Adapting Titanium Implants to the Elasticity of Bone by Comparison of Spring Stiffness
Source abstract
Abstract
The aim of adapted endoprostheses is to achieve at least the same percentage of negative elongation under the same stress in an implant-bone-combination. The negative elongation is like the downward deflexion of a spring under pressure. Due to the limitations mentioned above, the comparison of spring stiffness of bone and implant is the most promising way to achieve the adapted implant elasticity aimed at.
Source: openalex_inverted_index · Quality: baseline
Research context
OpenAlex topics: Orthopaedic implants and arthroplasty · Dental Implant Techniques and Outcomes · Bone Tissue Engineering Materials
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (13)
Show source-provided references
1. 10.1002/adem.20050006010.1002/adem.200500060 ↗
2. 10.1016/j.biomaterials.2004.01.02110.1016/j.biomaterials.2004.01.021 ↗
3. 10.1002/sia.125210.1002/sia.1252 ↗
4. 10.1016/S0300-9785(85)80078-810.1016/S0300-9785(85)80078-8 ↗
5. Thull R. · Med. Prog. Through Technol. · 1990 · 16 · 225
6. 10.1002/jbm.121210.1002/jbm.1212 ↗
7. J. A. Helsen H. J. Breme Met. as Biomaterials Wiley Ser. in Biomaterials Sci. and Eng. J. Wiley & Sons Ltd. England1998.
8. 10.1016/0142-9612(96)87281-410.1016/0142-9612(96)87281-4 ↗
9. 10.1007/s00198-004-1728-910.1007/s00198-004-1728-9 ↗
10. G. Hohenhoff H. Haferkamp A. Ostendorf O. Meier S. Czerner C. Stukenborg‐Colsman S. Ostermeier C. Hurschler;PICALOin Melbourne Australia 2006.
11. J. Wilden H. Frank C. Theiler T. Seefeld G. Sepold 47. Int. Scientific Colloq.TU Ilmenau Germany 23.–26. September2002.
12. 10.1002/(SICI)1527-2648(200005)2:5<270::AID-ADEM270>3.0.CO;2-910.1002/(SICI)1527-2648(200005)2:5<270::AID-ADEM270>3.0.CO;2-9 ↗
13. 10.1016/S0261-3069(99)00017-510.1016/S0261-3069(99)00017-5 ↗