Bone Temperature Variation Using a 3D-Printed Surgical Guide with Internal Irrigation
Source abstract
Abstract
Bone overheating is a possible cause of implants early failure. When a surgical guide is used, the risk of heat injury is greater due to the reduced efficacy of the irrigation. The aim of this ex vivo study was to evaluate the effect of an additional built-in irrigation on bone temperature variation during implant osteotomy. Twelve bovine ribs were used. Cone beam computerized tomography (CBCT) was performed and a 3D-printed surgical guide with additional built-in irrigation tubes was produced for each rib. A total of 48 osteotomies were prepared, to compare the supplementary internal irrigation system (Group A) with external irrigation alone (Group B), no irrigation (Group C) and with free-hand surgery with external irrigation (Group D). Temperature was measured by three thermocouples placed at depths of 1.5, 7, and 12 mm. The largest temperature variation at each thermocouple showed median values of 3.0 °C, 1.9 °C, and 2.3 °C in Group 1; 2.3 °C, 1.7 °C, and 0.9 °C in Group 2; 3.2 °C, 1.6 °C, and 2.0 °C in Group 3; 2.0 °C, 2.0 °C, and 1.3 °C in Group 4, respectively. No differences were found among the four groups. In general, the highest temperature increase was observed with the use of the first drill (cortical perforator). Post-experimental CBCT revealed the presence of radiopaque material clogging the aperture of the internal irrigation channels. Additional internal irrigation was not found to significantly contribute to decrease bone temperature in this ex vivo setting.
Source: crossref · Quality: baseline
Research context
OpenAlex topics: Dental Implant Techniques and Outcomes · Orthopaedic implants and arthroplasty · Bone Tissue Engineering Materials
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
Affiliations
Section of Dentistry, Department of Neurosciences, University of Padova, Via Giustiniani 2, 35128 Padova, Italy
Department of Oral and Maxillofacial Surgery and Oral Medicine, Faculty of Odontology, Malmö University, Carl Gustavs Väg 34, 20506 Malmö, Sweden
Independent Statistician, 36020 Solagna, Italy
Department of Management and Engineering, University of Padova, Stradella San Nicola 3, 36100 Vicenza, Italy
Funding
Biomax S.p.A.
References (36)
Show source-provided references
1. Osseointegrated Titanium Implants. Requirements for Ensuring a Long-Lasting, Direct Bone-to-Implant Anchorage in Man10.3109/17453678108991776 ↗
2. Biological Factors Contributing to Failures of Osseointegrated Oral Implants. (I). Success Criteria and Epidemiology10.1046/j.0909-8836..t01-2-.x ↗
3. Wagenberg, B., and Froum, S.J. (2006). A Retrospective Study of 1925 Consecutively Placed Immediate Implants from 1988 to 2004. Int. J. Oral Maxillofac. Implant., 21.10.1016/j.prosdent.2006.05.016 ↗
4. Heat Generation During Implant Drilling: The Significance of Motor Speed10.1053/joms.2002.34992 ↗
5. Novel Hybrid Drilling Protocol: Evaluation for the Implant Healing–Thermal Changes, Crestal Bone Loss, and Bone-to-Implant Contact10.1111/clr.12341 ↗
6. Stocchero, M., Jinno, Y., Toia, M., Ahmad, M., Papia, E., Yamaguchi, S., and Becktor, J.P. (2019). Intraosseous Temperature Change during Installation of Dental Implants with Two Different Surfaces and Different Drilling Protocols: An in Vivo Study in Sheep. J. Clin. Med., 8.10.3390/jcm8081198 ↗
7. Thermal Injury to Bone: A Vital-Microscopic Description of Heat Effects10.1016/S0300-9785(82)80020-3 ↗
8. The Effect of Heat on Bone Regeneration: An Experimental Study in the Rabbit Using the Bone Growth Chamber10.1016/0278-2391(84)90417-8 ↗
9. Cortical Bone Drilling and Thermal Osteonecrosis10.1016/j.clinbiomech.2011.10.010 ↗
10. Influence of Heat Stress to Matrix on Bone Formation10.1111/j.1600-0501.2009.01654.x ↗
