A Phase Error Correction System for Bioimpedance Measurement Circuits
Source abstract
Abstract
Bioimpedance sensing is widely used across a spectrum of biomedical applications. Among the different system architectures for measuring tissue impedance, synchronous detection or demodulation (SD) stands out for its lock-in amplifier utilising in-phase (I) and quadrature (Q) demodulation signals to derive real and imaginary impedance components. Typically, the current injected into the tissue is controlled by a voltage-controlled current source (VCCS). However, the VCCS can introduce phase shifts leading to discrepancies in real/imaginary outputs, especially at the highest end of the operating frequency bandwidth. Such discrepancies can significantly impact diagnostic accuracy in applications reliant on precise tissue phase profiling, such as cancer and neuromuscular evaluations. In the present work, we propose an automatic phase error compensation stage for bioimpedance measurement systems to minimise this systematic error. Our experimental findings demonstrated a considerable reduction in phase error, with the Phase Error Compensated Synchronous Detection (PECSD) system exhibiting a maximum phase error of 2° (≤5% error) compared with the uncompensated SD system where error exceeded 20%. The improvements made by our proposed SD system hold great potential for enhancing the accuracy of impedance measurements, particularly in clinical diagnosis and disease detection.
Source: crossref · Quality: baseline
Research context
OpenAlex topics: Electrical and Bioimpedance Tomography · Body Composition Measurement Techniques · Hemodynamic Monitoring and Therapy
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
Affiliations
Research Centre for Biomedical Engineering, City, University of London, London EC1V 0HB, UK
References (35)
Show source-provided references
1. Sunny, A.I., Rahman, M., Koutsoupidou, M., Cano-Garcia, H., Thanou, M., Rafique, W., Lipscombe, O., Kassanos, P., Triantis, I., and Kallos, E. (2019, January 23–27). Feasibility Experiments to Detect Skin Hydration Using a BioImpedance Sensor. Proceedings of the 2019 41st Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC), Berlin, Germany.10.1109/EMBC.2019.8857470 ↗
2. Skin cancer identification using multifrequency electrical impedance—A potential screening tool10.1109/TBME.2004.836523 ↗
3. Electrical impedance tomography: Tissue properties to image measures10.1109/TBME.2017.2728323 ↗
4. Bioimpedance measurement based evaluation of wound healing10.1088/1361-6579/aa63d6 ↗
5. Ching, C.T.-S., Chen, Y.-C., Lu, L.-H., Hsieh, P.F., Hsiao, C.-S., Sun, T.-P., Shieh, H.-L., and Chang, K.-M. (2013). Characterization of the Muscle Electrical Properties in Low Back Pain Patients by Electrical Impedance Myography. PLoS ONE, 8.10.1371/journal.pone.0061639 ↗
6. A CMOS Magnitude/Phase Measurement Chip for Impedance Spectroscopy10.1109/JSEN.2013.2251628 ↗
7. High-power CMOS current driver with accurate transconductance for electrical impedance tomography10.1109/TBCAS.2013.2285481 ↗
8. The bio-oscillator: A circuit for cell-culture assays
9. Fundamentals, Recent Advances, and Future Challenges in Bioimpedance Devices for Healthcare Applications
10. An integrated analog readout for multi-frequency bioimpedance measurements10.1109/JSEN.2014.2315963 ↗
11. Electrical Impedance Myography and Its Applications in Neuromuscular Disorders10.1007/s13311-016-0491-x ↗
12. Critcher, S., and Freeborn, T.J. (2021). Localized bioimpedance measurements with the max3000x integrated circuit: Characterization and demonstration. Sensors, 21.10.3390/s21093013 ↗
13. (2023, December 02). Texas Instruments AFE4300 4 Device and Documentation Support 4.1 Receiving Notification of Documentation Updates. Available online: www.ti.com.
