Learning Scatter Artifact Correction in Cone-Beam X-Ray CT Using Incomplete Projections with Beam Hole Array
Source abstract
Abstract
X-ray cone-beam computed tomography (CBCT) is a powerful tool for nondestructive testing and evaluation, yet the CT image quality can be compromised by artifact due to X-ray scattering within dense materials such as metals. This problem leads to the need for hardware- and software-based scatter artifact correction to enhance the image quality. Recently, deep learning techniques have merged as a promising approach to obtain scatter-free images efficiently. However, these deep learning techniques rely heavily on training data, often gathered through simulation. Simulated CT images, unfortunately, do not accurately reproduce the real properties of objects, and physically accurate X-ray simulation still requires significant computation time, hindering the collection of a large number of CT images. To address these problems, we propose a deep learning framework for scatter artifact correction using projections obtained solely by real CT scanning. To this end, we utilize a beam-hole array (BHA) to block the X-rays deviating from the primary beam path, thereby capturing scatter-free X-ray intensity at certain detector pixels. As the BHA shadows a large portion of detector pixels, we incorporate several regularization losses to enhance the training process. Furthermore, we introduce radiographic data augmentation to mitigate the need for long scanning time, which is a concern as CT devices equipped with BHA require two series of CT scans. Experimental validation showed that the proposed framework outperforms a baseline method that learns simulated projections where the entire image is visible and does not contain scattering artifacts.
Source: openalex_inverted_index · Quality: baseline
Research context
OpenAlex topics: Medical Imaging Techniques and Applications · Advanced X-ray and CT Imaging · Radiomics and Machine Learning in Medical Imaging
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (49)
Show source-provided references
1. Carmignato, S., Dewulf, W., Leach, R. (eds.): Industrial X-Ray Computed Tomography. Springer, Switzerland (2018). https://doi.org/10.1007/978-3-319-59573-310.1007/978-3-319-59573-3 ↗
2. Feldkamp, L.A., Davis, L.C., Kress, J.W.: Practical cone-beam algorithm. J. Opt. Soc. Am. A 1(6), 612–619 (1984). https://doi.org/10.1364/josaa.1.00061210.1364/josaa.1.000612 ↗
3. Lorensen, W.E., Cline, H.E.: Marching cubes: a high resolution 3D surface construction algorithm. In: Proceedings of the Annual Conference on Computer Graphics and Interactive Techniques, pp. 163–169. Association for Computing Machinery, New York (1987). https://doi.org/10.1145/37401.3742210.1145/37401.37422 ↗
4. Endo, M., Tsunoo, T., Nakamori, N., Yoshida, K.: Effect of scattered radiation on image noise in cone beam CT. Med. Phys. 28(4), 469–474 (2001). https://doi.org/10.1118/1.135745710.1118/1.1357457 ↗
5. Bhatia, N., Tisseur, D., Buyens, F., Létang, J.M.: Scattering correction using continuously thickness-adapted kernels. NDT & E Int. 78, 52–60 (2016). https://doi.org/10.1016/j.ndteint.2015.11.00410.1016/j.ndteint.2015.11.004 ↗
6. Hsieh, J. (ed.): Computed Tomography: Principles, Design, Artifacts, and Recent Advances. SPIE Press, Bellingham (2015). https://doi.org/10.1117/3.219775610.1117/3.2197756 ↗
7. Tkaczyk, J.E., Du, Y., Walter, D.J., Wu, X., Li, J., Toth, T.: Simulation of CT dose and contrast-to-noise as function of bowtie shape. In: Yaffe, M.J., Flynn, M.J. (eds.) Physics of Medical Imaging. SPIE, Bellingham (2004). https://doi.org/10.1117/12.53516110.1117/12.535161 ↗
