Evaluation of Specific Absorption Rate in Three-Layered Tissue Model at 13.56 MHz and 40.68 MHz for Inductively Powered Biomedical Implants
Source abstract
Abstract
This paper presents an optimized 3-coil inductive wireless power transfer (WPT) system at 13.56 MHz and 40.68 MHz to show and compare the specific absorption rate (SAR) effects on human tissue. This work also substantiates the effects of perfect alignment, lateral and/or angular misalignments on the power transfer efficiency (PTE) of the proposed WPT system. Additionally, the impacts of different tissue composition, input power and coil shape on the SAR are analyzed. The distance between the external and implantable coils is 10 mm. The results have been verified through simulations and measurements. The simulated results show that the SAR of the system at 40.68 MHz had crossed the limit designated by the Federal Communications Commission and hence, it is unsafe and causes tissue damage. Measurement results of the system in air medium show that the optimized printed circuit board coils at 13.56 MHz achieved a PTE of 41.7% whereas PTE waned to 18.2% and 15.4% at 10 mm of lateral misalignment and 60° of angular misalignment respectively. The PTE of a combination of 10 mm lateral misalignment and 60° angular misalignment is 21%. To analyze in a real-environment, a boneless pork sample with 10 mm of thickness is placed as a medium between the external and implantable coils. At perfect alignment, the PTE through pork sample is 30.8%. A RF power generator operating at 13.56 MHz provides 1 W input power to the external coil and the power delivered to load through the air and tissue mediums are 347 mW and 266 mW respectively.
Source: crossref · Quality: baseline
Research context
OpenAlex topics: Wireless Power Transfer Systems · Energy Harvesting in Wireless Networks · Wireless Body Area Networks
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
Affiliations
School of Electronics Engineering, College of IT Engineering, Kyungpook National University, Daegu 702-701, Korea
References (34)
Show source-provided references
1. Design and optimization of printed spiral coils for efficient transcutaneous inductive power transmission10.1109/TBCAS.2007.913130 ↗
2. Design and optimization of resonance-based efficient wireless power delivery systems for biomedical implants10.1109/TBCAS.2010.2072782 ↗
3. Midfield wireless power transfer for bioelectronics10.1109/MCAS.2015.2418999 ↗
4. Optimal design of wireless power transmission links for millimeter-sized biomedical implants10.1109/TBCAS.2014.2370794 ↗
5. Wireless power transfer strategies for implantable bioelectronics10.1109/RBME.2017.2683520 ↗
6. Ha, D., Lee, T.C., Weber, D.J., and Chappell, W.J. (2014, January 1–6). Power distribution to multiple implanted sensor devices using a multiport bandpass filter (BPF) approach. Proceedings of the 2014 IEEE MTT-S International Microwave Symposium (IMS2014), Tampa, FL, USA.
7. Poon, A.S.Y., O’Driscoll, S., and Meng, T.H. (2007, January 22–26). Optimal operating frequency in wireless power transmission for implantable devices. Proceedings of the 2007 29th Annual International Conference of the IEEE Engineering in Medicine and Biology Society, Lyon, France.10.1109/IEMBS.2007.4353634 ↗
8. Chang, T.C., Weber, M., Charthad, J., Nikoozadeh, A., Yakub, P.T.K., and Arbabian, A. (2015, January 21–24). Design of high-efficiency miniaturized ultrasonic receivers for powering medical implants with reconfigurable power levels. Proceedings of the 2015 IEEE International Ultrasonics Symposium (IUS), Taipei, Taiwan.
