journal article·2022-12-20Open access

Computer Simulations of Dual-Antenna Microwave Ablation and Comparison to Experimental Measurements

Jinying Wang, Shengyang Huang, Hongjian Gao, Ju Liu, Yubo Zhang, Shuicai Wu

Applied Sciences· Volume 13· Issue 1

Abstract

Single-antenna microwave ablation (MWA) is mainly used to treat small tumors less than 3 cm in diameter. To obtain a larger coagulation zone in a single ablation, a dual-antenna ablation approach was proposed. A three-dimensional finite element method (FEM) simulation model of parallel dual-antennas was developed. Ex vivo experiments at 50 W for 8 min were performed to verify the model. Both the temperature changes in tissue and the size of the coagulation zone were recorded. The effects of dual-antenna spacing, heating power, and blood perfusion on the coagulation zone were analyzed. Fifteen experiments were carried out. The errors between the mean measurements and simulated results at the set temperature points were 1.08 °C, 0.95 °C, and 2.1 °C, respectively. For the same conditions, the blood perfusion of 1.0, 1.5, and 3.0 kg/(m3·s) can result in a reduction in the coagulation volume by 18.4%, 25.4%, and 42.5%. As the spacing increased, the coagulation zone of each antenna started to fuse together later and the resulting integral coagulation zone became larger. Dual-antenna MWA is expected to be used for the treatment of tumors larger than 5 cm in diameter.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Ultrasound and Hyperthermia Applications · Medical Imaging and Pathology Studies · Ultrasound Imaging and Elastography

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

Faculty of Environmental and Life Sciences, Beijing University of Technology, Beijing 100124, China

Funding

National Natural Science Foundation of China

References (46)