8. Kwan, A.L.C., Boone, J.M., Shah, N.: Evaluation of x-ray scatter properties in a dedicated cone-beam breast CT scanner. Med. Phys. 32(9), 2967–2975 (2005). https://doi.org/10.1118/1.195490810.1118/1.1954908 ↗
9. Graham, S.A., Moseley, D.J., Siewerdsen, J.H., Jaffray, D.A.: Compensators for dose and scatter management in cone-beam computed tomography. Med. Phys. 34(7), 2691–2703 (2007). https://doi.org/10.1118/1.274046610.1118/1.2740466 ↗
10. Mail, N., Moseley, D.J., Siewerdsen, J.H., Jaffray, D.A.: The influence of bowtie filtration on cone-beam CT image quality. Med. Phys. 36(1), 22–32 (2009). https://doi.org/10.1118/1.301747010.1118/1.3017470 ↗
11. Liu, R., Zhang, S., Zhao, T., O’Sullivan, J.A., Williamson, J.F., Webb, T., Porras-Chaverri, M., Whiting, B.: Impact of bowtie filter and detector collimation on multislice CT scatter profiles: a simulation study. Med. Phys. 48(2), 852–870 (2020). https://doi.org/10.1002/mp.1465210.1002/mp.14652 ↗
12. Blessing, M., Bhagwat, M.S., Lyatskaya, Y., Bellon, J.R., Hesser, J., Zygmanski, P.: Kilovoltage beam model for flat panel imaging system with bow-tie filter for scatter prediction and correction. Physica Med. 28(2), 134–143 (2012). https://doi.org/10.1016/j.ejmp.2011.04.00110.1016/j.ejmp.2011.04.001 ↗
13. Persliden, J., Carlsson, G.A.: Scatter rejection by air gaps in diagnostic radiology. Calculations using a Monte Carlo collision density method and consideration of molecular interference in coherent scattering. Phys. Med. Biol. 42(1), 155–175 (1997). https://doi.org/10.1088/0031-9155/42/1/01110.1088/0031-9155/42/1/011 ↗
14. Zhu, L., Bennett, N.R., Fahrig, R.: Scatter correction method for X-ray CT using primary modulation: theory and preliminary results. IEEE Trans. Med. Imaging 25(12), 1573–1587 (2006). https://doi.org/10.1109/tmi.2006.88463610.1109/tmi.2006.884636 ↗
15. Alexeev, T., Kavanagh, B., Miften, M., Altunbas, C.: Two-dimensional antiscatter grid: a novel scatter rejection device for cone-beam computed tomography. Med. Phys. 45(2), 529–534 (2018). https://doi.org/10.1002/mp.1272410.1002/mp.12724 ↗
16. Ning, R., Tang, X., Conover, D.: X-ray scatter correction algorithm for cone beam CT imaging. Med. Phys. 31(5), 1195–1202 (2004). https://doi.org/10.1118/1.171147510.1118/1.1711475 ↗
17. Peterzol, A., Létang, J.M., Babot, D.: A beam stop based correction procedure for high spatial frequency scatter in industrial cone-beam X-ray CT. Nucl. Instrum. Methods Phys. Res. Sect. B 266(18), 4042–4054 (2008). https://doi.org/10.1016/j.nimb.2008.07.00510.1016/j.nimb.2008.07.005 ↗
18. Schörner, K., Goldammer, M., Stephan, J.: Comparison between beam-stop and beam-hole array scatter correction techniques for industrial X-ray cone-beam CT. Nucl. Instrum. Methods Phys. Res. Sect. B 269(3), 292–299 (2011). https://doi.org/10.1016/j.nimb.2010.11.05310.1016/j.nimb.2010.11.053 ↗
19. Rührnschopf, E.P., Klingenbeck, K.: A general framework and review of scatter correction methods in x-ray cone-beam computerized tomography. Part 1: scatter compensation approaches. Med. Phys. 38(7), 4296–4311 (2011). https://doi.org/10.1118/1.359903310.1118/1.3599033 ↗
20. Floyd, C.E., Jaszczak, R.J., Coleman, R.E.: Inverse Monte Carlo: a unified reconstruction algorithm for SPECT. IEEE Trans. Nucl. Sci. 32(1), 779–785 (1985). https://doi.org/10.1109/tns.1985.433694010.1109/tns.1985.4336940 ↗
21. National Research Council of Canada. Metrology Research Centre. Ionizing Radiation Standards: EGSnrc: Software for Monte Carlo Simulation of Ionizing Radiation. National Research Council of Canada (2000). https://doi.org/10.4224/4000130310.4224/40001303 ↗
