Non-Contact SpO2 Prediction System Based on a Digital Camera
Source abstract
Abstract
Patients with the COVID-19 condition require frequent and accurate blood oxygen saturation (SpO2) monitoring. The existing pulse oximeters, however, require contact-based measurement using clips or otherwise fixed sensor units or need dedicated hardware which may cause inconvenience and involve additional appointments with the patient. This study proposes a computer vision-based system using a digital camera to measure SpO2 on the basis of the imaging photoplethysmography (iPPG) signal extracted from the human’s forehead without the need for restricting the subject or physical contact. The proposed camera-based system decomposes the iPPG obtained from the red and green channels into different signals with different frequencies using a signal decomposition technique based on a complete Ensemble Empirical Mode Decomposition (EEMD) technique and Independent Component Analysis (ICA) technique to obtain the optical properties from these wavelengths and frequency channels. The proposed system is convenient, contactless, safe and cost-effective. The preliminary results for 70 videos obtained from 14 subjects of different ages and with different skin tones showed that the red and green wavelengths could be used to estimate SpO2 with good agreement and low error ratio compared to the gold standard of pulse oximetry (SA210) with a fixed measurement position.
Source: crossref · Quality: baseline
Research context
OpenAlex topics: Non-Invasive Vital Sign Monitoring · Hemodynamic Monitoring and Therapy · Optical Imaging and Spectroscopy Techniques
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
Affiliations
Electrical Engineering Technical College, Middle Technical University, Baghdad 10022, Iraq
School of Engineering, University of South Australia, Mawson Lakes, SA 5095, Australia
References (50)
Show source-provided references
1. (2021, May 01). COVID-19 Coronavirus Pandemic. Available online: https://www.worldometers.info/coronavirus/.
2. Signs and symptoms to determine if a patient presenting in primary care or hospital outpatient settings has COVID-19 disease
3. The international European Academy of Neurology survey on neurological symptoms in patients with COVID-19 infection10.1111/ene.14407 ↗
4. Current progress of photoplethysmography and SPO2 for health monitoring10.1007/s13534-019-00097-w ↗
5. Alharbi, S., Hu, S., Mulvaney, D., Barrett, L., Yan, L., Blanos, P., Elsahar, Y., and Adema, S. (2019). Oxygen saturation measurements from green and orange illuminations of multi-wavelength optoelectronic patch sensors. Sensors, 19.10.20944/preprints201810.0578.v1 ↗
6. Herrmann, C., and Metzler, J. (2020). Distant Pulse Oximetry. Laser Doppler Vibrometry for Non-Contact Diagnostics, Springer.10.1007/978-3-030-46691-6_7 ↗
7. Smartphone-Based SpO2 Measurement by Exploiting Wavelengths Separation and Chromophore Compensation10.1145/3360725 ↗
8. Non-contact measurement of oxygen saturation with an RGB camera10.1364/BOE.6.003320 ↗
9. Monitoring of cardiorespiratory signal: Principles of remote measurements and review of methods10.1109/ACCESS.2017.2735419 ↗
10. Reflectance photoplethysmography as noninvasive monitoring of tissue blood perfusion10.1109/TBME.2015.2417863 ↗
11. Pulse oximetry: Understanding its basic principles facilitates appreciation of its limitations10.1016/j.rmed.2013.02.004 ↗
12. Pulse oximetry in the oesophagus10.1088/0967-3334/27/1/R01 ↗
13. Jubran, A. (2012). Pulse oximetry. Applied Physiology in Intensive Care Medicine 1, Springer.10.1007/978-3-642-28270-6_12 ↗
14. Dedicated exposure control for remote photoplethysmography10.1109/ACCESS.2020.3003548 ↗
15. A low-cost, camera-based continuous ppg monitoring system using laplacian pyramid10.1016/j.smhl.2018.07.024 ↗
16. Robot Assisted Instantaneous Heart Rate Estimator using Camera based Remote Photoplethysmograpy via Plane-Orthogonal-to-Skin and Finite State Machine
17. Remote measurement of cardiopulmonary signal using an unmanned aerial vehicle10.1088/1757-899X/405/1/012001 ↗
18. Benezeth, Y., Li, P., Macwan, R., Nakamura, K., Gomez, R., and Yang, F. (2018, January 4–7). Remote Heart Rate Variability for Emotional State Monitoring. Proceedings of the 2018 IEEE EMBS International Conference on Biomedical & Health Informatics (BHI), Las Vegas, NV, USA.10.1109/BHI.2018.8333392 ↗
19. On the minimal adequate sampling frequency of the photoplethysmogram for pulse rate monitoring and heart rate variability analysis in mobile and wearable technology10.2478/msr-2019-0030 ↗
20. Nagpal, S. (2019). Contactless measurement of Heart Rate Variability by Extracting PPG Signals from Video. [Master’s Thesis, Eindhoven University of Technology].
