Experimental Assessment of the Spatial and Temporal Distribution of Non-Contact Tonometer Airflows

Daniela Oehring, David Jenkins, Phillip Jonathan Buckhurst, Hetal Buckhurst

Applied Sciences· Volume 11· Issue 6

Abstract

(1) The aim of the study was to investigate the spatial and temporal characteristics of the airflow created by two commercially available non-contact tonometers: the CorvisST and the Ocular Response Analyser (ORA). (2) The airflow pressure was measured using a microelectromechanical system (MEMS) pressure sensor to investigate the spatial and temporal distribution. The airflow from the CorvisST and Ocular Response Analyser were mapped to a 600 µm and a 1 mm resolution grid, respectively. (3) Central airflow pressure of the CorvisST (96.4 ± 1.4 mmHg) was higher than that of the Ocular Response Analyser (91.7 ± 0.7 mmHg). The duration of the air-puffs also differed, with the CorvisST showing a shorter duration (21.483 ± 0.2881 ms) than that of the ORA (23.061 ± 0.1872 ms). The rising edge of the CorvisST airflow pressure profile demonstrated a lower gradient (+8.94 mmHg/ms) compared to that of the Ocular Response Analyser (+11.00 mmHg/ms). Both had similar decay response edges: CorvisST −11.18 mmHg/ms, Ocular Response Analyser −11.65 mmHg/ms. (4) The study presents a valid method to investigate the physical dimensions of the airflow pressure of non-contact tonometers. Novel findings relating to the magnitude, duration and spatial characteristics of the respective airflow pressures are reported. It is anticipated that this information will better inform clinical studies and theoretical models relating to ocular biomechanics.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Corneal surgery and disorders · Ocular Surface and Contact Lens · Glaucoma and retinal disorders

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

Faculty of Health, School of Health Professions, University of Plymouth, Plymouth PL6 8AA, UK

Faculty of Science and Engineering, School of Engineering, Computing and Mathematics, University of Plymouth, Plymouth PL4 8AA, UK

Funding

Plymouth University · College of Optometrists

References (56)