22. Bundesanstalt für Materialforschung und -prüfung (BAM): aRTist—analytical RT inspection simulation tool (2019). http://www.artist.bam.de/
23. Jarry, G., Graham, S.A., Moseley, D.J., Jaffray, D.J., Siewerdsen, J.H., Verhaegen, F.: Characterization of scattered radiation in kV CBCT images using Monte Carlo simulations. Med. Phys. 33(11), 4320–4329 (2006). https://doi.org/10.1118/1.235832410.1118/1.2358324 ↗
24. Zbijewski, W., Beekman, F.J.: Efficient Monte Carlo based scatter artifact reduction in cone-beam micro-CT. IEEE Trans. Med. Imaging 25(7), 817–827 (2006). https://doi.org/10.1109/tmi.2006.87232810.1109/tmi.2006.872328 ↗
25. Poludniowski, G., Evans, P.M., Hansen, V.N., Webb, S.: An efficient Monte Carlo-based algorithm for scatter correction in keV cone-beam CT. Phys. Med. Biol. 54(12), 3847–3864 (2009). https://doi.org/10.1088/0031-9155/54/12/01610.1088/0031-9155/54/12/016 ↗
26. Rinkel, J., Gerfault, L., Estève, F., Dinten, J.-M.: A new method for x-ray scatter correction: first assessment on a cone-beam CT experimental setup. Phys. Med. Biol. 52(15), 4633–4652 (2007). https://doi.org/10.1088/0031-9155/52/15/01810.1088/0031-9155/52/15/018 ↗
27. Yao, W., Leszczynski, K.W.: An analytical approach to estimating the first order x-ray scatter in heterogeneous medium. Med. Phys. 36(7), 3145–3156 (2009). https://doi.org/10.1118/1.315211410.1118/1.3152114 ↗
28. Swindell, W., Evans, P.M.: Scattered radiation in portal images: a Monte Carlo simulation and a simple physical model. Med. Phys. 23(1), 63–73 (1996). https://doi.org/10.1118/1.59779210.1118/1.597792 ↗
29. Hansen, V.N., Swindell, W., Evans, P.M.: Extraction of primary signal from EPIDs using only forward convolution. Med. Phys. 24(9), 1477–1484 (1997). https://doi.org/10.1118/1.59803610.1118/1.598036 ↗
30. Ohnesorge, B., Flohr, T., Klingenbeck-Regn, K.: Efficient object scatter correction algorithm for third and fourth generation CT scanners. Eur. Radiol. 9(3), 563–569 (1999)10.1007/s003300050710 ↗
31. Star-Lack, J., Sun, M., Kaestner, A., Hassanein, R., Virshup, G., Berkus, T., Oelhafen, M.: Efficient scatter correction using asymmetric kernels. In: Samei, E., Hsieh, J. (eds.) Physics of Medical Imaging. SPIE, Bellingham, WA, USA (2009). https://doi.org/10.1117/12.81157810.1117/12.811578 ↗
32. Sun, M., Star-Lack, J.M.: Improved scatter correction using adaptive scatter kernel superposition. Phys. Med. Biol. 55(22), 6695–6720 (2010). https://doi.org/10.1088/0031-9155/55/22/00710.1088/0031-9155/55/22/007 ↗
33. Meyer, M., Kalender, W.A., Kyriakou, Y.: A fast and pragmatic approach for scatter correction in flat-detector CT using elliptic modeling and iterative optimization. Phys. Med. Biol. 55(1), 99–120 (2009). https://doi.org/10.1088/0031-9155/55/1/00710.1088/0031-9155/55/1/007 ↗
34. Lo, J.Y., Floyd, C.E., Baker, J.A., Ravin, C.E.: An artificial neural network for estimating scatter exposures in portable chest radiography. Med. Phys. 20(4), 965–973 (1993). https://doi.org/10.1118/1.59697810.1118/1.596978 ↗
35. Hansen, D.C., Landry, G., Kamp, F., Li, M., Belka, C., Parodi, K., Kurz, C.: Scatternet: a convolutional neural network for cone-beam CT intensity correction. Med. Phys. 45(11), 4916–4926 (2018). https://doi.org/10.1002/mp.1317510.1002/mp.13175 ↗
36. Lee, H., Lee, J.: A deep learning-based scatter correction of simulated X-ray images. Electronics 8(9), 944 (2019). https://doi.org/10.3390/electronics809094410.3390/electronics8090944 ↗
