journal article·2024-02-14Open access

Setup Optimization in Ocular Proton Therapy at the National Centre for Oncological Hadrontherapy: Comparison of Two Approaches to Refine the Position of an Eye-Tracking Device

Giulia Sellaro, Andrea Pella, Matteo Pepa, Federica Galante, Mario Ciocca, Maria Rosaria Fiore, Agnieszka Chalaszczyk, Chiara Paganelli, Marco Rotondi, Alessandro Vai, Ester Orlandi, Guido Baroni

Applied Sciences· Volume 14· Issue 4

Abstract

This study describes a method for setup optimization in patient simulation for ocular proton therapy (OPT) at the National Center for Oncological Hadrontherapy (CNAO) in Pavia, Italy, with the aim of minimizing the occupancy time of clinical areas and streamlining the actual procedure. Setup repeatability is ensured by patient-specific immobilization tools and relies on the patient’s ability to maintain a stable gaze direction according to the treatment plan. This is facilitated by aligning a light source (LED) on a patient-specific base along the prescribed gaze direction. At CNAO, a dedicated Eye-Tracking System (ETS) was designed to provide the patient with a visible source of light aligned to the desired gaze direction. The ETS position is defined prior to treatment planning, relying on optical-tracking guidance and comparing the position of passive markers fixed on the ETS chassis with patient-specific models prepared offline in accordance with the desired geometry. OPT at CNAO started in 2016 and may be considered as a consolidated clinical routine. However, all the preparation phases, including patient-specific ETS models and setup, still require long sessions in clinical areas such as the computed tomography (CT) and the treatment rooms, with a non-negligible impact on other activities. This study describes a novel approach for patient-specific definition of the ETS position and orientation, aiming at minimizing the time required for preparatory activities inside clinical areas. To minimize the occurrence of biases and to reproduce as much as possible a real end-to-end approach, we included in the analysis data of patients that received OPT in our facility. The study was performed in parallel, carrying out the alignment with the standard method currently used in the clinical workflow of CNAO and with the proposed method. Results are presented as 3D residuals and gaze deviations, comparing ETS alignment based on the new approach with respect to the clinical standard method. The preliminary results of this study are evidence of the capability of the procedure to align the ETS position, allowing performing of the procedure in a non-clinical dedicated room.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Ocular Oncology and Treatments · Photodynamic Therapy Research Studies · Retinal Diseases and Treatments

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

Bioengineering Unit, Clinical Department, National Centre for Oncological Hadrontherapy (CNAO), 27100 Pavia, Italy

Medical Physics Unit, Clinical Department, National Centre for Oncological Hadrontherapy (CNAO), 27100 Pavia, Italy

Radiotherapy Unit, Clinical Department, National Centre for Oncological Hadrontherapy (CNAO), 27100 Pavia, Italy

Department of Electronics, Information and Bioengineering, Politecnico di Milano (POLIMI), 20133 Milan, Italy

References (27)

Show source-provided references
1. Practice considerations for proton beam radiation therapy of uveal melanoma during the coronavirus disease pandemic: Particle therapy co-operative group ocular experience10.1016/j.adro.2020.04.010 ↗
2. Proton therapy for the management of uveal melanoma and other ocular tumors10.21037/cco.2016.07.06 ↗
3. Radiation therapy for choroidal melanoma10.1016/S0039-6257(97)00088-X ↗
4. Practice Patterns Analysis of Ocular Proton Therapy Centers: The International OPTIC Survey
5. Results of proton therapy of uveal melanomas treated in Nice
6. Advances in the clinical management of uveal melanoma10.1038/s41571-022-00714-1 ↗
7. Cui, Y., Yin, S., Qin, X., Jiao, W., Ren, A., Wang, F., and Zhao, B. (2022). Advances in the treatment of intraocular malignancies: A literature review. Front. Med., 9.10.3389/fmed.2022.975565 ↗
8. Proton beam therapy for iris melanomas in 107 patients10.1016/j.ophtha.2017.10.009 ↗
9. Clinical Outcomes of Proton Radiotherapy for Uveal Melanoma10.1016/j.clon.2016.01.034 ↗
10. Long-term risk of melanoma-related mortality for patients with uveal melanoma treated with proton beam therapy10.1001/jamaophthalmol.2015.0887 ↗
11. Outcomes After Proton Beam Therapy for Large Choroidal Melanomas in 492 Patients
12. Radiotherapy for ocular tumours10.1038/eye.2012.241 ↗
13. Proton beam radiotherapy10.1259/bjr.73.872.11026853 ↗
14. Diagnostic methods and therapeutic options of uveal melanoma with emphasis on MR imaging—Part I: MR imaging with pathologic correlation and technical considerations10.1186/s13244-021-01000-x ↗
15. Lane, S.A., Slater, J.M., and Yang, G.Y. (2023). Image-Guided Proton Therapy: A Comprehensive Review. Cancers, 15.10.3390/cancers15092555 ↗
16. Automated Treatment Planning System for Uveal Melanomas Treated with Proton Therapy: A Proof-of-Concept Analysis
17. Noninvasive eye localization in ocular proton therapy through optical eye tracking: A proof of concept10.1002/mp.12841 ↗
18. Development and validation of a new set-up simulator dedicated to ocular proton therapy at CNAO10.1016/j.ejmp.2021.01.070 ↗
19. The status of CNAO10.1140/epjp/i2011-11078-8 ↗
20. Optical eye tracking system for noninvasive and automatic monitoring of eye position and movements in radiotherapy treatments of ocular tumors10.1364/AO.51.002441 ↗
21. Optical eye tracking system for real-time noninvasive tumor localization in external beam radiotherapy10.1118/1.4915921 ↗
22. Development of a Camera Localization System for Three-Dimensional Digital Image Correlation Camera Triangulation10.1109/JSEN.2020.2997774 ↗
23. Ghost marker detection and elimination in marker-based optical tracking systems for real-time tracking in stereotactic body radiotherapy10.1118/1.4896126 ↗
24. A comparison between 2D plate calibration and wand calibration for 3D kinematic systems
25. A multi-camera calibration method using a 3-axis frame and wand10.1007/s12541-012-0035-1 ↗
26. Complete calibration of a stereo photogrammetric system through control points of unknown coordinates10.1016/S0021-9290(98)00104-3 ↗
27. Kacperek, A. (2012). Ion Beam Therapy, Springer.