The Effect of Undersized Drilling on the Coronal Surface Roughness of Microthreaded Implants: An In Vitro Study

Omer Cohen, Ofer Moses, Talia Gurevich, Roni Kolerman, Alina Becker, Zeev Ormianer

Applied Sciences· Volume 10· Issue 15

Abstract

This in-vitro study assessed the effect of an underdrilling implant placement protocol on the insertion torque, implant surface temperature and surface roughness (Sa) topography of the cervical microthreads of implants. Three groups of 25 implants (3.75 mm × 10 mm) were placed in osteotomies prepared in an artificial bone disc with final diameters of 3.65 mm according to the manufacturer’s instructions and in osteotomies prepared in accordance with an underdrilling protocol with final drill diameters of 3.2 and 2.8 mm (groups D3.65, D3.2, D2.8, respectively). Implants were inserted at a constant rate of 30 rpm. The surface temperature of the implants was measured with a thermal camera and temperature amplitude (Temp-Amp) was calculated by subtracting the room temperature from the measured implant surface temperature. Upon implant retrieval, coronal surface topography was assessed using a Nanofocus µsurf explorer and compared to a set of 25 new implants (control group). The differences between groups were compared using one-way ANOVA (p < 0.05). Significantly higher insertion torque, surface temperature values and significantly smaller average Sa values were measured in the implants inserted in undersized preparations. The highest temperature, insertion torque and Temp-Amp values and the largest decrease in Sa were measured in the D2.8 group. The lowest values were measured in the D3.65 group.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Dental Implant Techniques and Outcomes · Bone Tissue Engineering Materials · Dental materials and restorations

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

Department of Periodontology and Dental Implantology, School of Dental Medicine, Tel-Aviv University, P.O. Box 39040, Tel Aviv 6997801, Israel

Department of Oral Rehabilitation, School of Dental Medicine, Tel-Aviv University, P.O. Box 39040, Tel Aviv 6997801, Israel

References (31)

Show source-provided references
1. Implant stability in relation to osseointegration: An experimental study in the Labrador dog10.1111/j.1600-0501.2008.01646.x ↗
2. Optimal implant stabilization in low density bone10.1034/j.1600-0501.2001.120501.x ↗
3. The effect of thread pattern upon implant osseointegration10.1111/j.1600-0501.2009.01800.x ↗
4. Dental implant thread pitch and its influence on the osseointegration process: An in vivo comparison study
5. Misch, C.E. (2008). Scientific rationale for dental implant design. Contemporary Implant Dentistry, Mosby.
6. The effects of implant surface roughness and surgical technique on implant fixation in an in vitro model10.1111/j.1600-0501.2005.01202.x ↗
7. Effect of drilling dimension on implant placement torque and early osseointegration stages: An experimental study in dogs10.1016/j.joms.2011.08.006 ↗
8. Histological evaluation of oral implants inserted with different surgical techniques into the trabecular bone of goats10.1111/j.1600-0501.2007.01362.x ↗
9. Influence of the surgical technique and surface roughness on the primary stability of an implant in artificial bone with a density equivalent to maxillary bone: A laboratory study10.1111/j.1600-0501.2008.01692.x ↗
10. Differences in crestal bone-to-implant contact following an under-drilling compared to an over-drilling protocol. A study in the rabbit tibia10.1007/s00784-016-1765-8 ↗
11. Scarano, A., Carinci, F., Lorusso, F., Festa, F., Bevilacqua, L., Santos de Oliveira, P., and Maglione, M. (2018). Ultrasonic vs Drill Implant Site Preparation: Post-Operative Pain Measurement Through VAS, Swelling and Crestal Bone Remodeling: A Randomized Clinical Study. Materials, 11.10.3390/ma11122516 ↗
12. Herrero-Climent, M., Ferreira Lemos, B., Herrero-Climent, F., Falcao, C., Oliveira, H., Herrera, M., Javier Gil, F., Ríos-Carrasco, B., and Ríos-Santos, J.V. (2020). Influence of Implant Design and Under-Preparation of the Implant Site on Implant Primary Stability. An In Vitro Study. Int. J. Environ. Res. Public Health, 17.10.3390/ijerph17124436 ↗
13. Interaction of force-fitting and surface roughness of implants10.1111/j.1708-8208.2000.tb00120.x ↗
14. A rough surface implant neck with microthreads reduces the amount of marginal bone loss: A prospective clinical study10.1111/j.1600-0501.2009.01730.x ↗
15. Influence of implant length and diameter on stress distribution: A finite element analysis10.1016/j.prosdent.2003.08.008 ↗
16. The implant neck: Smooth or provided with retention elements. A biomechanical approach10.1034/j.1600-0501.1999.100506.x ↗
17. Tissue characteristics at microthreaded implants: An experimental study in dogs10.1111/j.1708-8208.2006.00016.x ↗
18. Effect of microthreads and platform switching on crestal bone stress levels: A finite element analysis10.1902/jop.2008.080178 ↗
19. Magnitude and direction of mechanical stress at the osseointegrated interface of the microthread implant10.1902/jop.2010.100237 ↗
20. Primary stability, insertion torque and bone density of cylindric implant ad modum Branemark: Is there a relationship? An in vitro study10.1111/j.1600-0501.2010.02036.x ↗
21. Evaluation of the surface damage of dental implants caused by different surgical protocols: An in vitro study10.1016/j.ijom.2018.12.009 ↗
22. A comparison of endosseous dental implant surfaces10.1902/jop.1999.70.12.1523 ↗
23. The effect of microthread design on magnitude and distribution of stresses in bone: A three-dimensional finite element analysis10.4103/1735-3327.240475 ↗
24. The effect of implant placement torque on crestal bone remodeling after 1 year of loading10.2147/CCIDE.S174895 ↗
25. Clinical outcome of dental implants placed with high insertion torques (up to 176Ncm)10.1111/j.1708-8208.2011.00351.x ↗
26. Peak insertion torque correlated to histologically and clinically evaluated bone density10.1097/ID.0b013e31821662b9 ↗
27. The mechanical response of commercially available bone simulants for quasi-static and dynamic loading10.1016/j.jmbbm.2018.10.032 ↗
28. Heat shock-induced necrosis and apoptosis in osteoblasts10.1002/jor.1100170614 ↗
29. Temperature threshold levels for heat-induced bone tissue injury: A vital-microscopic study in the rabbit10.1016/0022-3913(83)90174-9 ↗
30. Heat generated by dental implant drills during osteotomy—A review: Heat generated by dental implant drills10.1007/s13191-014-0350-6 ↗
31. Stocchero, M., Jinno, Y., Toia, M., Ahmad, M., Papia, E., Yamaguchi, S., and Becktor, J.P. (2019). Intraosseous temperature change during installation of dental implants with two different surfaces and different drilling protocols: An in vivo study in sheep. J. Clin. Med., 8.10.3390/jcm8081198 ↗