37. Maier, J., Sawall, S., Knaup, M., Kachelrieß, M.: Deep scatter estimation (DSE): accurate real-time scatter estimation for X-ray CT using a deep convolutional neural network. J. Nondestr. Eval. 37(3), 57 (2018). https://doi.org/10.1007/s10921-018-0507-z10.1007/s10921-018-0507-z ↗
38. Maier, J., Eulig, E., Vöth, T., Knaup, M., Kuntz, J., Sawall, S., Kachelrieß, M.: Real-time scatter estimation for medical CT using the deep scatter estimation: method and robustness analysis with respect to different anatomies, dose levels, tube voltages, and data truncation. Med. Phys. 46(1), 238–249 (2018). https://doi.org/10.1002/mp.1327410.1002/mp.13274 ↗
39. Nomura, Y., Xu, Q., Shirato, H., Shimizu, S., Xing, L.: Projection-domain scatter correction for cone beam computed tomography using a residual convolutional neural network. Med. Phys. 46(7), 3142–3155 (2019). https://doi.org/10.1002/mp.1358310.1002/mp.13583 ↗
40. Jiang, Y., Yang, C., Yang, P., Hu, X., Luo, C., Xue, Y., Xu, L., Hu, X., Zhang, L., Wang, J., Sheng, K., Niu, T.: Scatter correction of cone-beam CT using a deep residual convolution neural network (DRCNN). Phys. Med. Biol. 64(14), 145003 (2019). https://doi.org/10.1088/1361-6560/ab23a610.1088/1361-6560/ab23a6 ↗
41. Berker, Y., Maier, J., Kachelries, M.: Deep scatter estimation in PET: fast scatter correction using a convolutional neural network. In: IEEE Nuclear Science Symposium and Medical Imaging Conference Proceedings (NSS/MIC). IEEE, New York, NY, USA (2018). https://doi.org/10.1109/nssmic.2018.882459410.1109/nssmic.2018.8824594 ↗
42. Alsaffar, A., KießS, Sun, K., Simon, S.: Computational scatter correction in near real-time with a fast Monte Carlo photon transport model for high-resolution flat-panel CT. J. Real-Time Image Proc. 19(6), 1063–1079 (2022). https://doi.org/10.1007/s11554-022-01247-710.1007/s11554-022-01247-7 ↗
43. Dremel, K., Fuchs, T.: Scatter simulation and correction in computed tomography: a reconstruction-integrated approach modelling the forward projection. NDT & E Int. 86, 132–139 (2017). https://doi.org/10.1016/j.ndteint.2016.12.00210.1016/j.ndteint.2016.12.002 ↗
44. Wang, A., Maslowski, A., Wareing, T., Star-Lack, J., Schmidt, T.G.: A fast, linear Boltzmann transport equation solver for computed tomography dose calculation (Acuros CTD). Med. Phys. 46(2), 925–933 (2018). https://doi.org/10.1002/mp.1330510.1002/mp.13305 ↗
45. Principi, S., Wang, A., Maslowski, A., Wareing, T., Jordan, P., Schmidt, T.G.: Deterministic linear Boltzmann transport equation solver for patient-specific CT dose estimation: comparison against a Monte Carlo benchmark for realistic scanner configurations and patient models. Med. Phys. 47(12), 6470–6483 (2020). https://doi.org/10.1002/mp.1449410.1002/mp.14494 ↗
46. Baldi, P.: Autoencoders, unsupervised learning, and deep architectures. In: Proceedings of ICML Workshop on Unsupervised and Transfer Learning, pp. 37–49 (2012). JMLR Workshop and Conference Proceedings
47. Ronneberger, O., Fischer, P., Brox, T.: U-NET: convolutional networks for biomedical image segmentation. In: Proceedings of the International Conference on Medical Image Computing and Computer-assisted Intervention, pp. 234–241 (2015). Springer, Berlin10.1007/978-3-319-24574-4_28 ↗
48. Paszke, A., Gross, S., Massa, F., Lerer, A., Bradbury, J., Chanan, G., Killeen, T., Lin, Z., Gimelshein, N., Antiga, L., Desmaison, A., Kopf, A., Yang, E., DeVito, Z., Raison, M., Tejani, A., Chilamkurthy, S., Steiner, B., Fang, L., Bai, J., Chintala, S.: PyTorch: an imperative style, high-performance deep learning library. In: Advances in Neural Information Processing Systems, pp. 8024–8035 (2019)
49. Kingma, D.P., Ba, J.: Adam: A method for stochastic optimization (2014). arXiv:1412.6980