Optimization of Replanning Processes for Volumetric Modulated Arc Therapy Plans at Risk of QA Failure Predicted by a Machine Learning Model
Source abstract
Abstract
Patient-specific quality assurance (PSQA) procedures ensure the safe delivery of volumetric modulated arc therapy (VMAT) plans. PSQA requires extensive time and resources and may cause treatment delays if replanning is needed due to failures. Recently, our group developed a machine learning (ML) model predicting gamma passing rate (GPR) for VMAT arcs. This study explores automatable replanning strategies for plans identified at risk of failure, aiming to improve deliverability while maintaining dosimetric quality. Between 2022 and 2023, our ML model analyzed 1252 VMAT plans. Ten patients having a predicted GPR (pGPR) <95% were selected. Replanning strategies consisted of limiting monitor units (MUlimit) and employing the aperture shape controller (ASC) tool. Re-optimized plans were compared with the originals in terms of dose volume constraints (DVCs) for the target and organs-at-risk (OARs), and deliverability using the modulation complexity score (MCS), pGPR, and measured GPR (mGPR). Forty-five re-optimizations were performed. Replanning led to an increase in DVCs for OARs and a reduction for the target. Complexity decreased, reflected by the increase in the MCS from 0.17 to 0.21 (MUlimit) and 0.20 (ASC). The deliverability improved, with the pGPR increasing from 93.3% to 94.4% (MUlimit) and 95.1% (ASC), and the mGPR from 99.3% to 99.7% (MUlimit) and 99.8% (ASC). Limiting the MUs or utilizing the ASC reduced the complexity of plans and improved the GPR without compromising the dosimetric quality. These strategies can be used to automate replanning procedures, reduce the workload related to PSQA, and improve patient safety.
Source: crossref · Quality: baseline
Research context
OpenAlex topics: Advanced Radiotherapy Techniques · Advanced X-ray and CT Imaging · Radiation Dose and Imaging
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
Affiliations
Radiotherapy and Radiosurgery Department, IRCCS Humanitas Research Hospital, Via Manzoni 56, 20089 Milan, Italy
Department of Biomedical Sciences, Humanitas University, Via Rita Levi Montalcini 4, 20072 Milan, Italy
Dipartimento di Fisica “Aldo Pontremoli”, Università degli Studi di Milano, 20133 Milan, Italy
National Institute for Nuclear Physics (INFN), Milan Division, 20133 Milan, Italy
Funding
Ministero della Salute
References (30)
Show source-provided references
1. Tolerance Limits and Methodologies for IMRT Measurement-Based Verification QA: Recommendations of AAPM Task Group. No. 21810.1002/mp.12810 ↗
2. Dosimetry Tools and Techniques for IMRT10.1118/1.3514120 ↗
3. Challenges in Calculation of the Gamma Index in Radiotherapy—Towards Good Practice10.1016/j.ejmp.2017.03.001 ↗
4. Chan, M.F., Witztum, A., and Valdes, G. (2020). Integration of AI and Machine Learning in Radiotherapy QA. Front. Artif. Intell., 3.10.3389/frai.2020.577620 ↗
5. Applications of Machine and Deep Learning to Patient-specific IMRT/VMAT Quality Assurance10.1002/acm2.13375 ↗
6. Volumetric Modulated Arc Therapy: IMRT in a Single Gantry Arc10.1118/1.2818738 ↗
7. Volumetric Modulated Arc Therapy: A Review of Current Literature and Clinical Use in Practice10.1259/bjr/22373346 ↗
8. Multicentric Evaluation of a Machine Learning Model to Streamline the Radiotherapy Patient Specific Quality Assurance Process10.1016/j.ejmp.2023.102593 ↗
