Effects of repetitive insertion/removal and occlusal load on the retentive force of rest plate–I bar clasps made by selective laser melting

Yuki UEKUBO, Yoshimitsu KATO, Keita TOMONO, Mitsuo KATO, Juro WADACHI, Shinji TAKEMOTO, Shuichiro YAMASHITA

Dental Materials Journal· Volume 44· Issue 1

Abstract

The objective of this study was to investigate the effects of repetitive removal/insertion and a load equivalent to the occlusal force on the retentive force of rest plate-I bar (RPI) clasps fabricated by selective laser melting (SLM) on powder bed fusion. RPI clasp fabricated by SLM and casting were evaluated to the retentive force by repetitive removal/insertion (SLM-RI and Cast-RI), and the retentive force of RPI clasp by SLM was also evaluated by repetitive removal/insertion under load (SLM-L). SLM-RI group was kept superior retentive force during repetitive removal/insertion than Cast-RI group. SLM-L group showed the decrease of retentive force, and all specimens fractured during less than 7,000 repetitive removal/insertion. In SLM-L, the fracture of RPI clasp occurred at minor connector due to stress concentration. In conclusion, the RPI clasp by SLM demonstrated the superior retentive force and fitting for repetitive removal/insertion, yet improvements were necessary to withstand a load equivalent to occlusal force.
Source: openalex_inverted_index · Quality: baseline

Research context

OpenAlex topics: Additive Manufacturing Materials and Processes · Dental materials and restorations · Additive Manufacturing and 3D Printing Technologies

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

Department of Removable Partial Prosthodontics, Tokyo Dental College

Department of Biomedical Engineering, Iwate Medical University

References (40)

