Estimating Specific Patient Organ Dose for Chest CT Examinations with Monte Carlo Method
Source abstract
Abstract
Purpose: The purpose of this study was to preliminarily estimate patient-specific organ doses in chest CT examinations for Chinese adults, and to investigate the effect of patient size on organ doses. Methods: By considering the body-size and body-build effects on the organ doses and taking the mid-chest water equivalent diameter (WED) as a body-size indicator, the chest scan images of 18 Chinese adults were acquired on a multi-detector CT to generate the regional voxel models. For each patient, the lungs, heart, and breasts (glandular breast tissues for both breasts) were segmented, and other organs were semi-automated segmented based on their HU values. The CT scanner and patient models simulated by MCNPX 2.4.0 software (Los Alamos National LaboratoryLos Alamos, USA) were used to calculate lung, breast, and heart doses. CTDIvol values were used to normalize simulated organ doses, and the exponential estimation model between the normalized organ dose and WED was investigated. Results: Among the 18 patients in this study, the simulated doses of lung, heart, and breast were 18.15 ± 2.69 mGy, 18.68 ± 2.87 mGy, and 16.11 ± 3.08 mGy, respectively. Larger patients received higher organ doses than smaller ones due to the higher tube current used. The ratios of lung, heart, and breast doses to the CTDIvol were 1.48 ± 0.22, 1.54 ± 0.20, and 1.41 ± 0.13, respectively. The normalized organ doses of all the three organs decreased with the increase in WED, and the normalized doses decreased more obviously in the lung and the heart than that in the breasts. Conclusions: The output of CT scanner under ATCM is positively related to the attenuation of patients, larger-size patients receive higher organ doses. The organ dose normalized by CTDIvol was negatively correlated with patient size. The organ doses could be estimated by using the indicated CTDIvol combined with the estimated WED.
Source: crossref · Quality: baseline
Research context
OpenAlex topics: Radiation Dose and Imaging · Advanced X-ray and CT Imaging · Digital Radiography and Breast Imaging
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
Affiliations
Institute of Radiation Medicine, Fudan University, Shanghai 200032, China
Division of Nuclear Medicine and Molecular Imaging, Geneva University Hospital, CH-1211 Geneva, Switzerland
Funding
National Natural Science Foundation of China · National key R&D Program of China
References (32)
Show source-provided references
1. International Commission on Radiological Protection (2007). Annals of the ICRP-Publication 102-Managing Patient Dose in Multi-Detector Computed Tomography (MDCT), ELSEVIER.
2. Potential for adult-based epidemiological studies to characterize overall cancer risks associated with a lifetime of CT scans10.1667/RR13622.1 ↗
3. What we know and what we don’t know about cancer risks associated with radiation doses from radiological imaging10.1259/bjr.20130629 ↗
4. United Nations Scientific Committee on the Effects of Atomic Radiation (2020). UNSCEAR 2019 Report: Sources and Effects of Ionizing Radiation. Report to the General Assembly with Scientific Annexes, United Nations.
5. Computed tomography—An increasing source of radiation exposure10.1056/NEJMra072149 ↗
6. CT dose index and patient dose: They are not the same thing10.1148/radiol.11101800 ↗
7. American Association of Physicists in Medicine (2011). Size-Specific Dose Estimates (SSDE) in Pediatric and Adult Body CT, American Association of Physicists in Medicine. Report of AAPM task group 204.
8. International Electrotechnical Commission (2019). Methods for Calculating Size Specific Dose Estimates (SSDE) on Computed Tomography, International Electrotechnical Commission. IEC 62985.
9. International Commission on Radiological Protection (2009). Annals of the ICRP-Publication 110-Adult Reference Computational Phantoms, ELSEVIER.
10. VirtualDose: A software for reporting organ doses from CT for adult and pediatric patients10.1088/0031-9155/60/14/5601 ↗
11. NCICT: A computational solution to estimate organ doses for pediatric and adult patients undergoing CT scans10.1088/0952-4746/35/4/891 ↗
12. DukeSim: A realistic, rapid, and scanner-specific simulation framework in computed tomography10.1109/TMI.2018.2886530 ↗
13. Dose to radiosensitive organs during routine chest CT: Effects of tube current modulation10.2214/AJR.09.2886 ↗
14. The feasibility of a regional CTDIvol to estimate organ dose from tube current modulated CT exams10.1118/1.4798561 ↗
15. Pediatric chest and abdominopelvic CT: Organ dose estimation based on 42 patient models10.1148/radiol.13122617 ↗
16. Attenuation-based size metric for estimating organ dose to patients undergoing tube current modulated CT exams10.1118/1.4906132 ↗
17. Estimating organ doses from tube current modulated CT examinations using a generalized linear model10.1002/mp.12119 ↗
18. Assessment of organ doses for CT patients based on x-ray attenuation using water equivalent diameter10.1016/j.radphyschem.2020.109332 ↗
19. Estimating patient dose from CT exams that use automatic exposure control: Development and validation of methods to accurately estimate tube current values10.1002/mp.12314 ↗
20. American Association of Physicists in Medicine (2014). Use of Water Equivalent Diameter for Calculating Patient Size and Size-Specific Dose Estimates (SSDE) in CT, American Association of Physicists in Medicine. Report of AAPM task group 220.
21. International Commission on Radiological Protection (2003). Annals of the ICRP-Publication 89-Basic Anatomical and Physiological Data for Use in Radiological Protection, International Commission on Radiological Protection.
22. National Standard of the People’s Republic of China (NSPRC) (2007). Reference Individuals for Use in Radiation Protection: Part 1 Physique Parameters, Chinese Standard GBZ/T200.1.
23. National Standard of the People’s Republic of China (NSPRC) (2007). Reference Individuals for Use in Radiation Protection: Part 2 Masses of Main Organs and Tissues, Chinese Standard GBZ/T200.2.
24. Techniques and applications of automatic tube current modulation for CT10.1148/radiol.2333031150 ↗
25. Investigation and analysis of the application frequency of medical X-CT and the dose in Fujian Province in 2016
26. International variation in radiation dose for computed tomography examinations: Prospective cohort study
27. Huang, Y. (2018). Study on the Method for Estimating Organ Dose from X-CT Scans with Automatic Tube Current Modulation. [Master’s Thesis, Radiation Medicine, Fudan University].
28. A new phantom developed to test the ATCM performance of chest CT scanners10.1088/1361-6498/abf900 ↗
29. Technical Note:Spektr 3.0 A computational tool for X-Ray spectrum modeling and analysis10.1118/1.4955438 ↗
30. The feasibility of a scanner-independent technique to estimate organ dose from MDCT scans: Using CTDIvol to account for differences between scanners10.1118/1.3368596 ↗
31. The study and prediction of the total number of radio-diagnosis and radiotherapy examination in hospitals
32. Survey on frequency of medical X-ray diagnosis in Shanghai