Scanning Electrochemical Microscopy with Simultaneous Independent Topography
Source abstract
Abstract
A new method to perform scanning electrochemical microscopy (SECM) and topography simultaneously is described here. The new method uses a conventional scanning microelectrode to sense species released by local electrochemical reactions on the surface of a sample, combined with shear‐force feedback to maintain the probe at a constant distance from the surface of the material. By using shear‐force feedback, larger electrochemical currents can be detected at the microelectrode because the probe is scanned at a closer distance from the surface of the sample. The new method has yielded high lateral resolution topography and SECM images are reported here.
Source: openalex_inverted_index · Quality: baseline
Research context
OpenAlex topics: Force Microscopy Techniques and Applications · Electrochemical Analysis and Applications · Thermography and Photoacoustic Techniques
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
Affiliations
Corrosion Research Center, Department of Chemical Engineering and Materials Science, University of Minnesota, Minneapolis, Minnesota 55455
Department of Physics, University of North Carolina at Charlotte, Charlotte, North Carolina 28223
References
Reference metadata was not supplied for this record.