Show source-provided references
1. 1) Singh BP, Gauthier G, Rompre P, De Grandmont P, Emami E. A 30-year follow-up of partial removable dental prostheses in a university dental school setting. J Prosthodont 2016; 25: 544-549.10.1111/jopr.12336 ↗
2. 2) Suenaga H, Kubo K, Hosokawa R, Kuriyagawa T, Sasaki K. Effects of occlusal rest design on pressure distribution beneath the denture base of a distal extension removable partial denture —An in vivo study. Int J Prosthodont 2014; 27: 469-471.10.11607/ijp.3847 ↗
3. 3) Richert R, Alsheghri AA, Alageel O, Caron E, Song J, Ducret M, et al. Analytical model of I-bar clasps for removable partial dentures. Dent Mater 2021; 37: 1066-1072.10.1016/j.dental.2021.03.018 ↗
4. 4) Tasaka A, Kato Y, Odaka K, Matsunaga S, Goto TK, Abe S, et al. Accuracy of clasps fabricated with three different CAD/CAM technologies: Casting, milling, and selective laser sintering. Int J Prosthodont 2019; 32: 526-529.10.11607/ijp.6363 ↗
5. 5) Tasaka A, Shimizu T, Kato Y, Okano H, Ida Y, Higuchi S, et al. Accuracy of removable partial denture framework fabricated by casting with a 3D printed pattern and selective laser sintering. J Prosthodont Res 2020; 64: 224-230.10.1016/j.jpor.2019.07.009 ↗
6. 6) Carneiro Pereira AL, Bezerra de Medeiros AK, de Sousa Santos K, Oliveira de Almeida É, Seabra Barbosa GA, da Fonte Porto Carreiro A. Accuracy of CAD-CAM systems for removable partial denture framework fabrication: A systematic review. J Prosthet Dent 2021; 125: 241-248.10.1016/j.prosdent.2020.01.003 ↗
7. 7) Bajunaid SO, Altwaim B, Alhassan M, Alammari R. The fit accuracy of removable partial denture metal frameworks using conventional and 3D printed techniques: An in vitro study. J Contemp Dent Pract 2019; 20: 476-481.10.5005/jp-journals-10024-2542 ↗
8. 8) Yunizar MF, Watanabe M, Ichikawa T. Current development status of additive manufacturing technologies for fabricating removable partial denture frameworks: A literature review. Int J Comput Dent 2022; 25: 57-70.
9. 9) Hwang S, An S, Robles U, Rumpf RC. Process parameter optimization for removable partial denture frameworks manufactured by selective laser melting. J Prosthet Dent 2023; 129: 191-198.10.1016/j.prosdent.2021.04.021 ↗
10. 10) Takaichi A, Suyalatu, Nakamoto T, Joko N, Nomura N, Tsutsumi Y, et al. Microstructures and mechanical properties of Co-29Cr-6Mo alloy fabricated by selective laser melting process for dental applications. J Mech Behav Biomed Mater 2013; 21: 67-76.10.1016/j.jmbbm.2013.01.021 ↗
11. 11) Qian B, Saeidi K, Kvetková L, Lofaj F, Xiao C, Shen Z. Defects-tolerant Co-Cr-Mo dental alloys prepared by selective laser melting. Dent Mater 2015; 31: 1435-1444.10.1016/j.dental.2015.09.003 ↗
12. 12) Kajima Y, Takaichi A, Nakamoto T, Kimura T, Yogo Y, Ashida M, et al. Fatigue strength of Co-Cr-Mo alloy clasps prepared by selective laser melting. J Mech Behav Biomed Mater 2016; 59: 446-458.10.1016/j.jmbbm.2016.02.032 ↗
13. 13) Hedberg YS, Qian B, Shen Z, Virtanen S, Wallinder IO. In vitro biocompatibility of CoCrMo dental alloys fabricated by selective laser melting. Dent Mater 2014; 30: 525-534.10.1016/j.dental.2014.02.008 ↗
14. 14) Okazaki Y, Ishino A, Higuchi S. Chemical, physical, and mechanical properties and microstructures of laser-sintered Co-25Cr-5Mo-5W (SP2) and W-free Co-28Cr-6Mo alloys for dental applications. Materials (Basel) 2019; 12: 4039.10.3390/ma12244039 ↗
15. 15) Kim HR, Jang SH, Kim YK, Son JS, Min BK, Kim KH, et al. Microstructures and mechanical properties of Co-Cr dental alloys fabricated by three CAD/CAM-based processing techniques. Materials (Basel) 2016; 9: 596.10.3390/ma9070596 ↗
16. 16) Kobayashi H, Tasaka A, Higuchi S, Yamashita S. Influence of molding angle on the trueness and defects of removable partial denture frameworks fabricated by selective laser melting. J Prosthodont Res 2022; 66: 589-599.10.2186/jpr.JPR_D_21_00175 ↗
17. 17) Wang H, Shu X, Zhao J. Influence of build angle and polishing roughness on corrosion resistance of 316L stainless steel fabricated by SLM method. Materials (Basel) 2022; 15: 4020.10.3390/ma15114020 ↗
18. 18) Cheng H, Xu M, Zhang H, Wu W, Zheng M, Li X. Cyclic fatigue properties of cobalt-chromium alloy clasps for partial removable dental prostheses. J Prosthet Dent 2010; 104: 389-396.10.1016/S0022-3913(10)60173-4 ↗
19. 19) Marie A, Keeling A, Hyde TP, Nattress BR, Pavitt S, Murphy RJ, et al. Deformation and retentive force following in vitro cyclic fatigue of cobalt-chrome and aryl ketone polymer (AKP) clasps. Dent Mater 2019; 35: e113-e121.10.1016/j.dental.2019.02.028 ↗