9. Practical inductive link design for biomedical wireless power transfer: A tutorial10.1109/TBCAS.2018.2846020 ↗
10. Position and orientation insensitive wireless power transmission for enercage-homecage system10.1109/TBME.2017.2691720 ↗
11. Conformal phased surfaces for wireless powering of bioelectronic microdevices10.1038/s41551-017-0043 ↗
12. Analysis and optimization of spiral circular inductive coupling link for bio-implanted applications on air and within human tissue10.3390/s140711522 ↗
13. Design and optimization of a 3-coil inductive link for efficient wireless power pransmission10.1109/TBCAS.2011.2158431 ↗
14. Yao, Y., Meng, X., Tsui, C.Y., and Ki, W.H. (2018, January 26–30). Polyimide-based flexible 3-coil inductive link design and optimization. Proceedings of the 2018 IEEE Asia Pacific Conference on Circuits and Systems (APCCAS), Chengdu, China.10.1109/APCCAS.2018.8605614 ↗
15. Robust wireless power transmission to mm-sized free-floating distributed implants10.1109/TBCAS.2017.2663358 ↗
16. Assessing human exposure to electromagnetic fields from wireless power transmission systems10.1109/JPROC.2013.2245851 ↗
17. Wireless power transfer to deep-tissue microimplants10.1073/pnas.1403002111 ↗
18. An end-to-end implanted brain–machine interface antenna system performance characterizations and development10.1109/TAP.2017.2700163 ↗
19. Near-field inductive-coupling link to power a three-dimensional millimeter-sze antenna for brain implantable medical devices10.1109/TBME.2017.2778729 ↗
20. Yeon, P., Mirbozorgi, S.A., Ash, B., Eckhardt, H., and Ghovanloo, M. (2016). Fabrication and microassembly of a mm-sized floating probe for a distributed wireless neural interface. Micromachines, 7.10.3390/mi7090154 ↗
21. Design of spiral circular coils in wet and dry tissue for bio-implanted micro-system applications10.2528/PIERM13052707 ↗
22. Comparative analysis of two-coil and three-coil structures for wireless power transfer10.1109/TPEL.2016.2526780 ↗
23. A methodology for making a three-coil wireless power transfer system more energy efficient than a two-coil counterpart for extended transfer distance10.1109/TPEL.2014.2312020 ↗
24. Mirbozorgi, S.A., Gosselin, B., and Sawan, M. (September, January 28). A transcutaneous power transfer interface based on a multicoil inductive link. Proceedings of the 2012 Annual International Conference of the IEEE Engineering in Medicine and Biology Society, San Diego, CA, USA.10.1109/EMBC.2012.6346265 ↗
25. Nair, V.V., and Choi, J.R. (2016). An efficiency enhancement technique for a wireless power transmission system based on a multiple coil switching technique. Energies, 9.10.3390/en9030156 ↗
26. Radio-frequency coils in implantable devices: Misalignment analysis and design procedure10.1109/TBME.1987.326088 ↗
27. Analysis and optimization of wireless power transfer efficiency considering the tilt angle of a coil10.26866/jees.2018.18.1.13 ↗
28. Khan, S.R., and Choi, G.S. (2016). Analysis and optimization of four-coil planar magnetically coupled printed spiral resonators. Sensors, 16.10.3390/s16081219 ↗
29. Ultracompact implantable design with integrated wireless power transfer and RF transmission capabilities10.1109/TBCAS.2017.2787649 ↗
30. (2018, July 15). Dielectric Properties of Body Tissues. Available online: http://niremf.ifac.cnr.it/tissprop/htmlclie/htmlclie.php.
31. Yellappa, P., Lee, Y.R., and Choi, J.R. (2018, January 27–30). Analysis of coil misalignment issue in resonance-based wireless power transmission system for implantable biomedical applications. Proceedings of the 2018 IEEE International Symposium on Circuits and Systems (ISCAS), Florence, Italy.10.1109/ISCAS.2018.8351407 ↗
32. Modeling and optimization of printed spiral coils in air, saline, and muscle tissue environments10.1109/TBCAS.2009.2025366 ↗
33. High-efficiency wireless power transfer for biomedical implants by optimal resonant load transformation10.1109/TCSI.2012.2209297 ↗
34. Efficient four-coil wireless power transfer for deep brain stimulation10.1109/TMTT.2017.2658560 ↗