Abstract

Diagnostic reference levels (DRLs) are critical operational standards for monitoring radiological imaging parameters and ensuring patient safety in hospitals. A regular review of DRLs is necessary to optimize scanners and protocol parameters. This retrospective study collected data on the CT dose index volume and dose–length product for 250 children who underwent head CT scans in the region of Abu Dhabi, United Arab Emirates. Descriptive statistics were used to present median, range, and interquartile range values for each pediatric age group, based on region and protocol. The 75th percentile was used as the reference point for local DRLs. Proposed values of DRLs were suggested per age groups; 304 mGy∙cm (children < 1 year), 385 mGy∙cm (children ≥ 1 year to <5 years), 441 mGy∙cm (children ≥ 5 years to <10 years), and 568 mGy∙cm (patients ≥ 10 years to <16 years. A comparison of the local DRLs with previously established ones was carried out, and differences were discussed. To the best of our knowledge, this is the first study on pediatric head CT scans in this region. We believe the results can be used as a baseline for establishing local DRLs in the Emirate of Abu Dhabi and national DRLs in the United Arab Emirates.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Radiation Dose and Imaging · Advanced X-ray and CT Imaging · Ultrasound in Clinical Applications

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

School of Physics, Universiti Sains Malaysia (USM), Gelugor 11800, Malaysia

Department of Physics, Khalifa University of Science and Technology, Abu Dhabi 127788, United Arab Emirates

Funding

Khalifa University of Science and Technology

References (22)

Show source-provided references
1. Morigi, M.P., and Albertin, F. (2022). X-ray Digital Radiography and Computed Tomography. J. Imaging, 8.10.3390/jimaging8050119 ↗
2. Medical Physics in Europe Following Recommendations of the International Atomic Energy Agency10.1515/raon-2016-0004 ↗
3. Patient Radiation Exposure: Imaging during Radiation Oncology Procedures: Executive Summary of NCRP Report No. 184. J. Am10.1016/j.jacr.2020.02.002 ↗
4. European Commission (2014). Directive 2013/59/EURATOM of 5 December 2013 laying down basic safety standards for protection against the dangers arising from exposure to ionising radiation, and repealing Directives 89/618/Euratom, 90/641/Euratom, 96/29/Euratom, 97/43/Euratom and 2003/122/Euratom. Off. J. Eur. Union, L13, 1–73. Available online: https://eur-lex.europa.eu/legal-content/EN/TXT/PDF/?uri=CELEX:32013L0059&from=EN.
5. ICRP, Khong, P.L., Ringertz, H., Donoghue, V., Frush, D., Rehani, M., Appelgate, K., and Sanchez, R. (2013). ICRP Publication 121: Radiological Protection in Paediatric Diagnostic and Interventional Radiology. Ann. ICRP, 42, 1–63.10.1016/j.icrp.2012.10.001 ↗
6. Zheng, X., Gutsche, L., Al-Hayek, Y., Stanton, J., Elshami, W., and Jensen, K. (2021). Impacts of Phantom Off-Center Positioning on CT Numbers and Dose Index CTDIv: An Evaluation of Two CT Scanners from GE. J. Imaging, 7.10.3390/jimaging7110235 ↗
7. ICRP Publication 135: Diagnostic Reference Levels in Medical Imaging10.1177/0146645317717209 ↗
8. Al-Othman, A.Y., Al-Sharydah, A.M., Abuelhia, E.I., Mohtasib, R., Bin Dahmash, A., Hegazi, T.M., Tajaldeen, A.A., Alshehri, S.S., Al-Malki, F.M., and Alghamdi, S. (2022). Radiation Dose Optimization Based on Saudi National Diagnostic Reference Levels and Effective Dose Calculation for Computed Tomography Imaging: A Unicentral Cohort Study. Appl. Sci., 12.10.3390/app122211504 ↗
9. Radiation exposure from CT scans in childhood and subsequent risk of leukaemia and brain tumours: A retrospective cohort study10.1016/S0140-6736(12)60815-0 ↗
10. Cancer risk in 680,000 people exposed to computed tomography scans in childhood or adolescence: Data linkage study of 11 million Australians10.1136/bmj.f2360 ↗
11. (2022, April 13). ICRP Publication 73. Available online: https://www.icrp.org/publication.asp?id=ICRP%20Publication%2073.
12. (2022, April 13). Radiation Protection No 185—European Guidelines on Diagnostic Reference Levels for Pediatric Imaging. Available online: http://www.eurosafeimaging.org/wp/wp-content/uploads/2018/10/European-guidelines-on-diagnostic-reference-levels-for-pediatric-imaging.pdf.
13. Long-term experience and analysis of data on diagnostic reference levels: The good, the bad, and the ugly10.1007/s00330-019-06422-2 ↗
14. (2021, April 13). SEHA Website. Available online: https://www.seha.ae.
15. (2021, April 13). FANR Website. National Diagnostic Reference Levels, Available online: https://www.fanr.gov.ae/en/services/others/national-diagnostic-reference-levels.
16. The Use of Computed Tomography in Pediatrics and the Associated Radiation Exposure and Estimated Cancer Risk10.1001/jamapediatrics.2013.311 ↗
17. Diagnostic reference levels for paediatric CT in Jordan10.1088/1361-6498/ab3ee2 ↗
18. A Multi-Provincial Survey and Analysis of Radiation Doses from Pediatric CT in China10.1016/j.radmp.2021.02.001 ↗
19. Patient Doses from CT Examinations in Turkey10.5152/dir.2015.14306 ↗
20. Muhammad, N.A., Abdul Karim, M.K., Abu Hassan, H., Ahmad Kamarudin, M., Ding Wong, J.H., and Ng, K.H. (2020). Diagnostic Reference Level of Radiation Dose and Image Quality among Paediatric CT Examinations in A Tertiary Hospital in Malaysia. Diagnostics, 10.10.3390/diagnostics10080591 ↗
21. Developing Diagnostic Reference Levels in Japan10.1007/s11604-020-01066-5 ↗
22. European Guidelines on Diagnostic Reference Levels for Paediatric Imaging