Abstract The present paper shows an effective methodology which is based on Finite Element Analysis (FEA) for the fatigue life prediction (FLP) of a Balloon-Expandable (BE) stent made of stainless steel material (AISI316L). BE stent is a tubular, often mesh-like, structure which is expanded inside a diseased (stenosed) artery segment to restore blood flow and keep the vessel open following angioplasty. With design lives of 10–15 years, fatigue life prediction of these devices is critical for the designer, as fatigue failure may occur after deployment inside the blood vessel due to the heart beating (typically 4 × 10 7 cycles/year). Therefore, it is crucial to study the effect of stenting conditions and stent design on the fatigue life of BE stents. The objective of this paper is to provide quantitative measures of the stent’s stress amplitude and mean stress which are generated by the cardiac pulse pressure. This allows prediction of the device’s life and optimization of stent designs. In this paper, numerical investigations were conducted using an effective approach based on the well-known stress-based Goodman Diagram (GD) for the FLP of a given stent design. It has been demonstrated that the percentage of artery’s expansion as well as the dimensions of the struts (the fundamental structural unit that forms the stent), have both an impact on the fatigue behavior after stent deployment.
Source: openaire · Quality: candidate unverified · Source record ↗
Research context
OpenAlex topics: Hydrogen embrittlement and corrosion behaviors in metals · Mechanical Failure Analysis and Simulation · Fatigue and fracture mechanics
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (26)
Show source-provided references
1. Stoeckel · Minimally Invasive Therapy and Allied Technology · 2002 · 11 · 13710.1080/136457002760273340 ↗
2. Marrey · Biomaterials · 2006 · 27 · 198810.1016/j.biomaterials.2005.10.012 ↗
3. Schievano · Journal of Biomechanics · 2010 · 43 · 68710.1016/j.jbiomech.2009.10.024 ↗
4. Hsiao · Computer-Aided Design · 2012 · 44 · 75710.1016/j.cad.2012.03.009 ↗
5. Sweeney · International Journal of Fatigue · 2012 · 44 · 20210.1016/j.ijfatigue.2012.04.022 ↗
6. Santos · Journal of the Mechanical Behavior of Biomedical Materials · 2012 · 15 · 7810.1016/j.jmbbm.2012.06.011 ↗
7. Barrera · Computer Methods in Biomechanics and Biomedical Engineering · 2012 · 15 · 93
8. Bathias · International Journal of Fatigue · 2006 · 28 · 143810.1016/j.ijfatigue.2005.09.020 ↗
9. Dixon · Journal of the American Statistical · 1948 · 43 · 10910.1080/01621459.1948.10483254 ↗
10. C. Bathias, A. Pineau, Fatigue des matériaux et des structures 1. Lavoisier, Hermès (2008).
11. Wiersma · Bio-Medical Materials and Engineering · 2006 · 16 · 137
12. Taylor · 2007
13. Park · International Journal of Mathematical, Physical and Engineering Sciences · 2007 · 2 · 103
14. Abaqus Manual, Version 6.11. Dassault systems. May 2011, .
15. James · Biomaterials · 2010 · 31 · 18110.1016/j.biomaterials.2009.10.028 ↗
16. Matthys · Journal of Biomechanics · 2007 · 40 · 347610.1016/j.jbiomech.2007.05.027 ↗
17. C. Lally, P.J. Prendergast, An investigation into the applicability of a Mooney–Rivlin constitutive equation for modeling vascular tissue in cardiovascular stenting procedures, in: Proceedings of the International Congress on Computational Biomechanics, Zaragoza, Spain, 2003, pp. 542–550.
18. Holzapfel · Journal of Biomechanical Engineering · 2004 · 126 · 65710.1115/1.1800557 ↗
19. Gervaso · Journal of Biomechanics · 2008 · 41 · 120610.1016/j.jbiomech.2008.01.027 ↗
20. Lally · Journal of Biomechanics · 2005 · 38 · 157410.1016/j.jbiomech.2004.07.022 ↗
21. Python, Version 2.7, July 2010, .
22. Scheinert · Journal of the American College of Cardiology · 2006 · 45 · 31210.1016/j.jacc.2004.11.026 ↗
23. Murphy · Journal of Materials Science: Materials in Medicine · 2006 · 17 · 1
24. Savage · Annals of Biomedical Engineering · 2004 · 32 · 20210.1023/B:ABME.0000012740.47963.9e ↗
25. Murphy · Annals of Biomedical Engineering · 2003 · 31 · 68610.1114/1.1569268 ↗
26. Optimization based simulation of self-expanding Nitinol stent10.1016/j.matdes.2013.03.012 ↗