journal article·2022-06-17Open access

Novel Dual-Threaded Pedicle Screws Provide Fixation Stability That Is Comparable to That of Traditional Screws with Relative Bone Preservation: An In Vitro Biomechanical Study

Ming-Kai Hsieh, Yun-Da Li, Yun-Ja Hsu, Tsung-Ting Tsai, Po-Liang Lai, De-Mei Lee, Ching-Lung Tai

Applied Sciences· Volume 12· Issue 12

Abstract

Replacement with larger diameter screws is always used in pedicle screw loosening but carries a risk of pedicle wall violation. A pedicle screw with more preserved bone stock is the preferred primary fixation choice. The purpose of this study was to evaluate whether a newly designed proximal-conical dual-thread screw with less bone occupancy provides fixation strength comparable to that of a traditional screw. Six types of pedicle screws based on three different shapes (cylindrical, conical, and proximal-conical) and two thread profiles (single-thread and dual-thread) were grouped. Conical and proximal-conical screws differed mainly in the slope of the outer diameter from the hub to the tip. Conical screws had an outer diameter (6.5 mm) that differed from the hub and tapered by 30% to an outer diameter (4.5 mm) at the tip and proximal-conical screws had the same outer diameter from the hub and tapered by 30% (4.5 mm) at 20 mm from the hub and then maintained the outer diameter (45 mm) to the tip. A total of 36 L4 Sawbones® vertebrae were used in the study and six trials for each screw group. The results of the imaging, screw volume in bone, insertion torque, and pullout force were analyzed. For screws with the same shape, insertion torque and pullout force were significantly higher for those in the dual-thread groups than for those in the single-thread groups (p < 0.05). For screws with the same thread profile, there was no significant difference in either biomechanical test between the different screw shapes (p > 0.05). Our results demonstrated that these proximal-conical dual-thread screws, with the property of relative bone stock preservation, display a comparable biomechanical performance to traditional dual-thread screws and a better performance than single-thread screws. This screw design could serve as the primary pedicle screw choice to reduce revision difficulty.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Spinal Fractures and Fixation Techniques · Pelvic and Acetabular Injuries · Spine and Intervertebral Disc Pathology

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

Department of Orthopedic Surgery, Spine Section, Bone and Joint Research Center, Chang Gung Memorial Hospital, Taoyuan 333, Taiwan

Department of Biomedical Engineering, Chang Gung University, Taoyuan 333, Taiwan

Department of Mechanical Engineering, Chang Gung University, Taoyuan 333, Taiwan

Funding

Chang Gung Memorial Hospital

References (38)

