Real-Time HIFU Treatment Monitoring Using Pulse Inversion Ultrasonic Imaging
Source abstract
Abstract
Real-time monitoring of high-intensity focused ultrasound (HIFU) surgery is essential for safe and accurate treatment. However, ultrasound imaging is difficult to use for treatment monitoring during HIFU surgery because of the high intensity of the HIFU echoes that are received by an imaging transducer. Here, we propose a real-time HIFU treatment monitoring method based on pulse inversion of imaging ultrasound; an imaging transducer fires ultrasound twice in 0° and 180° phases for one scanline while HIFUs of the same phase are transmitted in synchronization with the ultrasound transmission for imaging. By doing so, HIFU interferences can be eliminated after subtracting the two sets of the signals received by the imaging transducer. This function was implemented in a commercial research ultrasound scanner, and its performance was evaluated using the excised bovine liver. The experimental results demonstrated that the proposed method allowed ultrasound images to clearly show the echogenicity change induced by HIFU in the excised bovine liver. Additionally, it was confirmed that the moving velocity of the organs in the abdomen due to respiration does not affect the performance of the proposed method. Based on the experimental results, we believe that the proposed method can be used for real-time HIFU surgery monitoring that is a pivotal function for maximized treatment efficacy.
Source: crossref · Quality: baseline
Research context
OpenAlex topics: Ultrasound and Hyperthermia Applications · Ultrasound Imaging and Elastography · Photoacoustic and Ultrasonic Imaging
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
Affiliations
Department of Electronic Engineering, Sogang University, Seoul 04107, Korea
Department of Biomedical Engineering, Sogang University, Seoul 04107, Korea
Funding
National Research Foundation of Korea
References (29)
Show source-provided references
1. Scanning ultrasound removes amyloid-β and restores memory in an Alzheimer’s disease mouse model10.1126/scitranslmed.aaa2512 ↗
2. Focused ultrasound as a novel strategy for Alzheimer’s disease therapeutics10.1002/ana.24933 ↗
3. Microfocused ultrasound for skin tightening
4. Jang, J., and Chang, J.H. (2016). Design and fabrication of double-focused ultrasound transducers to achieve tight focusing. Sensors, 16.10.3390/s16081248 ↗
5. Temperature elevation by HIFU in ex vivo porcine muscle: MRI measurement and simulation study10.1118/1.4870965 ↗
6. Real-time 2-D temperature imaging using ultrasound10.1109/TBME.2009.2035103 ↗
7. Real-time visualization of high-intensity focused ultrasound treatment using ultrasound imaging10.1016/S0301-5629(00)00279-9 ↗
8. Using the acoustic interference pattern to locate the focus of a high-intensity focused ultrasound (HIFU) transducer10.1016/j.ultrasmedbio.2007.07.001 ↗
9. MR techniques for guiding high-intensity focused ultrasound (HIFU) treatments10.1002/jmri.25770 ↗
10. Compensating for bone interfaces and respiratory motion in high-intensity focused ultrasound10.1080/02656730701209996 ↗
11. Real-time passive acoustic monitoring of HIFU-induced tissue damage10.1016/j.ultrasmedbio.2011.02.016 ↗
12. A method to synchronize high-intensity, focused ultrasound with an arbitrary ultrasound imager10.1109/TUFFC.2006.1610574 ↗
13. New integrated imaging high intensity focused ultrasound probe for transrectal prostate cancer treatment10.1016/j.ultrasmedbio.2007.12.005 ↗
14. Ultrasound transducer and system for real-time simultaneous therapy and diagnosis for noninvasive surgery of prostate tissue10.1109/TUFFC.2009.1267 ↗
15. Pulse compression technique for simultaneous HIFU surgery and ultrasonic imaging: A preliminary study10.1016/j.ultras.2012.01.016 ↗
16. Real-time monitoring of HIFU treatment using pulse inversion10.1088/0031-9155/58/15/5333 ↗
17. An effective pulse sequence for simultaneous HIFU insonation and monitoring10.1109/TUFFC.2014.3072 ↗
18. Improvement of tissue harmonic imaging using the pulse-inversion technique10.1016/j.ultrasmedbio.2005.03.006 ↗
19. Pulse inversion techniques in ultrasonic nonlinear imaging10.1016/S0929-6441(09)60073-4 ↗
20. Thermal therapeutic method for selective treatment of deep-lying tissue by combining laser and high-intensity focused ultrasound energy10.1364/OL.39.002806 ↗
21. Low-cost, high-speed back-end processing system for high-frequency ultrasound B-mode imaging10.1109/TUFFC.2009.1205 ↗
22. Increased light penetration due to ultrasound-induced air bubbles in optical scattering media10.1038/s41598-017-16444-9 ↗
23. Acoustic impedance evaluation of high-intensity-focused-ultrasound exposed chicken breast muscle using ultrasonic microscopy10.1143/JJAP.49.07HF04 ↗
24. Ultrasound-assisted photothermal therapy and real-time treatment monitoring10.1364/BOE.9.004472 ↗
25. Real-time spatiotemporal control of high-intensity focused ultrasound thermal ablation using echo decorrelation imaging in ex vivo bovine liver10.1016/j.ultrasmedbio.2017.09.007 ↗
26. Optimized echo decorrelation imaging feedback for bulk ultrasound ablation control10.1109/TUFFC.2018.2847599 ↗
27. Motion artifacts of pulse inversion-based tissue harmonic imaging10.1109/TUFFC.2002.1041536 ↗
28. An efficient motion estimation and compensation method for ultrasound synthetic aperture imaging10.1177/016173460202400202 ↗
29. Ultrasound quantitation of respiratory organ motion in the upper abdomen10.1259/0007-1285-67-803-1096 ↗