journal article·2022-05-27Open access

Descriptive Time Series Analysis for Downtime Prediction Using the Maintenance Data of a Medical Linear Accelerator

Kwang Hyeon Kim, Moon-Jun Sohn, Suk Lee, Hae-Won Koo, Sang-Won Yoon, Ahmad Khalid Madadi

Applied Sciences· Volume 12· Issue 11

Abstract

A medical linear accelerator (LINAC) delivers high-energy X-rays or electrons to the patient’s tumor. In this study, we categorized failures and predicted downtime leading to discontinuous radiation treatment using a descriptive time series analysis of a 20-year maintenance dataset of a medical LINAC. A LINAC dataset of failure records for 359 instances was collected from 2001 to 2021. Next, we performed institution-specific seasonal autoregressive integrated moving average (ARIMA) modeling to analyze the causes of the failure categories and predict the downtime. Furthermore, we evaluated the performance of the predictive model using standard error metrics and statistical methods. Our results show that the downtime will increase by 95 h/year after 2022 and 100 h/year after 2023. The accumulated downtime in 2029 is predicted to be a maximum of 2820 h. The modeled seasonal ARIMA showed statistical significance (p < 0.001) with a residual error of σ2 (328.33 ± 9.4). In addition, the forecasting performance of the model was assessed using the mean absolute percentage error (MAPE). The failure parts where the major downtime occurred were the multileaf collimator (25.2%), gantry and couch motion part (15.4%), dosimetric part (11.7%), and computer console (10.0%). Using the development of the ARIMA model specific to our institution, the downtime is predicted to reach up to 2820 h.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Advanced Radiotherapy Techniques · Radiation Therapy and Dosimetry · Advanced X-ray and CT Imaging

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

Department of Neurosurgery, College of Medicine, Inje University Ilsan Paik Hospital, Goyang 10380, Korea

Department of Radiation Oncology, Anam Hospital, College of Medicine, Korea University, Seoul 02841, Korea

Funding

the Basic Science Research Program through the National Research Foundation of Korea (NRF) funded by the Ministry of Education

References (32)

Show source-provided references
1. Back to the future: The history and development of the clinical linear accelerator10.1088/0031-9155/51/13/R20 ↗
2. Rapid acceptance testing of modern linac using on-board MV and kV imaging systems10.1002/mp.12294 ↗
3. Das, S., Krishnan, R., Bhagwat, A., and Pethe, S. (2009, January 28–30). Linac for Medical Applications. Proceedings of the 2009 IEEE International Vacuum Electronics Conference, Rome, Italy.10.1109/IVELEC.2009.5193402 ↗
4. Podgorsak, E.B. (2005). Treatment Machines for External Beam Radiotherapy. IAEA Radiation Oncology Physics: A Handbook for Teachers and Students, International Atomic Energy Agency.
5. Real-time tumor tracking using sequential kV imaging combined with respiratory monitoring: A general framework applicable to commonly used IGRT systems10.1088/0031-9155/55/12/003 ↗
6. Precision of image-guided radiotherapy (IGRT) in six degrees of freedom and limitations in clinical practice10.1007/s00066-007-1695-0 ↗
7. Image guided radiation therapy (IGRT) practice patterns and IGRT’s impact on workflow and treatment planning: Results from a national survey of American Society for Radiation Oncology members10.1016/j.ijrobp.2015.09.035 ↗
8. The report of Task Group 100 of the AAPM: Application of risk analysis methods to radiation therapy quality management10.1118/1.4947547 ↗
9. Task Group 142 report: Quality assurance of medical accelerators a10.1118/1.3190392 ↗
10. Quantifying risk using FMEA: An alternate approach to AAPM TG-100 for scoring failures and evaluating clinical workflow10.1016/j.brachy.2021.02.005 ↗
11. Predicting machine’s performance record using the stacked long short-term memory (LSTM) neural networks10.1002/acm2.13558 ↗
12. Topic modeling of maintenance logs for linac failure modes and trends identification10.1002/acm2.13477 ↗
13. Development of Novalis Quality Assurance Protocol for Radiosurgery
14. Spinal radiosurgical treatment for thoracic epidural cavernous hemangioma presenting as radiculomyelopathy: Technical case report10.1227/01.NEU.0000345940.21674.AE ↗
15. In-House Developed Surface-Guided Repositioning and Monitoring System to Complement In-Room Patient Positioning System for Spine Radiosurgery10.14316/pmp.2021.32.2.40 ↗
16. Fractionated Stereotactic Radiotherapy in Pediatric Diffuse Intrinsic Brain Stem Gliomas
17. An ARIMA-model-based approach to seasonal adjustment10.1080/01621459.1982.10477767 ↗
18. Time series analysis using autoregressive integrated moving average (ARIMA) models10.1111/j.1553-2712.1998.tb02493.x ↗
19. Schaffer, A.L., Dobbins, T.A., and Pearson, S.-A. (2021). Interrupted time series analysis using autoregressive integrated moving average (ARIMA) models: A guide for evaluating large-scale health interventions. BMC Med. Res. Methodol., 21.10.1186/s12874-021-01235-8 ↗
20. The Akaike information criterion: Background, derivation, properties, application, interpretation, and refinements10.1002/wics.1460 ↗
21. A primer on model selection using the Akaike Information Criterion
22. The coefficient of determination R-squared is more informative than SMAPE, MAE, MAPE, MSE and RMSE in regression analysis evaluation10.7717/peerj-cs.623 ↗
23. Predictive quality assurance of a linear accelerator based on the machine performance check application using statistical process control and ARIMA forecast modeling10.1002/acm2.12917 ↗
24. Multivariate log file analysis for multi-leaf collimator failure prediction in radiotherapy delivery10.1016/j.phro.2020.07.011 ↗
25. Donon, Y., Kupriyanov, A., Kirsh, D., Di Meglio, A., Paringer, R., Rytsarev, I., Serafimovich, P., and Syomic, S. (2020, January 26–29). Extended Anomaly Detection and Breakdown Prediction in LINAC 4’s RF Power Source Output. Proceedings of the 2020 International Conference on Information Technology and Nanotechnology (ITNT), Samara, Russia.10.1109/ITNT49337.2020.9253296 ↗
26. A model for preemptive maintenance of medical linear accelerators—Predictive maintenance10.1186/s13014-016-0602-1 ↗
27. Drought forecasting based on the remote sensing data using ARIMA models10.1016/j.mcm.2009.10.031 ↗
28. Jan, F., Shah, I., and Ali, S. (2022). Short-Term Electricity Prices Forecasting Using Functional Time Series Analysis. Energies, 15.10.3390/en15093423 ↗
29. Short-term forecasting for the electricity spot prices with extreme values treatment10.1109/ACCESS.2021.3100076 ↗
30. Combining seasonal ARIMA models with computational intelligence techniques for time series forecasting10.1007/s00500-012-0805-9 ↗
31. A comparison of the forecasting ability of ARIMA models10.1108/14635780710746902 ↗
32. Stamatis, D.H. (2003). Failure Mode and Effect Analysis: FMEA from Theory to Execution, Quality Press.