14. (2024, January 12). Analog Devices MAX30001-Ultra-Low-Power, Single-Channel Integrated Biopotential (ECG, R-to-R, and Pace Detection) and Bioimpedance (BioZ) AFE. Available online: www.maximintegrated.com/thermal-tutorial.
15. (2024, January 12). Analog Devices MAX30009EVKIT# Evaluation Kit General Description. Available online: www.maximintegrated.com/products/MAX30009.
16. Lu, F., Wang, C., Zhao, R., Du, L., Fang, Z., Guo, X., and Zhao, Z. (2018). Review of stratum corneum impedance measurement in non-invasive penetration application. Biosensors, 8.10.3390/bios8020031 ↗
17. Guidelines to electrode positioning for human and animal electrical impedance myography research10.1038/srep32615 ↗
18. Discriminating neurogenic from myopathic disease via measurement of muscle anisotropy10.1002/mus.21115 ↗
19. Thin film non-symmetric microelectrode array for impedance monitoring of human skin10.1016/S0040-6090(03)00389-4 ↗
20. Bioimpedance basics and phase angle fundamentals10.1007/s11154-022-09780-3 ↗
21. High Accurate Howland Current Source: Output Constraints Analysis10.4236/cs.2013.47059 ↗
22. Algueta-Miguel, J.M., Beato-López, J.J., and López-Martín, A.J. (2023). Analog Lock-In Amplifier Design Using Subsampling for Accuracy Enhancement in GMI Sensor Applications. Sensors, 23.10.3390/s23010057 ↗
23. Synchronous sampling and demodulation in an instrument for multifrequency bioimpedance measurement10.1109/TIM.2007.900163 ↗
24. A fully differential synchronous demodulator for AC signals10.1109/TIM.2019.2895481 ↗
25. Zamani, M., Rezaeiyan, Y., Shoaei, O., and Moradi, F. (2021, January 26–27). A 7.2μW Magnitude/Phase Bio-impedance Measurement Front-End with PWM Output in 0.18 μm CMOS. Proceedings of the 2021 IEEE Nordic Circuits and Systems Conference, NORCAS 2021—Proceedings, Oslo, Norway.10.1109/NorCAS53631.2021.9599647 ↗
26. Howland current source for high impedance load applications10.1063/1.5005330 ↗
27. Mirrored enhanced Howland current source with feedback control10.1063/1.5079872 ↗
28. Electrical impedance characterization of normal and cancerous human hepatic tissue10.1088/0967-3334/31/7/009 ↗
29. Yamaji, T., Tanimoto, H., and Kokatsu, H. (1998, January 5–7). An I/Q Active Balanced Harmonic Mixer with IM2 Cancelers and a 45 Phase Shifter. Proceedings of the 1998 IEEE International Solid-State Circuits Conference. Digest of Technical Papers, ISSCC. First Edition (Cat. No.98CH36156), San Francisco, CA, USA.10.1109/4.735708 ↗
30. Analog Devices (1978). Multiplication Application Guide, Analog Devices, Inc.
31. Spatial resolution in electrical impedance tomography: A topical review10.5617/jeb.3350 ↗
32. (2024, January 10). Texas Instruments Noise Analysis in Operational Amplifier Circuits 2007 Digital Signal Processing Solutions Application Report. Available online: https://www.ti.com/lit/an/slva043b/slva043b.pdf?ts=1718065789407&ref_url=https%253A%252F%252Fwww.google.com.hk%252F.
33. (2024, January 10). I. Corporation Making Accurate Voltage Noise and Current Noise Measurements on Operational Amplifiers Down to 0.1 Hz. Available online: www.intersil.com.
34. (2023, December 02). High Precision Foil Resistor with TCR of ± 2.0 ppm/°C, Tolerance of ± 0.005% and Load Life Stability of ± 0.005% RCK Series Vishay Foil Resistors. Available online: www.vishayfoilresistors.com.
35. Multi-frequency bioelectrical impedance analysis of skin rubor with two-electrode technique10.1016/j.jtv.2008.01.002 ↗