journal article·1996-04-01

Determination of Improved Selectivity Coefficients of Polymer Membrane Ion‐Selective Electrodes by Conditioning with a Discriminated Ion

Eric Bakker

Publisher page

Abstract

A new method for determining selectivity coefficients of carrier based ion‐selective electrodes toward highly discriminated ions is proposed. For each separately measured ion solution, the interfacial layer of the membrane contacting the sample must contain only the ion to be measured for the experimental determination of Nikolskii coefficients that reflect the true ion‐exchange selectivity of the membrane. For this reason, it is recommended here that the electrode membrane is conditioned in a solution of a discriminated ion that, preferably, is equal to the cation of the tetraphenylborate salt incorporated in the membrane. The selectivity of valinomycin based plasticized poly(vinyl chloride) membrane electrodes has been characterized by using this new method and logarithmic Nikolskii coefficients for over , and have been determined as −4.5, −7.5, and −6.9, respectively. All measured ions showed near‐Nernstian electrode slopes.
Source: openalex_inverted_index · Quality: baseline

Research context

OpenAlex topics: Analytical Chemistry and Sensors · Electrochemical sensors and biosensors · Fuel Cells and Related Materials

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

Department of Chemistry, Auburn University, Auburn, Alabama 36849

References

Reference metadata was not supplied for this record.