9. Use of Metrics to Quantify IMRT and VMAT Treatment Plan Complexity: A Systematic Review and Perspectives10.1016/j.ejmp.2019.05.024 ↗
10. Complexity Metrics for IMRT and VMAT Plans: A Review of Current Literature and Applications10.1259/bjr.20190270 ↗
11. Small Fields: Nonequilibrium Radiation Dosimetry10.1118/1.2815356 ↗
12. Influence of the Jaw Tracking Technique on the Dose Calculation Accuracy of Small Field VMAT Plans10.1002/acm2.12029 ↗
13. A Closer Look at RapidArc® Radiosurgery Plans Using Very Small Fields10.1088/0031-9155/56/6/020 ↗
14. Wolfs, C.J.A., Swinnen, A.C.C., Nijsten, S.M.J.J.G., and Verhaegen, F. (2018). Should Dose from Small Fields Be Limited for Dose Verification Procedures?: Uncertainty versus Small Field Dose in VMAT Treatments. Phys. Med. Biol., 63.10.1088/1361-6560/aae338 ↗
15. Pseudo Skin Flash on VMAT in Breast Radiotherapy: Optimization of Virtual Bolus Thickness and HU Values10.1016/j.ejmp.2019.05.010 ↗
16. Assessment of Monitor Unit Limiting Strategy Using Volumetric Modulated Arc Therapy for Cancer of Hypopharynx10.1016/j.ejmp.2017.01.016 ↗
17. Total Monitor Units Influence on Plan Quality Parameters in Volumetric Modulated Arc Therapy for Breast Case10.1016/j.ejmp.2013.08.142 ↗
18. Limiting Treatment Plan Complexity by Applying a Novel Commercial Tool10.1002/acm2.12908 ↗
19. Investigating the Use of Aperture Shape Controller in VMAT Treatment Deliveries10.1016/j.meddos.2020.02.003 ↗
20. A New Metric for Assessing IMRT Modulation Complexity and Plan Deliverability10.1118/1.3276775 ↗
21. Automation in Radiotherapy Treatment Planning: Examples of Use in Clinical Practice and Future Trends for a Complete Automated Workflow10.1016/j.canrad.2021.06.006 ↗
22. Automation in Intensity Modulated Radiotherapy Treatment Planning—A Review of Recent Innovations10.1259/bjr.20180270 ↗
23. The Current and Future Role of Automation in Radiotherapy Treatment Planning10.1088/1742-6596/1515/2/022025 ↗
24. Mancosu, P., Lambri, N., Castiglioni, I., Dei, D., Iori, M., Loiacono, D., Russo, S., Talamonti, C., Villaggi, E., and Scorsetti, M. (2022). Applications of Artificial Intelligence in Stereotactic Body Radiation Therapy. Phys. Med. Biol., 67.10.1088/1361-6560/ac7e18 ↗
25. Franzese, C., Dei, D., Lambri, N., Teriaca, M.A., Badalamenti, M., Crespi, L., Tomatis, S., Loiacono, D., Mancosu, P., and Scorsetti, M. (2023). Enhancing Radiotherapy Workflow for Head and Neck Cancer with Artificial Intelligence: A Systematic Review. JPM, 13.10.3390/jpm13060946 ↗
26. Taciuc, I.-A., Dumitru, M., Vrinceanu, D., Gherghe, M., Manole, F., Marinescu, A., Serboiu, C., Neagos, A., and Costache, A. (2024). Applications and Challenges of Neural Networks in Otolaryngology (Review). Biomed. Rep., 20.10.3892/br.2024.1781 ↗
27. Prospective Clinical Validation of Virtual Patient-Specific Quality Assurance of Volumetric Modulated Arc Therapy Radiation Therapy Plans10.1016/j.ijrobp.2022.04.040 ↗
28. The Sensitivity of Patient-Specific IMRT QA Methods in Detecting Systematic Errors: Field-by-Field versus Single-Gantry-Angle Composite10.1088/1742-6596/1248/1/012063 ↗
29. Parallel/Opposed: IMRT QA Using Treatment Log Files Is Superior to Conventional Measurement-based Method10.1120/jacmp.v16i1.5385 ↗
30. What Is Plan Quality in Radiotherapy? The Importance of Evaluating Dose Metrics, Complexity, and Robustness of Treatment Plans10.1016/j.radonc.2020.09.038 ↗