journal article·2024-04-22Open access

Comparison of the Operator and Surrounding Dose When Using Portable Intraoral X-ray Devices

Mehrdad Abdinian, Maedeh Aminian, Forouzan Keymasi, Parisa Soltani, Mariangela Cernera, Niccolo Giuseppe Armogida, Gianrico Spagnuolo

Applied Sciences· Volume 14· Issue 8

Abstract

This study aimed to investigate the scattered radiation dose using three portable dental radiographic units: iRay D3, EZRay Air, and Epix. The absorbed dose was measured at 0.5 and 1 m distances, every 15° in the horizontal plane, using an ionization chamber. The maximum number of radiographs per day using the portable units was calculated considering a dose limit of 50 mSv/year and 20 mSv/year. The doses were higher in the Epix unit compared to the other two devices. Anterior exposure was generally higher than the sides or posterior exposure. With a dose limit of 50 mSv/year, considering a distance of 0.5 m between the operator and the X-ray unit, a maximum of 961, 565, and 38 radiographs are permitted daily using iRay D3, EZRay Air, and Epix, respectively. Considering a dose limit of 20 mSv/year, with a distance of 0.5 m between the operator and the radiographic device, a maximum of 384, 226, and 15 radiographs are permitted daily using iRay D3, EZRay Air, and Epix portable units, respectively. It is highly unlikely that an operator would reach occupational dose limits when using iRay D3 and EZRay Air. The Epix radiographic device allows for fewer daily radiographs and should be avoided for daily use.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Radiation Dose and Imaging · Dental Radiography and Imaging · Advanced X-ray and CT Imaging

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

Department of Oral and Maxillofacial Radiology, Dental Implants Research Center, Dental Research Institute, School of Dentistry, Isfahan University of Medical Sciences, Isfahan 8174673461, Iran

Department of Oral and Maxillofacial Radiology, School of Dentistry, Ardabil University of Medical Sciences, Ardabil 5618953141, Iran

Student Research Committee, School of Dentistry, Isfahan University of Medical Sciences, Isfahan 8174673461, Iran

Department of Neurosciences, Reproductive and Odontostomatological Sciences, University of Naples “Federico II”, 80138 Naples, Italy

Therapeutic Dentistry Department, Institute for Dentistry, Sechenov University, 119991 Moscow, Russia

Funding

Isfahan University of Medical Sciences

References (25)

Show source-provided references
1. White, S.C., and Pharoah, M.J. (2018). White and Pharoah’s Oral Radiology E-Book: Principles and Interpretation, Elsevier Health Sciences.
2. Portable dental equipment: Dental units and X-ray equipment
3. Hand-held dental X-ray (HDX) with medical collimator: Use in casualty radiology10.1093/milmed/161.7.428 ↗
4. Use of the hand-held dental X-ray machine during joint operation, NATO exercise Display Determination-9210.1093/milmed/162.8.575 ↗
5. Information yield from a hand-held dental X-ray unit10.1016/0030-4220(93)90272-6 ↗
6. In vitro evaluation of the XR-15 portable X-ray unit for forensic odontology
7. Maintaining radiation exposures as low as reasonably achievable (ALARA) for dental personnel operating portable hand-held X-ray equipment10.1097/HP.0b013e318259fa29 ↗
8. The effects of device position on the operator’s radiation dose when using a handheld portable X-ray device10.1259/dmfr.20150245 ↗
9. Handheld intraoral dental X-ray devices should supplement but not replace conventional radiographic equipment10.1016/j.oooo.2021.06.001 ↗
10. Assessment of the radiological safety of a Genoray portable dental X-ray unit10.1259/dmfr.20140255 ↗
11. Operator Exposure to Scatter Radiation from a Portable Hand-held Dental Radiation Emitting Device (Aribex™ NOMAD™) While Making 915 Intraoral Dental Radiographs10.1111/j.1556-4029.2008.00960.x ↗
12. Evaluation of radiation exposure to operators of portable hand-held dental X-ray units10.1007/s13246-021-00985-5 ↗
13. Determination of scattered radiation around PORT-XII portable dental radiography unit
14. ICRP (2007). The 2007 Recommendations of the International Commission on Radiological Protection, ICRP.
15. The study of protection of operators and surrounding workers at the time of using portable intraoral X-ray unit10.1016/j.legalmed.2018.05.007 ↗
16. On the safe use of portable intraoral X-ray units in large-scale disasters10.15214/jsodom.30.311 ↗
17. Altındağ, A., Eren, H., Orhan, K., and Görgün, S. (2023). Evaluation of Operator and Patient Doses after Irradiation with Handheld X-ray Devices. Appl. Sci., 13.10.3390/app131810414 ↗
18. An assessment of the usefulness of handheld X-ray devices in general radiography based on a performance evaluation experiment10.61186/ijrr.21.3.545 ↗
19. Diagnostic image quality of hand-held and wall-mounted X-ray devices in bitewing radiography: A non-inferiority clinical trial10.1259/dmfr.20200471 ↗
20. Image Quality Assessment of Digital Radiographs Captured by Hand-Held Devices versus Wall-Mounted Devices: A Retrospective Comparative Study
21. Radiation protection recommendations on dose limits: The role on the NCRP and the ICRP and future developments10.1016/0360-3016(94)00275-P ↗
22. A comprehensive review of dose limits, triage systems and measurement tools for consequence management of nuclear and radiological emergencies10.1016/j.radphyschem.2024.111533 ↗
23. The scientific basis for the use of the linear no-threshold (LNT) model at low doses and dose rates in radiological protection10.1088/1361-6498/acdfd7 ↗
24. Radiation safety characteristics of the NOMAD™ portable X-ray system
25. Comparison of air dose and operator exposure from portable X-ray units10.1016/j.legalmed.2020.101787 ↗