20. 20) Tanaka A, Miyake N, Hotta H, Takemoto S, Yoshinari M, Yamashita S. Change in the retentive force of Akers clasp for zirconia crown by repetitive insertion and removal test. J Prosthodont Res 2019; 63: 447-452.10.1016/j.jpor.2019.02.005 ↗
21. 21) Torii M, Nakata T, Takahashi K, Kawamura N, Shimpo H, Ohkubo C. Fitness and retentive force of cobalt-chromium alloy clasps fabricated with repeated laser sintering and milling. J Prosthodont Res 2018; 62: 342-346.10.1016/j.jpor.2018.01.001 ↗
22. 22) Zlatarić DK, Celebić A, Valentić-Peruzović M. The effect of removable partial dentures on periodontal health of abutment and non-abutment teeth. J Periodontol 2002; 73: 137-144.10.1902/jop.2002.73.2.137 ↗
23. 23) Kato Y, Tasaka A, Kato M, Wadachi J, Takemoto S, Yamashita S. Effects of repetitive insertion/removal cycles and simulated occlusal loads on retention of denture retainers. Dent Mater J 2021; 40: 1277-1283.10.4012/dmj.2020-462 ↗
24. 24) Davis H, Victor LS. Mccracken’s removable partial prosthodontics, 5th ed. Missouri, USA, Mosby Inc. 1977, p.74-75.
25. 25) Tatematsu M, Mori T, Kawaguchi T, Takeuchi K, Hattori M, Morita I, et al. Masticatory performance in 80-year-old individuals. Gerodontology 2004; 21: 112-119.10.1111/j.1741-2358.2004.00018.x ↗
26. 26) Sato Y, Abe Y, Yuasa Y, Akagawa Y. Effect of friction coefficient on Akers clasp retention. J Prosthet Dent 1997; 78: 22-27.10.1016/S0022-3913(97)70083-0 ↗
27. 27) Rues S, Tasaka A, Fleckenstein I, Yamashita S, Rammelsberg P, Boehm S, et al. Fit and retention of cobalt-chromium removable partial denture frameworks fabricated with selective laser melting. J Funct Biomater 2023; 14: 416.10.3390/jfb14080416 ↗
28. 28) Zhang M, Gan N, Qian H, Jiao T. Retentive force and fitness accuracy of cobalt-chrome alloy clasps for removable partial denture fabricated with SLM technique. J Prosthodont Res 2022; 66: 459-465.10.2186/jpr.JPR_D_21_00017 ↗
29. 29) Ma KN, Chen H, Shen YR, Zhou YS, Wang Y, Sun YC. [Finite element analyses of retention of removable partial denture circumferential clasps manufactured by selective laser melting]. Beijing Da Xue Xue Bao Yi Xue Ban 2022; 54: 105-112.
30. 30) Peng PW, Hsu CY, Huang HY, Chao JC, Lee WF. Trueness of removable partial denture frameworks additively manufactured with selective laser melting. J Prosthet Dent 2022; 127: 122-127.10.1016/j.prosdent.2020.06.035 ↗
31. 31) Sokolowski A, Horak D, Behlau A, Madreiter-Sokolowski C, Lorenzoni M, Sokolowski A. Evaluation of two printing techniques for maxillary removable partial denture frameworks. J Prosthet Dent 2024; 131: 707e1-707e8.10.1016/j.prosdent.2024.01.015 ↗
32. 32) Sato Y. Clinical methods for adjusting retention force of cast clasps. J Prosthet Dent 1999; 82: 557-561.10.1016/S0022-3913(99)70054-5 ↗
33. 33) Zhou Y, Li N, Yan J, Zeng Q. Comparative analysis of the microstructures and mechanical properties of Co-Cr dental alloys fabricated by different methods. J Prosthet Dent 2018; 120: 617-623.10.1016/j.prosdent.2017.11.015 ↗
34. 34) Tomono K, Kato Y, Wadachi J, Tasaka A, Takemoto S, Yamashita S. Influence of different undercut depths of clasp fabricated by selective laser melting on retentive force. Eur J Prosthodont Restor Dent 2024; 32: 261-269.
35. 35) Alageel O, Alsheghri AA, Algezani S, Caron E, Tamimi F. Determining the retention of removable partial dentures. J Prosthet Dent 2019; 122: 55-62.10.1016/j.prosdent.2018.06.015 ↗
36. 36) Fu W, Liu S, Jiao J, Xie Z, Huang X, Lu Y, et al. Wear resistance and biocompatibility of Co-Cr dental alloys fabricated with CAST and SLM techniques. Materials (Basel) 2022; 15: 3263.10.3390/ma15093263 ↗
37. 37) Kajima Y, Takaichi A, Nakamoto T, Kimura T, Kittikundecha N, Tsutsumi Y, et al. Effect of adding support structures for overhanging part on fatigue strength in selective laser melting. J Mech Behav Biomed Mater 2018; 78: 1-9.10.1016/j.jmbbm.2017.11.009 ↗
38. 38) Xie W, Zheng M, Wang J, Li X. The effect of build orientation on the microstructure and properties of selective laser melting Ti-6Al-4V for removable partial denture clasps. J Prosthet Dent 2020; 123: 163-172.10.1016/j.prosdent.2018.12.007 ↗
39. 39) Yan X, Jiang R, Li W, Lin H. Oriented face-centered cubic to hexagonal close-packed martensitic transition, grain morphology, and mechanical properties of Co-Cr alloy fabricated by selective laser melting. J Prosthet Dent 2022; 127: 282-287.10.1016/j.prosdent.2020.07.036 ↗
40. 40) Ma K, Chen H, Shen Y, Guo Y, Li W, Wang Y, et al. Feasibility study and material selection for powder-bed fusion process in printing of denture clasps. Comput Biol Med 2023; 157: 106772.10.1016/j.compbiomed.2023.106772 ↗