Show source-provided references
1. Cancer statistics10.3322/caac.21708 ↗
2. Global cancer statistics 2020: GLOBOCAN estimates of incidence and mortality worldwide for 36 cancers in 185 countries10.3322/caac.21660 ↗
3. Innovative approaches for cancer treatment: Current perspectives and new challenges10.3332/ecancer.2019.961 ↗
4. Real-time in vivo assessment of radiofrequency ablation of human colorectal liver metastases using diffuse reflectance spectroscopy10.1016/j.ejso.2015.12.005 ↗
5. Hepatic resection for malignant liver tumors in the elderly: A systematic review and meta-analysis10.1111/ans.13211 ↗
6. Radiofrequency Ablation and Microwave Ablation in Liver Tumors: An Update10.1634/theoncologist.2018-0337 ↗
7. Microwave ablation Technology: What every User should know10.1067/j.cpradiol.2007.08.011 ↗
8. Evaluation of a New CT-Guided Robotic System for Percutaneous Antenna Insertion for Thermal Ablation of Liver Tumors: A Prospective Pilot Study10.1007/s00270-022-03267-z ↗
9. Role of Contrast-Enhanced Ultrasound in the Detection of Complications After Ultrasound-Guided Liver Interventional Procedures
10. Hepatic microwave ablation with multiple antennae results in synergistically larger zones of coagulation necrosis10.1245/ASO.2003.03.045 ↗
11. Prognostic Factors in Hepatitis C Patients with a Single Small Hepatocellular Carcinoma after Radiofrequency Ablation
12. Finite-Element Analysis and In Vitro Experiments of Placement Configurations Using Triple Antennas in Microwave Hepatic Ablation10.1109/TBME.2009.2027128 ↗
13. Clibbon, K.L., and McCowen, A. (1993, January 14–18). Thermal modeling of non-ideal interstitial microwave antenna array hyperthermia for the treatment of cancer. Proceedings of the IEEE MTT-S International Microwave Symposium Digest, Atlanta, GA, USA.
14. Hepatocellular carcinoma: MWA with multiple straight and loop antenna clusters-Pilot comparison with pathologic findings10.1148/radiol.2383041592 ↗
15. Interstitial microwave transition from hyperthermia to ablation: Historical perspectives and current trends in thermal therapy10.3109/02656731003639356 ↗
16. The Effects of Driving Frequency and Antenna Length on Power Deposition Within a Microwave Antenna Array Used for Hyperthermia10.1109/TBME.1985.325436 ↗
17. Mukherjee, S., Curto, S., Albin, N., Natarajan, B., and Prakash, P. (2015, January 8–9). Multiple-antenna MWA: Analysis of non-parallel antenna implants. Proceedings of the PIE Conference on Energy-Based Treatment of Tissue and Assessment VIII, San Francisco, CA, USA.10.1117/12.2080349 ↗
18. Aparna, V.C., Gandhi, A.S., and Harsh, R. (2022, January 24–27). Minimally Invasive MWA Antenna Designs at 915 MHz and 2.45 GHz. Proceedings of the 28th National Conference on Communications (NCC), Mumbai, India.10.1109/NCC55593.2022.9806765 ↗
19. Study on interstitial microwave hyperthermia with multi-slot coaxial antenna
20. The vascular cooling effect in hepatic multipolar radiofrequency ablation leads to incomplete ablation ex vivo10.1080/02656736.2016.1196395 ↗
21. Microwave ablation: Results with double 915 MHz antennae in ex vivo bovine livers10.1016/j.ejrad.2010.03.015 ↗
22. Treatment planning facilitates clinical decision making for hyperthermia treatments10.1080/02656736.2021.1903583 ↗
23. Tissue shrinkage in microwave ablation of liver: An ex vivo predictive model10.1080/02656736.2016.1208292 ↗
24. Influences of blood flow parameters on temperature distribution during liver tumor microwave ablation10.52586/4963 ↗
25. Gas, P., and Kurgan, E. (2016, January 14–17). Comparative Analysis Between the 2D and 3D Models of Interstitial Microwave Hyperthermia. Proceedings of the 2016 17th International Conference Computational Problems of Electrical Engineering (CPEE), Sandomierz, Poland. IEEE Xplore.10.1109/CPEE.2016.7738757 ↗
26. Computational modeling of microwave tumor ablations10.3109/02656736.2013.799295 ↗
27. Recent developments in modeling heat transfer in blood-perfused tissues10.1109/10.284920 ↗
28. 2450-MHz microwave ablation temperature simulation using temperature-dependence feedback of characteristic parameters10.1002/mmce.21488 ↗
29. Numerical models to evaluate the temperature increase induced by ex vivo microwave thermal ablation10.1088/0031-9155/60/8/3287 ↗
30. Low-Cost Mini-Invasive Microwave Needle Applicator for Cancer Thermal Ablation: Feasibility Investigation10.1109/JSEN.2021.3060499 ↗
31. Effects of saline volume on lesion formation during saline-infused radiofrequency ablation10.1016/j.apm.2016.11.032 ↗
32. Thermal therapy in urologic systems: A comparison of Arrhenius and thermal isoeffective dose models in predicting hyperthermic injury10.1115/1.3128671 ↗
33. Gas, P., and Kurgan, E. (2018, January 18–22). Evaluation of thermal damage of hepatic tissue during thermotherapy based on the arrhenius model. Proceedings of the 2018 Progress in Applied Electrical Engineering (PAEE), Koscielisko, Poland.10.1109/PAEE.2018.8441065 ↗
34. Review of temperature dependence of thermal properties, dielectric properties, and perfusion of biological tissues at hyperthermic and ablation temperatures10.1615/CritRevBiomedEng.2015012486 ↗
35. Thermal analysis of induced damage to the healthy cell during RFA of breast tumor10.1016/j.jtherbio.2016.04.002 ↗
36. Comparison of temperature curve and ablation zone between 915- and 2450-MHz cooled-shaft microwave antenna: Results in ex vivo porcine livers10.1016/j.ejrad.2011.02.013 ↗
37. Hyperthermia-induced changes in liver physiology and metabolism: A rationale for hyperthermic machine perfusion10.1152/ajpgi.00101.2020 ↗
38. A comparative analysis of thermal blood perfusion measurement techniques10.1115/1.3138672 ↗
39. Computer-aided dynamic simulation of microwave-induced thermal distribution in coagulation of liver cancer10.1109/10.930907 ↗
40. Thermal ablation of biological tissues in disease treatment: A review of computational models and future directions10.1080/15368378.2020.1741383 ↗
41. Errors between two-dimensional and three-dimensional thermal model predictions of hyperthermia treatments10.3109/02656739009140814 ↗
42. Interactive multi-criteria planning for radiofrequency ablation10.1007/s11548-015-1201-6 ↗
43. Heating characteristics of antenna arrays used in microwave ablation: A theoretical parametric study10.1016/j.compbiomed.2013.07.013 ↗
44. A Dual-Slot Microwave Antenna for More Spherical Ablation Zones: Ex Vivo and in vivo Validation10.1148/radiol.13122128 ↗
45. Simulation study of the cooling effect of blood vessels and blood coagulation in hepatic radio-frequency ablation10.1080/02656736.2020.1866217 ↗
46. Vein wall shrinkage induced by thermal coagulation with high-intensity-focused ultrasound: Numerical modeling and in vivo experiments in sheep10.1080/02656736.2020.1834626 ↗