SCS.0b013e318270fe0e ↗
4. A systematic review of the 5-year survival and complication rates of implant-supported single crowns10.1111/j.1600-0501.2007.01453.x ↗
5. Heat generation during removal of an abutment screw fragment from dental implants10.1016/j.prosdent.2017.06.011 ↗
6. Retrieval of Fractured Implant Abutment Screws: A Narrative Review10.4103/jispcd.JISPCD_318_21 ↗
7. In vitro comparison of the efficacy of two fractured implant-prosthesis screw extraction methods: Conventional versus mechanical10.1016/j.prosdent.2019.10.014 ↗
8. Pérez, M.B., Zubizarreta-Macho, Á., Sánchez, J.B., Rodríguez, J.H., Pérez-Barquero, J.A., Deglow, E.R., and Montero, S.H. (2022). Removal capability, implant-abutment connection damage and thermal effect using ultrasonic and drilling techniques for the extraction of fractured abutment screws: An in vitro study. BMC Oral Health, 22.10.1186/s12903-022-02653-w ↗
9. A conservative approach to retrieve a fractured abutment screw—Case report10.1016/j.jpor.2015.09.003 ↗
10. In Vitro Evaluation of Three Methods for the Retrieval of Fractured Screw Fragments from Dental Implants10.11607/jomi.5174 ↗
11. (2016). Dentistry. Implants. Dynamic Loading Test for Endosseous Dental Implants (Standard No. ISO 14801:2016).
12. Retrospective analysis of prosthetic complications of implant-supported fixed partial dentures after an observation period of 5 to 10 years10.11607/jomi.2999 ↗
13. A systematic review of biologic and technical complications with fixed implant rehabilitations for edentulous patients
14. Load fatigue performance of a single-tooth implant abutment system: Effect of diameter
15. The influence of implant-abutment connection on the screw loosening and microleakage10.1186/s40729-018-0121-y ↗
16. Screw loosening and changes in removal torque relative to abutment screw length in a dental implant with external abutment connection after oblique cyclic loading10.4047/jap.2018.10.6.415 ↗
17. Mechanical complications of dental implants10.1034/j.1600-0501.2000.011S1156.x ↗
18. Relationship between implant preload and screw loosening on implant-supported prostheses10.1563/1548-1336(2002)028<0067:TRBIPA>2.3.CO;2 ↗
19. Influence of implant/abutment joint designs on abutment screw loosening in a dental implant system10.1002/jbm.b.30328 ↗
20. Mechanism of and factors associated with the loosening of the implant abutment screw: A review10.1111/jerd.12494 ↗
21. Removal of a fractured implant abutment screw: A clinical report10.1016/j.prosdent.2004.03.010 ↗
22. Rapid retrieval of fractured implant abutment screws with a custom-made screwdriver10.1016/j.jds.2023.05.028 ↗
23. The cotton driver: An alternative technique for removing fractured screw fragments10.1016/j.prosdent.2014.07.012 ↗
24. Various Treatment Modalities for Retrieval of Broken Abutment Screw in Implant Prosthodontics10.5005/jp-journals-10004-1035 ↗
25. Safe Removal of a Broken Abutment Screw with Customized Drill Guide and Rotary Instrument: A Clinical Report10.1111/jopr.12291 ↗
26. Management of a Fractured Implant Abutment Screw10.1563/aaid-joi-D-16-00105 ↗
27. Heat generation in 1-piece implants during abutment preparations with high-speed cutting instruments10.1097/ID.0b013e318277af53 ↗
28. Heat generation during implant site preparation: An in vitro study
29. Assessment of bone viability after heat trauma. A histological, histochemical and vital microscopic study in the rabbit
30. Management of a fractured implant abutment screw: A clinical report10.1111/jopr.12111 ↗
31. The effect of screw length on fracture load and abutment strain in dental implants with external abutment connections
32. Alternative approach to salvaging an implant with a fractured screw fragment: A clinical report10.1016/j.prosdent.2019.12.015 ↗