11. Ferhat, Mahmut Sumer, Murat Yenisey, and Erol Ergioglu. Effect of Surgical Drill Guide on Heat Generated from Implant Drilling10.1016/j.joms.2009.07.056 ↗
12. Effect of Implant Drill Characteristics on Heat Generation in Osteotomy Sites: A Pilot Study10.1111/j.1600-0501.2010.02051.x ↗
13. The Effect of Irrigation on Osteotomy Depth and Bur Diameter
14. The Influence of Drill Wear on Cutting Efficiency and Heat Production During Osteotomy Preparation for Dental Implants: A Study of Drill Durability
15. Insufficient Irrigation Induces Peri-Implant Bone Resorption: An in Vivo Histologic Analysis in Sheep10.1111/clr.12127 ↗
16. Dos Santos, P.L., Pereira Queiroz, T., Margonar, R., de Souza Carvalho, A.C.G., Betoni, W., Rodrigues Rezende, R.R., and Garcia, R. (2014). Evaluation of Bone Heating, Drill Deformation, and Drill Roughness after Implant Osteotomy: Guided Surgery and Classic Drilling Procedure. Int. J. Oral Maxillofac. Implant., 29.10.11607/jomi.2919 ↗
17. The Influence of Drill Length and Irrigation System on Heat Production During Osteotomy Preparation for Dental Implants: An Ex Vivo Study10.1111/clr.12827 ↗
18. Factors Affecting Heat Generation During Implant Site Preparation: A Review of Biologic Observations and Future Considerations
19. Guided Implant Surgery Risks and Their Prevention10.1111/prd.12292 ↗
20. Internal Bone Temperature Change During Guided Surgery Preparations for Dental Implants: An in Vitro Study10.11607/jomi.2854 ↗
21. Flapless Implant Surgery: Review of the Literature and Report of 2 Cases with Computer-Guided Surgical Approach10.1016/j.joms.2007.12.010 ↗
22. Kivovics, M., Pénzes, D., Németh, O., and Mijiritsky, E. (2020). The Influence of Surgical Experience and Bone Density on the Accuracy of Static Computer-Assisted Implant Surgery in Edentulous Jaws Using a Mucosa-Supported Surgical Template with a Half-Guided Implant Placement Protocol—A Randomized Clinical Study. Materials, 13.10.3390/ma13245759 ↗
23. Computer-Supported Implant Planning and Guided Surgery: A Narrative Review10.1111/clr.12638 ↗
24. Comparison of the Accuracy of Implant Position Using Surgical Guides Fabricated by Additive and Subtractive Techniques10.1111/jopr.13161 ↗
25. How Accurate Are Implant Surgical Guides Produced with Desktop Stereolithographic 3-Dimentional Printers?10.1016/j.joms.2017.08.001 ↗
26. Team R Core (2013). R: A Language and Environment for Statistical Computing, Team R Core.
27. Guided Implant Placement Using an Internally Cooling Surgical Template–a Technical Note10.1563/aaid-joi-D-19-00312 ↗
28. Comparison of Heat Generation between Internally Guided (Cannulated) Single Drill and Traditional Sequential Drilling with and without a Drill Guide for Dental Implants
29. Thermal Properties of Cancellous Bone10.1002/jbm.820090206 ↗
30. Thermal Effects of a Combined Irrigation Method during Implant Site Drilling. A Standardized in Vitro Study Using a Bovine Rib Model10.1111/clr.12032 ↗
31. Aghvami, M., Brunski, J.B., Serdar Tulu, U., Chen, C.H., and Helms, J.A. (2018). A Thermal and Biological Analysis of Bone Drilling. J. Biomech. Eng., 140.10.1115/1.4040312 ↗
32. Numerical and Experimental Analyses on the Temperature Distribution in the Dental Implant Preparation Area When Using a Surgical Guide10.1111/jopr.12488 ↗
33. Intraosseous Heat Generation During Osteotomy Performed Freehand and through Template with an Integrated Metal Guide Sleeve: An in Vitro Study10.1097/ID.0000000000000763 ↗
34. Measurement of Thermal Conductivity of Bovine Cortical Bone10.1016/S1350-4533(01)00003-0 ↗
35. Experimental Analysis of Temperature Differences During Implant Site Preparation: Continuous Drilling Technique Versus Intermittent Drilling Technique10.1563/aaid-joi-D-17-00077 ↗
36. Effect of Surgical Drill Guide and Irrigans Temperature on Thermal Bone Changes During Drilling Implant Sites-Thermographic Analysis on Bovine Ribs10.2298/VSP141208041M ↗