21. Chen, M., Zhu, Q., Zhang, H., Wu, M., and Wang, Q. (2019, January 19–22). Respiratory Rate Estimation from Face Videos. Proceedings of the 2019 IEEE EMBS International Conference on Biomedical & Health Informatics (BHI), Chicago, IL, USA.10.1109/BHI.2019.8834499 ↗
22. Luguern, D., Benezeth, Y., Moser, V., Dunbar, L.A., Braun, F., Lemkaddem, A., Nakamura, K., Gomez, R., and Dubois, J. (2020, January 20–22). Remote Photoplethysmography Combining Color Channels with SNR Maximization for Respiratory Rate Assessment. Proceedings of the 2020 14th International Symposium on Medical Information Communication Technology (ISMICT), Nara, Japan.10.1109/ISMICT48699.2020.9152720 ↗
23. Luguern, D., Perche, S., Benezeth, Y., Moser, V., Dunbar, L.A., Braun, F., Lemkaddem, A., Nakamura, K., Gomez, R., and Dubois, J. (2020, January 14–19). An Assessment of Algorithms to Estimate Respiratory Rate from the Remote Photoplethysmogram. Proceedings of the 2020 IEEE/CVF Conference on Computer Vision and Pattern Recognition Workshops, Seattle, WA, USA.10.1109/CVPRW50498.2020.00160 ↗
24. Web-Based Smartphone Algorithm for Calculating Blood Pressure From Photoplethysmography Remotely in a General Adult Population: Validation Study10.2196/19187 ↗
25. Chowdhury, M.H., Shuzan, M.N.I., Chowdhury, M.E., Mahbub, Z.B., Uddin, M.M., Khandakar, A., and Reaz, M.B.I. (2020). Estimating blood pressure from the photoplethysmogram signal and demographic features using machine learning techniques. Sensors, 20.10.3390/s20113127 ↗
26. Huang, P.-W., Lin, C.-H., Chung, M.-L., Lin, T.-M., and Wu, B.-F. (2017, January 12–15). Image Based Contactless Blood Pressure Assessment Using Pulse Transit Time. Proceedings of the 2017 International Automatic Control Conference (CACS), Pingtung, Taiwan.10.1109/CACS.2017.8284275 ↗
27. An overview of remote photoplethysmography methods for vital sign monitoring
28. Khanam, F.-T.-Z., Al-Naji, A., and Chahl, J. (2019). Remote monitoring of vital signs in diverse non-clinical and clinical scenarios using computer vision systems: A review. Appl. Sci., 9.10.3390/app9204474 ↗
29. Pulse oximetry based on photoplethysmography imaging with red and green light10.1007/s10877-019-00449-y ↗
30. Posture effects on the calibratability of remote pulse oximetry in visible light10.1088/1361-6579/ab051a ↗
31. Contactless multiple wavelength photoplethysmographic imaging: A first step toward “SpO 2 camera” technology10.1007/s10439-005-5763-2 ↗
32. New principle for measuring arterial blood oxygenation, enabling motion-robust remote monitoring
33. Calibration of contactless pulse oximetry10.1213/ANE.0000000000001381 ↗
34. Non-contact video-based vital sign monitoring using ambient light and auto-regressive models10.1088/0967-3334/35/5/807 ↗
35. Continuous non-contact vital sign monitoring in neonatal intensive care unit10.1049/htl.2014.0077 ↗
36. Non-contact estimation of heart rate and oxygen saturation using ambient light10.1364/BOE.6.000086 ↗
37. Casalino, G., Castellano, G., and Zaza, G. (2020, January 7–10). A mHealth Solution for Contact-Less Self-Monitoring of Blood Oxygen Saturation. Proceedings of the 2020 IEEE Symposium on Computers and Communications (ISCC), Rennes, France.10.1109/ISCC50000.2020.9219718 ↗
38. Torres, M.E., Colominas, M.A., Schlotthauer, G., and Flandrin, P. (2011, January 22–27). A Complete Ensemble Empirical Mode Decomposition with Adaptive Noise. Proceedings of the 2011 IEEE International Conference on Acoustics, Speech and Signal Processing (ICASSP), Prague, Czech Republic.10.1109/ICASSP.2011.5947265 ↗
39. Hyvärinen, A., Hurri, J., and Hoyer, P.O. (2009). Independent component analysis. Natural Image Statistics, Springer.10.1007/978-1-84882-491-1 ↗
40. Non-contact detection of oxygen saturation based on visible light imaging device using ambient light10.1364/OE.21.017464 ↗
41. Tkalcic, M., and Tasic, J.F. (2003, January 22–24). Colour Spaces: Perceptual, Historical and Applicational Background; In Proceedings of the IEEE Region 8 EUROCON 2003. Computer as a Tool, Ljubljana, Slovenia.
42. Soil color analysis based on a RGB camera and an artificial neural network towards smart irrigation: A pilot study10.1016/j.heliyon.2021.e06078 ↗
43. Cardiopulmonary signal acquisition from different regions using video imaging analysis10.1080/21681163.2018.1441075 ↗
44. Xu, Y., Luo, M., Li, T., and Song, G. (2017). ECG signal de-noising and baseline wander correction based on CEEMDAN and wavelet threshold. Sensors, 17.10.3390/s17122754 ↗
45. Filtering of surface EMG using ensemble empirical mode decomposition10.1016/j.medengphy.2012.10.009 ↗
46. Remote optical cardiopulmonary signal extraction with noise artifact removal, multiple subject detection & long-distance10.1109/ACCESS.2018.2811392 ↗
47. Removal of muscle artifact from EEG data: Comparison between stochastic (ICA and CCA) and deterministic (EMD and wavelet-based) approaches10.1186/1687-6180-2012-127 ↗
48. ICA-based procedures for removing ballistocardiogram artifacts from EEG data acquired in the MRI scanner10.1016/j.neuroimage.2004.09.041 ↗
49. Zhou, W., and Gotman, J. (2004, January 1–5). Removal of EMG and ECG artifacts from EEG based on wavelet transform and ICA. Proceedings of the 26th Annual International Conference of the IEEE Engineering in Medicine and Biology Society, San Francisco, CA, USA.
50. Limits of agreement (Bland-Altman method)10.1136/bmj.f1630 ↗