Assessment of MicroPET Image Quality Based on Reconstruction Methods and Post-Filtering

Hyeon-Sik Kim, Byeong-il Lee, Jae-Sung Ahn

Applied Sciences· Volume 11· Issue 18

Abstract

The accuracy of positron emission tomography (PET) imaging is hampered by the partial volume effect (PVE), which causes image blurring and sampling. The PVE produces spillover phenomena, making PET analysis difficult. Generally, the PVE values vary based on reconstruction methods and filtering. Thus, selection of the proper reconstruction and filtering method can ensure accurate and high-quality PET images. This study compared the values of factors (recovery coefficient (RC), uniformity, and spillover ratio (SOR)) associated with different reconstruction and post-filtering methods using a mouse image quality phantom (NEMA NU 4), and we present an effective approach for microPET images. The PET images were obtained using a microPET scanner (Inveon, Siemens Medical Solutions, Malvern, PA, USA). PET data were reconstructed and/or post-filtered. For tumors smaller than 3 mm, iterative reconstruction methods provided better image quality. For tumor sizes bigger than 3 mm, reconstruction methods without post-filtering showed better results.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Medical Imaging Techniques and Applications · Radiomics and Machine Learning in Medical Imaging · Radiation Detection and Scintillator Technologies

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

Medical Photonics Research Center, Korea Photonics Technology Institute, Gwangju 61007, Korea

Department of Smart Healthcare, Pukyoung National University, Busan 48513, Korea

Funding

Basic Science Research Program of the National Research Foundation (NRF) of Korea

References (17)

Show source-provided references
1. Di Martino, F., Barca, P., Bortoli, E., Giuliano, A., and Volterrani, D. (2021). Correction for the Partial Volume Effects (PVE) in Nuclear Medicine Imaging: A Post-Reconstruction Analytic Method. Appl. Sci., 11.10.3390/app11146460 ↗
2. Validation of iterative multi-resolution method for partial volume correction and quantification improvement in PET image10.1016/j.bspc.2020.101954 ↗
3. Partial volume correction analysis for 11C-UCB-J PET studies of Alzheimer’s disease10.1016/j.neuroimage.2021.118248 ↗
4. Comparison of Correction Techniques for the Spillin Effect in Emission Tomography10.1109/TRPMS.2020.2980443 ↗
5. Pitfalls in quantitative myocardial PET perfusion I: Myocardial partial volume correction10.1007/s12350-020-02073-9 ↗
6. Performance evaluation of the inveon dedicated PET preclinical tomograph based on the NEMA NU-4 standards10.2967/jnumed.108.056374 ↗
7. Image-quality assessment for several positron emitters using the NEMA NU 4-2008 standards in the Siemens Inveon small-animal PET scanner10.2967/jnumed.109.068858 ↗
8. NEMA NU-04-based performance characteristics of the LabPET-8 small animal PET scanner10.1088/0031-9155/56/20/009 ↗
9. NEMA NU 4-2008 comparison of preclinical PET imaging systems10.2967/jnumed.111.099382 ↗
10. NEMA NU-4 performance evaluation of PETbox4, a high sensitivity dedicated PET preclinical tomograph10.1088/0031-9155/58/11/3791 ↗
11. Liu, Q., Li, C., Liu, J., Krish, K., Fu, X., Zhao, J., and Chen, J.C. (2021). Performance evaluation of a small-animal PET/CT system based on NEMA NU 4–2008 standards. Med. Phys.10.1002/mp.15088 ↗
12. Fighting Together against the Pandemic: Learning Multiple Models on Tomography Images for COVID-19 Diagnosis10.3390/ai2020016 ↗
13. An improved patch-based regularization method for PET image reconstruction10.21037/qims-20-19 ↗
14. Comparison of reconstruction methods and quantitative accuracy in Siemens Inveon PET scanner10.1088/1748-0221/10/04/P04001 ↗
15. Quantitative comparison of analytic and iterative reconstruction methods in 2- and 3-dimensional dynamic cardiac 18F-FDG PET
16. Evaluation of iterative reconstruction (OSEM) versus filtered back-projection for the assessment of myocardial glucose uptake and myocardial perfusion using dynamic PET10.1007/s00259-006-0198-z ↗
17. Comparison of Reconstruction Methods in a Small Animal Cardiac Positron Emission Tomography Study Using a 18F-Labeled Myocardial Agent, [18F] FPTP