Assessing the Impact of Resin Type, Post-Processing Technique, and Arch Location on the Trueness and Precision of 3D-Printed Full-Arch Implant Surgical Guides
Source abstract
Abstract
Objective: The purpose of this study was to assess the impact of photopolymer resin type, arch location, and post-processing techniques on the trueness and precision of three-dimensionally printed (3DP) full-arch surgical implant guides. Methods: Stereolithography reference images of an upper and lower surgical guide with six drill holes from a full-mouth rehabilitation clinical case were used. The files were imported into the Asiga MAX UV slicing software (Asiga Composer) where build orientation, print resolution, and support structures were added. A digital light processing 3D printer (MAX UV, Asiga Max) was used for printing the samples. The samples assessed were printed using two different, manufacturer-validated resins, DentaGuide (n = 35) and DentaClear (n = 20). The samples were subdivided and measured based on the post-processing technique used: handwashing (n = 20), sonication (n = 25), a mix of handwashing and sonication (n = 10), and post-curing using 385 nm UVA light with nitrogen (n = 50) or without nitrogen (n = 5). The diameter of each drill hole per guide was measured using a coordinate measuring machine (Absolute Arm 7-Axis, Hexagon) and compared with the reference STL to calculate each sample’s trueness (median error) and precision (interquartile range). The Mann–Whitney and Kruskal–Wallis tests were used for statistical analyses. Results: All samples demonstrated a dimensional error of <70 µm. No significant differences (p > 0.05) were observed between upper and lower arches and between post-processing techniques using nitrogen, irrespective of the use of hand- or ultrasonic washing. In contrast, DentaClear resin was significantly (p < 0.001) more accurate with a trueness of 26 µm and precision of 12 to 34 µm versus the DentaGuide at −31 µm and −54 to −17 µm, respectively. The samples post-cured without nitrogen were significantly (p < 0.05) the least accurate of all surgical guides, with a trueness of −42 µm and precision of −68 to −39 µm. Conclusion: The resin type and nitrogen post-processing are parameters that can significantly impact the accuracy of surgical guides. The tolerance of 3DP surgical guides needs to account for the dimensional changes occurring during the manufacturing process to minimise implant positioning errors.
Source: crossref · Quality: baseline
Research context
OpenAlex topics: Dental Implant Techniques and Outcomes · Dental materials and restorations · Anatomy and Medical Technology
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
Affiliations
School of Medicine and Dentistry, Griffith University, Gold Coast, QLD 4215, Australia
Advanced Design and Prototyping Technologies Institute, Griffith University, Gold Coast, QLD 4215, Australia
Glasgow Dental Hospital and School, Glasgow G2 3JZ, UK
Dentistry and Oral Health, Department of Rural Clinical Sciences, La Trobe Rural Health School, La Trobe University, Bendigo, VIC 3552, Australia
Violet Vines Marshman Centre for Rural Health Research, La Trobe Rural Health School, La Trobe University, Bendigo, VIC 3552, Australia
References (34)
Show source-provided references
1. Lin, L.H., Granatelli, J., Alifui-Segbaya, F., Drake, L., Smith, D., and Ahmed, K.E. (2021). A Proposed In Vitro Methodology for Assessing the Accuracy of Three-Dimensionally Printed Dental Models and the Impact of Storage on Dimensional Stability. Appl. Sci., 11.10.3390/app11135994 ↗
2. Production tolerance of additive manufactured polymeric objects for clinical applications10.1016/j.dental.2016.03.020 ↗
3. Anadioti, E., Musharbash, L., Blatz, M.B., Papavasiliou, G., and Kamposiora, P. (2020). 3D printed complete removable dental prostheses: A narrative review. BMC Oral Health, 20.10.1186/s12903-020-01328-8 ↗
4. Trueness and fitting accuracy of maxillary 3D printed complete dentures10.2186/jpr.JPR_D_20_00240 ↗
5. CAD-CAM complete denture resins: An evaluation of biocompatibility, mechanical properties, and surface characteristics10.1016/j.jdent.2021.103785 ↗