Show source-provided references
1. Corneal deformation measurement using Scheimpflug noncontact tonometry10.1097/OPX.0b013e318279eb87 ↗
2. Biomechanical diagnostics of the cornea10.1186/s40662-020-0174-x ↗
3. Determination of Corneal Biomechanical Behavior in-vivo for Healthy Eyes Using CorVis ST Tonometry: Stress-Strain Index10.3389/fbioe.2019.00105 ↗
4. Determining in vivo biomechanical properties of the cornea with an ocular response analyzer10.1016/j.jcrs.2004.10.044 ↗
5. The range of the waveform score of the ocular response analyzer (ora) in healthy subjects
6. Relationship among corneal stiffness, thickness, and biomechanical parameters measured by Corvis ST, Pentacam and ORA in keratoconus10.3389/fphys.2019.00740 ↗
7. Comparisons of the corneal biomechanics between low and high myopic eyes–a meta-analysis10.1016/j.ajo.2019.07.007 ↗
8. Corneal Biomechanical Properties Characterization Using Air-jet Indentation Based Optical Coherence Tomography System (AIOCT)10.1051/matecconf/201925601004 ↗
9. Biomechanical changes in the cornea following cataract surgery: A prospective assessment with the Corneal Visualisation Scheimpflug Technology
10. Corneal Curvature: The Influence of Corneal Accommodation and Biomechanics on Corneal Shape10.1167/tvst.8.4.5 ↗
11. Changes of corneal biomechanics with keratoconus10.1097/ICO.0b013e318243e42d ↗
12. Detection of Keratoconus With a New Biomechanical Index10.3928/1081597X-20160629-01 ↗
13. Fundamental physics of ultrasound and its propagation in tissue10.1148/radiographics.13.3.8316679 ↗
14. Fatt, I., and Weissman, B.A. (2013). Physiology of the Eye: An. Introduction to the Vegetative Functions, Elsevier Science.
15. A new tonometer system10.1097/00006324-197208000-00005 ↗
16. Dynamic OCT measurement of corneal deformation by an air puff in normal and cross-linked corneas10.1364/BOE.3.000473 ↗
17. Assessment of the Ocular Response Analyzer as an Instrument for Measurement of Intraocular Pressure and Corneal Biomechanics10.3109/02713683.2014.978479 ↗
18. Theoretical and numerical analysis of the corneal air puff test10.1016/j.jmps.2016.04.012 ↗
19. The influence of intraocular pressure and air jet pressure on corneal contactless tonometry tests10.1016/j.jmbbm.2015.07.030 ↗
20. Fluid–structure simulation of a general non-contact tonometry. A required complexity?10.1016/j.cma.2018.05.031 ↗
21. Kling, S., Bekesi, N., Dorronsoro, C., Pascual, D., and Marcos, S. (2014). Corneal viscoelastic properties from finite-element analysis of in vivo air-puff deformation. PLoS ONE, 9.10.1371/journal.pone.0104904 ↗
22. Metzler, K., Roberts, C., Whitaker, S., Lawrence, M., Malik, J., and Bons, J. (2013). Modeling corneal response to an air puff using deformation data to derive Young’s modulus. ARVO, Investigative Ophthalmology & Visual Science.
23. Metzler, K., Roberts, C., Whitaker, S., Lawrence, M., Malik, J., and Bons, J. (2013). The Influence of IOP, Corneal Stiffness, and Sclera on Corneal Deformation Response to an Air Puff Using the CorVis ST, The Ohio State University, Investigative Ophthalmology & Visual Science.
24. Bahr, N., Ali, N., Patel, D., McGhee, C., Hunter, P., and Ho, H. (2015). Modelling the Deformation of the Human Cornea Produced by a Focussed Air Pulse. Comput. Biomech. Med., 93–100.10.1007/978-3-319-15503-6_9 ↗
25. Air–jet temporal and spatial pressure properties of the reichert ocular response analyzer (ORA)
26. Asaoka, R., Nakakura, S., Tabuchi, H., Murata, H., Nakao, Y., Ihara, N., Rimayanti, U., Aihara, M., and Kiuchi, Y. (2015). The Relationship between Corvis ST Tonometry Measured Corneal Parameters and Intraocular Pressure, Corneal Thickness and Corneal Curvature. PLoS ONE, 10.10.1371/journal.pone.0140385 ↗
27. Correlation of Biomechanic Parameters Measured by Corvis ST (Oculus®) and by Ocular Response Analyzer (ORA, Reichert®)
28. Matsuura, M., Hirasawa, K., Murata, H., Yanagisawa, M., Nakao, Y., Nakakura, S., Kiuchi, Y., and Asaoka, R. (2016). The Relationship between Corvis ST Tonometry and Ocular Response Analyzer Measurements in Eyes with Glaucoma. PLoS ONE, 11.10.1371/journal.pone.0161742 ↗
29. Tejwani, S., Shetty, R., Kurien, M., Dinakaran, S., Ghosh, A., and Roy, A.S. (2014). Biomechanics of the cornea evaluated by spectral analysis of waveforms from ocular response analyzer and Corvis-ST. PLoS ONE, 9.10.1371/journal.pone.0097591 ↗
30. Size does matter: What is the corneo-limbal diameter?10.1111/cxo.12583 ↗
31. The shape of the anterior and posterior surface of the aging human cornea10.1016/j.visres.2005.09.021 ↗
32. A review of mathematical descriptors of corneal asphericity
33. Corneal Viscous Properties Cannot Be Determined From Air-Puff Applanation10.3928/1081597X-20191010-03 ↗
34. Can the Corvis ST Estimate Corneal Viscoelasticity?10.3928/1081597X-20200212-04 ↗
35. Enhanced Ectasia Detection Using Corneal Tomography and Biomechanics10.1016/j.ajo.2018.08.054 ↗
36. Biomechanically-Corrected Intraocular Pressure Compared To Pressure Measured With Commonly Used Tonometers In Normal Subjects10.2147/OPTO.S220776 ↗
37. Corneal hysteresis measured with the Ocular Response Analyzer in normal and glaucomatous eyes10.1111/j.1755-3768.2009.01554.x ↗
38. Biomechanical properties of the cornea in Fuchs' corneal dystrophy10.1167/iovs.08-3312 ↗
39. A pilot study on the corneal biomechanical changes in short-term orthokeratology10.1111/j.1475-1313.2008.00625.x ↗
40. Biomechanical properties of the cornea measured by the Ocular Response Analyzer and their association with intraocular pressure and the central corneal curvature10.1111/j.1444-0938.2009.00414.x ↗
41. Corneal hysteresis using the Reichert ocular response analyser: Findings pre-and post-LASIK and LASEK10.1111/j.1600-0420.2007.01023.x ↗
42. Assessment of the biomechanical properties of the cornea with the ocular response analyzer in normal and keratoconic eyes10.1167/iovs.04-0694 ↗
43. Intraoffice variability of corneal biomechanical parameters and intraocular pressure (IOP)10.1097/OPX.0b013e3181783a5f ↗
44. Effect of aging on corneal biomechanical parameters using the ocular response analyzer10.3928/1081597X-20090917-10 ↗
45. Relationship between corneal biomechanical properties, central corneal thickness, and intraocular pressure across the spectrum of glaucoma10.1016/j.ajo.2011.10.032 ↗
46. Corneal hysteresis and intraocular pressure measurement in children using the Reichert ocular response analyzer10.1016/j.ajo.2006.07.058 ↗
47. Comparison of IOP measurements between ORA and GAT in normal Chinese10.1097/OPX.0b013e3181559db2 ↗
48. Cornea biomechanical characteristics and their correlates with refractive error in Singaporean children10.1167/iovs.07-1670 ↗
49. Central corneal thickness and corneal hysteresis during corneal swelling induced by contact lens wear with eye closure10.1016/j.ajo.2006.12.031 ↗
50. Diurnal variation of ocular hysteresis, corneal thickness, and intraocular pressure10.1097/OPX.0b013e31818e8abe ↗
51. Corneal hysteresis and axial length among Chinese secondary school children: The Xichang Pediatric Refractive Error Study (X-PRES) report no. 410.1016/j.ajo.2007.12.034 ↗
52. Correlations between corneal hysteresis, intraocular pressure, and corneal central pachymetry10.1016/j.jcrs.2007.11.051 ↗
53. Intraocular pressure measurement using Ocular response analyzer, dynamic contour tonometer, and scheimpflug analyzer Corvis ST10.1155/2019/3879651 ↗
54. Agreement in central corneal thickness measurement between Corvis ST and ocular response analyzer10.1007/s10792-020-01436-y ↗
55. The relationship between corvis ST tonometry parameters and ocular response analyzer corneal hysteresis10.1097/IJG.0000000000001486 ↗
56. Effect of central corneal thickness and corneal hysteresis on tonometry as measured by dynamic contour tonometry, ocular response analyzer, and Goldmann tonometry in glaucomatous eyes10.1097/IJG.0b013e31815c3ad3 ↗