Accuracy of Two Variants of 3D-Printed Insertion Guides for Orthodontic Mini-Implants: An Ex Vivo Study in Human Cadavers
Source abstract
Abstract
Insertion guides are becoming popular for orthodontic mini-implant positioning. The aim of this study was to evaluate and compare the accuracy of two different mini-implant insertion guides, with or without pre-drilling, in a human cadaveric model. Maxillary casts of six fresh frozen specimens were digitized to create insertion guides. Sixty mini-implants were randomly inserted with full-arch or skeletonized guides, either with or without predrilling. Pre- and post-treatment CBCTs were superimposed using rigid registration. Transformation matrices of the planned and real positions were obtained, and distances at the mini-implant neck and apex, as well as the angular deviation, were calculated. The Kruskal–Wallis test was performed, followed by a post hoc test when indicated. Out of 60 inserted mini-implants, 46 could be evaluated. Of these, 10 initially assigned to no pre-drilling required this procedure due to very high bone density. Therefore, 32 implants were inserted with pre-drilling (n = 15 full-arch; n = 17 skeletonized) and 14 without (n = 7 full-arch; n = 7 skeletonized). The lowest mean deviation at the neck was 1.22 ± 0.6 mm, registered in the full-arch/pre-drilling group. The skeletonized/no pre-drilling group presented the lowest mean values at the apex, i.e., 1.72 ± 1.22 mm, as well as the lowest mean angular deviation, i.e., 8.23 ± 4.24°. Significant differences among groups were observed only at the neck, with higher mean deviation in the skeletonized/pre-drilling group than in the full-arch/pre-drilling one (p = 0.014). In conclusion, within the limitations of the study, rather high deviations between planned and real mini-implant positions were found. Further studies are needed on how to improve the accuracy within in vivo settings.
Source: crossref · Quality: baseline
Research context
OpenAlex topics: Orthodontics and Dentofacial Orthopedics · Dental Implant Techniques and Outcomes · Temporomandibular Joint Disorders
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
Affiliations
Department of Orthodontics, University Hospital of Düsseldorf, 40225 Düsseldorf, Germany
Department of Neurosciences, School of Dentistry, University of Padova, 35128 Padova, Italy
Department of Oral Surgery, University Hospital of Düsseldorf, 40225 Düsseldorf, Germany
Institute for Anatomy II, University Hospital of Düsseldorf, 40225 Düsseldorf, Germany
Funding
PSM Medical · Easy Driver
References (32)
Show source-provided references
1. Risk factors for orthodontic mini-implants in skeletal anchorage biological stability: A systematic literature review and meta-analysis10.1038/s41598-020-62838-7 ↗
2. Acceptance rate of palatal implants: A questionnaire study10.1016/j.ajodo.2004.06.031 ↗
3. Comparison of treatment outcomes between skeletal anchorage and extraoral anchorage in adults with maxillary dentoalveolar protrusion10.1016/j.ajodo.2006.12.022 ↗
4. Maxillary protraction with rapid maxillary expansion and facemask versus skeletal anchorage with mini-implants in class III patients: A non-randomized clinical trial10.1186/s40510-019-0288-7 ↗
5. Safe zones for miniscrews in maxillary dentition distalization assessed with cone-beam computed tomography10.1016/j.ajodo.2016.07.021 ↗
6. Sbricoli, L., Ricci, S., Cattozzo, A., Favero, R., Bressan, E., and Sivolella, S. (2022). Mandibular Molar Uprighting Using Skeletal Anchorage: A Novel Approach. J. Clin. Med., 11.10.3390/jcm11133565 ↗
7. Skeletally anchored mesialization of molars using digitized casts and two surface-matching approaches: Analysis of treatment effects10.1007/s00056-017-0108-y ↗
8. The T-Zone: Median vs. Paramedian Insertion of Palatal Mini-Implants
9. Survival estimates and risk factors for failure of palatal and buccal mini-implants10.2319/090720-777.1 ↗
10. Survival of palatal miniscrews used for orthodontic appliance anchorage: A retrospective cohort study10.1016/j.ajodo.2013.01.018 ↗
11. Three dimensional anatomical exploration of the anterior hard palate at the level of the third ruga for the placement of mini-implants--a cone-beam CT study10.1093/ejo/cju093 ↗
12. Vertical palatal bone dimensions on lateral cephalometry and cone-beam computed tomography: Implications for palatal implant placement10.1111/j.1600-0501.2010.02021.x ↗
13. Procedure using CAD/CAM-manufactured insertion guides for purely mini-implant-borne rapid maxillary expanders10.1007/s00056-022-00375-w ↗
14. Comparative evaluation of insertion torque and mechanical stability for self-tapping and self-drilling orthodontic miniscrews—An in vitro study10.1186/s13005-017-0143-3 ↗
15. Impact of bone quality, implant type, and implantation site preparation on insertion torques of mini-implants used for orthodontic anchorage10.1016/j.ijom.2010.08.008 ↗
16. Effects of pilot hole size and bone density on miniscrew implants’ stability10.1111/j.1708-8208.2010.00269.x ↗
17. Placement and removal torque values of orthodontic miniscrew implants10.1016/j.ajodo.2010.11.017 ↗
18. Biomechanical and histological comparison of self-drilling and self-tapping orthodontic microimplants in dogs10.1016/j.ajodo.2007.01.023 ↗
19. Influence of implant taper on the primary and secondary stability of osseointegrated titanium implants10.1111/j.1600-0501.2004.01041.x ↗
20. Relationship between insertion torque and removal torque analyzed in fresh temporal bone
21. Influence of pre-drilling diameter on the stability of orthodontic anchoring screws in the mid-palatal area10.2334/josnusd.21-0012 ↗
22. Accuracy evaluation of computer-guided implant surgery associated with prototyped surgical guides10.1016/j.prosdent.2019.07.010 ↗
23. Current state of the art of computer-guided implant surgery10.1111/prd.12175 ↗
24. The easy driver for placement of palatal mini-implants and a maxillary expander in a single appointment
25. Accuracy of orthodontic mini-implants placed at the anterior palate by tooth-borne or gingiva-borne guide support: A cadaveric study10.1007/s00784-019-02885-1 ↗
26. Accuracy of fully guided orthodontic mini-implant placement evaluated by cone-beam computed tomography: A study involving human cadaver heads10.1007/s00784-020-03436-9 ↗
27. CAD-CAM-fabricated mini-implant insertion guides for the delivery of a distalization appliance in a single appointment10.1016/j.ajodo.2018.12.017 ↗
28. R Core Team (2023, July 03). R: A Language and Environment for Statistical Computing. R Foundation for Statistical Computing, Vienna, Austria. Available online: https://www.R-project.org/.
29. Accurate pre-surgical determination for self-drilling miniscrew implant placement using surgical guides and cone-beam computed tomography10.1093/ejo/cjq012 ↗
30. Insertion angle impact on primary stability of orthodontic mini-implants10.2319/100707-484.1 ↗
31. Accuracy of palatal orthodontic mini-implants placed by conventionally or CAD/CAM-based surgical guides: A comparative in vitro study10.2319/011722-55.1 ↗
32. An in vitro investigation of peak insertion torque values of six commercially available mini-implants10.1093/ejo/cjq129 ↗