6. Effect of Manufacturing Technique on the Accuracy of Surgical Guides for Static Computer-Aided Implant Surgery10.11607/jomi.8186 ↗
7. Influences of build angle on the accuracy, printing time, and material consumption of additively manufactured surgical templates10.1016/j.prosdent.2020.09.012 ↗
8. Etemad-Shahidi, Y., Qallandar, O.B., Evenden, J., Alifui-Segbaya, F., and Ahmed, K.E. (2020). Accuracy of 3-Dimensionally Printed Full-Arch Dental Models: A Systematic Review. J. Clin. Med., 9.10.3390/jcm9103357 ↗
9. 3D printed versus conventionally cured provisional crown and bridge dental materials10.1016/j.dental.2017.10.003 ↗
10. The accuracy of static computer-aided implant surgery: A systematic review and meta-analysis10.1111/clr.13346 ↗
11. Accuracy of commercial 3D printers for the fabrication of surgical guides in dental implantology10.1016/j.jdent.2021.103909 ↗
12. Wegmüller, L., Halbeisen, F., Sharma, N., Kühl, S., and Thieringer, F.M. (2021). Consumer vs. High-End 3D Printers for Guided Implant Surgery—An In Vitro Accuracy Assessment Study of Different 3D Printing Technologies. J. Clin. Med., 10.10.3390/jcm10214894 ↗
13. Current state of the art of computer-guided implant surgery10.1111/prd.12175 ↗
14. Does residual cement around implant-supported restorations cause peri-implant disease? A retrospective case analysis10.1111/j.1600-0501.2012.02570.x ↗
15. A systematic review on the accuracy and the clinical outcome of computer-guided template-based implant dentistry10.1111/j.1600-0501.2009.01788.x ↗
16. Accuracy of computer-aided implant placement10.1111/j.1600-0501.2012.02552.x ↗
17. Establishing a point-of-care additive manufacturing workflow for clinical use10.1557/s43578-021-00270-x ↗
18. Effects of two Post-Processing Methods Onto Surface Dimension of in-Office Fabricated Stereolithographic Implant Surgical Guides10.1111/jopr.13227 ↗
19. Temporary 3D printed fixed dental prosthesis materials: Impact of post printing cleaning methods on degree of conversion as well as surface and mechanical properties10.11607/ijp.7048 ↗
20. Development and evaluation of an interdisciplinary teaching model via 3D printing10.1002/cre2.334 ↗
21. Wada, J., Wada, K., Gibreel, M., Wakabayashi, N., Iwamoto, T., Vallittu, P.K., and Lassila, L. (2022). Effect of Nitrogen Gas Post-Curing and Printer Type on the Mechanical Properties of 3D-Printed Hard Occlusal Splint Material. Polymers, 14.10.3390/polym14193971 ↗
22. An In Vitro Evaluation of the Effect of 3D Printing Orientation on the Accuracy of Implant Surgical Templates Fabricated By Desktop Printer10.1111/jopr.13485 ↗
23. Accuracy of 3D printed polymers intended for models and surgical guides printed with two different 3D printers10.4012/dmj.2020-039 ↗
24. Comparison of the fit of surgical guides manufactured with low-cost 3D printers
25. An assessment of template-guided implant surgery in terms of accuracy and related factors10.4047/jap.2013.5.4.440 ↗
26. Digital workflow: In vitro accuracy of 3D printed casts generated from complete-arch digital implant scans10.1016/j.prosdent.2019.10.029 ↗
27. Effect of additive manufacturing process and storage condition on the dimensional accuracy and stability of 3D-printed dental casts10.1016/j.prosdent.2021.02.028 ↗
28. On different 3D printing methods and fracture performance in DCB composite specimens including structured interfaces10.1016/j.tafmec.2022.103552 ↗
29. Characterization of 3D-printed PLA parts with different raster orientations and printing speeds10.1038/s41598-022-05005-4 ↗
30. Accuracy of digital light processing printing of 3-dimensional dental models10.1016/j.ajodo.2019.10.012 ↗
31. Influence of the three-dimensional printing technique and printing layer thickness on model accuracy10.1007/s00056-019-00180-y ↗
32. (2022, February 01). Available online: https://www.iso.org/standard/66262.html.
33. Pop, S.I., Dudescu, M., Mihali, S.G., Păcurar, M., and Bratu, D.C. (2022). Effects of Disinfection and Steam Sterilization on the Mechanical Properties of 3D SLA- and DLP-Printed Surgical Guides for Orthodontic Implant Placement. Polymers, 14.10.3390/polym14102107 ↗
34. Accuracy of surgical guides from 2 different desktop 3D printers for computed tomography-guided surgery10.1016/j.prosdent.2018.08.009 ↗