Diagnostic Performance and Radiation Dose of the EOS System to Image Enchondromatosis: A Phantom Study

Domenico Albano, Alessandro Loria, Cristiana Fanciullo, Alberto Bruno, Carmelo Messina, Antonella del Vecchio, Luca Maria Sconfienza

Applied Sciences· Volume 10· Issue 24

Abstract

Background: Radiation doses and capability of EOS, conventional radiography (CR), and computed tomography (CT) to detect and measure enchondromas in a dedicated five-year-old anthropomorphic phantom were compared. Methods: To simulate enchondromas, minced pieces of chicken bone and cartilage were packed in conventional kitchen plastic foil to create ovoidal/rounded masses and randomly hung on the phantom. The phantom was imaged five times with CR, CT, and EOS, each time changing the number and position of inserts. All images were reviewed by a senior radiologist and a radiology resident. Results: EOS and CR detected all inserts in 4/5 cases (80%), while in one case 1/17 inserts was not seen. Excellent agreement of EOS with CR (88% reproducibility; bias = 14 mm; repeatability coefficient (CoR) 2.9; 95% CI from −2.8 to 3.1 mm; p = 0.5) and CT (81% reproducibility; bias = 15 mm; CoR 5.2; 95% CI from −5.5 to 5.2 mm; p = 0.7) was found. EOS showed 71% interobserver reproducibility (CoR 7.2; bias = 0.6 mm; 95% CI from −6.6 to 7.8 mm; p = 0.25). The EOS-Fast radiation dose was also significantly lower than the median radiation dose of CR (644.7 (599.4–651.97) mGy•cm2, p = 0.004). Conclusions: Low-dose EOS has the same capability as CR to detect and measure enchondroma-like inserts on a phantom and may be considered to monitor patients with multiple enchondromas.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Bone Tumor Diagnosis and Treatments · Oral and Maxillofacial Pathology · Sarcoma Diagnosis and Treatment

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

IRCCS Istituto Ortopedico Galeazzi, 20161 Milano, Italy

Sezione di Scienze Radiologiche, Dipartimento di Biomedicina, Neuroscienze e Diagnostica Avanzata, Università di Palermo, 90127 Palermo, Italy

Servizio di Fisica Sanitaria, IRCCS Ospedale San Raffaele, 20132 Milano, Italy

Postgraduate School of Diagnostic and Interventional Radiology, University of Milan, 20122 Milano, Italy

Centro Salus, 90145 Palermo, Italy

Dipartimento di Scienze Biomediche per la Salute, Università degli Studi di Milano, 20133 Milano, Italy

References (18)

Show source-provided references
1. Differential Diagnosis of Spine Tumors: My Favorite Mistake10.1055/s-0038-1675551 ↗
2. MRI radiomics-based machine-learning classification of bone chondrosarcoma10.1016/j.ejrad.2020.109043 ↗
3. Maroteaux, P., and Le Merrer, M. (2002). Les Maladies Osseuses de L’enfant, Médecine-Sciences Flammarion.
4. Cartilaginous lesions of bone10.1007/s007760170015 ↗
5. Whyte, M. (2003). Acquired disorders of cartilage and bone. Am. Soc. Bone Miner. Res.
6. Fletcher, C.D.M., Unni, K.K., and Mertens, F. (2002). Pathology and Genetics of Tumours of Soft Tissue and Bone, Iarc.
7. A mutant PTH/PTHrP type I receptor in enchondromatosis10.1038/ng844 ↗
8. ALARA in pediatric interventional and fluoroscopic imaging: Striving to keep radiation doses as low as possible during fluoroscopy of pediatric patients—A white paper executive summary10.1016/j.jacr.2006.04.008 ↗
9. (2007). The 2007 Recommendations of the International Commission on Radiological Protection. Ann. ICRP, 37, 1–332. ICRP publication 103.
10. EOS((R)) biplanar X-ray imaging: Concept, developments, benefits, and limitations10.1007/s11832-016-0713-0 ↗
11. A new 2D and 3D imaging approach to musculoskeletal physiology and pathology with low-dose radiation and the standing position: The EOS system
12. EOS stereo-radiography system: Whole-body simultaneous anteroposterior and lateral radiographs with very low radiation dose10.1016/S0035-1040(07)92729-4 ↗
13. Segmented lordotic angles to assess lumbosacral transitional vertebra on EOS10.1007/s00586-020-06565-7 ↗
14. Diagnostic imaging of spinal deformities: Reducing patients radiation dose with a new slot-scanning X-ray imager10.1097/BRS.0b013e3181bdcaa4 ↗
15. Statistical methods for assessing agreement between two methods of clinical measurement10.1016/S0140-6736(86)90837-8 ↗
16. Kerma-area product in diagnostic radiology10.2214/AJR.14.12513 ↗
17. Statistical significance: P value, 0.05 threshold, and applications to radiomics-reasons for a conservative approach10.1186/s41747-020-0145-y ↗
18. Evaluation of reproducibility of the MOCART score in patients with osteochondral lesions of the talus repaired using the autologous matrix-induced chondrogenesis technique10.1007/s11547-017-0794-y ↗