Show source-provided references
1. Additional vertebral augmentation with posterior instrumentation for unstable thoracolumbar burst fractures10.1016/j.injury.2017.06.015 ↗
2. Percutaneous Balloon Kyphoplasty and Short Instrumentation Compared with Traditional Long Instrumentation for Thoracolumbar Metastatic Spinal Cord Compression10.1016/j.wneu.2019.06.182 ↗
3. Is open surgery for metastatic spinal cord compression secondary to lung cancer really beneficial? A systematic review10.1016/j.wneu.2020.08.098 ↗
4. Liao, J.C., Chen, W.P., and Wang, H. (2017). Treatment of thoracolumbar burst fractures by short-segment pedicle screw fixation using a combination of two additional pedicle screws and vertebroplasty at the level of the fracture: A finite element analysis. BMC Musculoskelet. Disord., 18.10.1186/s12891-017-1623-0 ↗
5. Su, K.C., Chen, K.H., Pan, C.C., and Lee, C.H. (2021). Biomechanical evaluation of cortical bone trajectoryfixation with traditional pedicle screw in the lumbar spine: A finite element study. Appl. Sci., 11.10.3390/app112210583 ↗
6. Hsieh, M.K., Chen, L.H., Niu, C.C., Fu, T.S., Lai, P.L., and Chen, W.J. (2015). Combined anterior lumbar interbody fusion and instrumented posterolateral fusion for degenerative lumbar scoliosis: Indication and surgical outcomes. BMC Surg., 15.10.1186/s12893-015-0006-4 ↗
7. Tsuang, F.Y., Chen, C.H., Wu, L.C., Kuo, Y.J., Hsieh, Y.Y., and Chiang, C.J. (2021). Partial threading of pedicle screws in a standard construct increases fatigue life: A biomechanical analysis. Appl. Sci., 11.10.3390/app11041503 ↗
8. Intraoperative image guidance compared with free-hand methods in adolescent idiopathic scoliosis posterior spinal surgery: A systematic review on screw-related complications and breach rates10.1016/j.spinee.2017.04.001 ↗
9. Pedicle screw loosening: A clinically relevant complication?10.1007/s00586-015-3768-6 ↗
10. Improved fixation stability for repairing pedicle screw loosening using a modified cement filling technique in porcine vertebrae10.1038/s41598-022-06724-4 ↗
11. Xu, F., Zhou, S., Zou, D., Li, W., Sun, Z., and Jiang, S. (2022). The relationship between S1 screw loosening and postoperative outcome in patients with degenerative lumbar scoliosis. BMC Musculoskelet. Disord., 23.10.1186/s12891-022-05107-0 ↗
12. Dual pitch screw design provides equivalent fixation to upsized screw diameter in revision pedicle screw instrumentation: A cadaveric biomechanical study10.1016/j.spinee.2021.07.010 ↗
13. Use of longer sized screws is a salvage method for broken pedicles in osteoporotic vertebrae10.1038/s41598-020-67489-2 ↗
14. Biomechanical comparison of expansive pedicle screw and polymethylmethacrylate-augmented pedicle screw in osteoporotic synthetic bone in primary implantation: An experimental study10.1097/BSD.0b013e31828bfc85 ↗
15. Hsieh, M.K., Liu, M.Y., Chen, J.K., Tsai, T.T., Lai, P.L., Niu, C.C., and Tai, C.L. (2019). Biomechanical study of the fixation stability of broken pedicle screws and subsequent strategies. PLoS ONE, 14.10.1371/journal.pone.0219189 ↗
16. Intraoperative revision of initially loosened pedicle screws10.1007/s00064-019-0611-y ↗
17. The Mechanisms of Medial Pedicle Wall Violation: Insertion Method Is as Important as Correct Cannulation of the Pedicle10.1155/2014/283783 ↗
18. The pullout strength of pedicle screws following redirection after lateral wall breach or end-plate breach10.1097/BRS.0000000000001600 ↗
19. Irie, T.Y., Irie, T., Espinoza Orías, A.A., Segami, K., Iwasaki, N., An, H.S., and Inoue, N. (2021). Micro-computed tomography analysis of the lumbar pedicle wall. PLoS ONE, 16.10.1371/journal.pone.0253019 ↗
20. Pullout strength of pedicle screws following redirection after lateral or medial wall breach10.1097/BRS.0000000000002611 ↗
21. Influence of novel design alteration of pedicle screw on pull-out strength: A finite element study10.1016/j.jos.2019.03.002 ↗
22. Biomechanical analysis of pedicle screw thread differential design in an osteoporotic cadaver model10.1016/j.clinbiomech.2011.10.004 ↗
23. Characteristics of immediate and fatigue strength of a dual-threaded pedicle screw in cadaveric spines10.1016/j.spinee.2013.03.010 ↗
24. Liu, M.Y., Tsai, T.T., Lai, P.L., Hsieh, M.K., Chen, L.H., and Tai, C.L. (2020). Biomechanical comparison of pedicle screw fixation strength in synthetic bones: Effects of screw shape, core/thread profile and cement augmentation. PLoS ONE, 15.10.1371/journal.pone.0229328 ↗
25. Cortical threaded pedicle screw improves fatigue strength in decreased bone quality10.1007/s00586-020-06593-3 ↗
26. Pedicle Screws Loosening in Patients With Degenerative Diseases of the Lumbar Spine: Potential Risk Factors and Relative Contribution10.1177/2192568218772302 ↗
27. Tai, C.L., Tsai, T.T., Lai, P.L., Chen, Y.L., Liu, M.Y., and Chen, L.H. (2015). A Biomechanical Comparison of Expansive Pedicle Screws for Severe Osteoporosis: The Effects of Screw Design and Cement Augmentation. PLoS ONE, 10.10.1371/journal.pone.0146294 ↗
28. National inpatient complications and outcomes after surgery for spinal metastasis from 1993–200210.1002/cncr.22819 ↗
29. Adverse events in emergency oncological spine surgery: A prospective analysis10.3171/2014.7.SPINE131007 ↗
30. Complications After Percutaneous Pedicle Screw Fixation for the Treatment of Unstable Spinal Metastases10.1245/s10434-016-5156-9 ↗
31. Formiga, M.C., Gehrke, A.F., DeBortoli, J.P., and Gehrke, S.A. (2021). Can the design of the instruments used for undersized osteotomies influence the initial stability of implants installed in low-density bone? An in vitro pilot study. PLoS ONE, 16.10.1371/journal.pone.0257985 ↗
32. Characterization of commercial rigid polyurethane foams used as bone analogs for implant testing10.1007/s10856-010-4024-6 ↗
33. Characteristics of pullout failure in conical and cylindrical pedicle screws after full insertion and back-out10.1016/S1529-9430(01)00119-X ↗
34. Influence of thread design on pedicle screw fixation. Laboratory investigation10.3171/SPI/2008/9/7/090 ↗
35. Screw insertion torque as parameter to judge the fixation. Assessment of torque and pull-out strength in different bone densities and screw-pitches10.1016/j.clinbiomech.2019.12.004 ↗
36. A comparison of screw insertion torque and pullout strength10.1097/BOT.0b013e3181c4a655 ↗
37. A comparison of pullout strength for pedicle screws of different designs: A study using tapped and untapped pilot holes10.1097/01.brs.0000244658.35865.59 ↗
38. Hsieh, M.K., Li, Y.D., Liu, M.Y., Lin, C.X., Tsai, T.T., Lai, P.L., and Tai, C.L. (2021). Biomechanical Comparison of Fixation Stability among Various Pedicle Screw Geometries: Effects of Screw Outer/Inner Projection Shape and Thread Profile. Appl. Sci., 11.10.3390/app11219901 ↗