journal article·2018-09Open access

Deep Scatter Estimation (DSE): Accurate Real-Time Scatter Estimation for X-Ray CT Using a Deep Convolutional Neural Network

Joscha Maier, Stefan Sawall, Michael Knaup, Marc Kachelrieß

Abstract

X-ray scatter is a major cause of image quality degradation in dimensional CT. Especially, in case of highly attenuating components scatter-to-primary ratios may easily be higher than 1. The corresponding artifacts which appear as cupping or dark streaks in the CT reconstruction may impair a metrological assessment. Therefore, an appropriate scatter correction is crucial. Thereby, the gold standard is to predict the scatter distribution using a Monte Carlo (MC) code and subtract the corresponding scatter estimate from the measured raw data. MC, however, is too slow to be used routinely. To correct for scatter in real-time, we developed the deep scatter estimation (DSE). It uses a deep convolutional neural network which is trained to reproduce the output of MC simulations using only the acquired projection data as input. Once trained, DSE can be applied in real-time. The present study demonstrates the potential of the proposed approach using simulations and measurements. In both cases the DSE yields highly accurate scatter estimates that differ by< 3% from our MC scatter predictions. Further, DSE clearly outperforms kernel-based scatter estimation techniques and hybrid approaches, as they are in use today.
Source: openalex_inverted_index · Quality: baseline

Research context

OpenAlex topics: Advanced X-ray and CT Imaging · Medical Imaging Techniques and Applications · Digital Radiography and Breast Imaging

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (39)

Show source-provided references
1. Baer, M., Kachelrieß, M.: Hybrid scatter correction for CT imaging. Phys. Med. Biol. 57(21), 6849–6867 (2012). https://doi.org/10.1088/0031-9155/57/21/684910.1088/0031-9155/57/21/6849 ↗
2. Bhatia, N., Tisseur, D., Buyens, F., Létang, J.M.: Scattering correction using continuously thickness-adapted kernels. NDTE Int. 78, 52–60 (2016). https://doi.org/10.1016/j.ndteint.2015.11.00410.1016/j.ndteint.2015.11.004 ↗
3. Bier, B., Berger, M., Maier, A., Kachelrieß, M., Ritschl, L., Müller, K., J.H, Choi, R., Fahrig: Scatter correction using a primary modulator on a clinical angiography C-arm CT system. Med. Phys. 44(9), e125–e137 (2017). https://doi.org/10.1002/mp.1209410.1002/mp.12094 ↗
4. Cullen, D.E., Hubbell, J.H., Kissel, L.: EPDL97: The Evaluated Photon Data Library, ’97 Version. Tech. Rep. (1997)10.2172/295438 ↗
5. Feldkamp, L.A., Davis, L.C., Kress, J.W.: Practical cone-beam algorithm. J. Optical Soc. Am. A 1(6), 612–619 (1984). https://doi.org/10.1364/JOSAA.1.00061210.1364/JOSAA.1.000612 ↗
6. Glover, G.H.: Compton scatter effects in CT reconstructions. Med. Phys. 9(6), 860–867 (1982). https://doi.org/10.1118/1.59519710.1118/1.595197 ↗
7. Grimmer, R., Fahrig, R., Hinshaw, W., Gao, H., Kachelrieß, M.: Empirical cupping correction for CT scanners with primary modulation (ECCP). Med. Phys. 39(2), 825–831 (2012). https://doi.org/10.1118/1.367618010.1118/1.3676180 ↗
8. Hsieh, S.: Estimating scatter in cone beam CT with striped ratio grids: a preliminary investigation. Med. Phys. 43(9), 5084–5092 (2016). https://doi.org/10.1118/1.496063110.1118/1.4960631 ↗
9. Joseph, P.M., Spital, R.D.: The effects of scatter in X-ray computed tomography. Med. Phys. 9(4), 464–472 (1982). https://doi.org/10.1118/1.59511110.1118/1.595111 ↗
10. Leinweber, C., Maier, J., Kachelrieß, M.: X-ray spectrum estimation for accurate attenuation simulation. Med. Phys. 44(12), 6183–6194 (2017). https://doi.org/10.1002/mp.1260710.1002/mp.12607 ↗
11. Li, H., Mohan, R., Zhu, X.R.: Scatter kernel estimation with an edge-spread function method for cone-beam computed tomography imaging. Phys. Med. Biol. 53(23), 6729–6748 (2008). https://doi.org/10.1088/0031-9155/53/23/00610.1088/0031-9155/53/23/006 ↗
12. Lifton, J.J., Carmignato, S.: Simulating the influence of scatter and beam hardening in dimensional computed tomography. Meas. Sci. Technol. 28(10), 104,001 (2017). https://doi.org/10.1088/1361-6501/aa80b210.1088/1361-6501/aa80b2 ↗
13. Lifton, J.J., Malcolm, A.A., McBride, J.W.: An experimental study on the influence of scatter and beam hardening in X-ray CT for dimensional metrology. Meas. Sci. Technol. 27(1), 015,007 (2016). https://doi.org/10.1088/0957-0233/27/1/01500710.1088/0957-0233/27/1/015007 ↗
14. Lo, J.Y., Floyd, C.E., Baker, J.A., Ravin, C.E.: An artificial neural network for estimating scatter exposures in portable chest radiography. Med. Phys. 20(4), 965–973 (1993). https://doi.org/10.1118/1.59697810.1118/1.596978 ↗
15. Love, L.A., Kruger, R.A.: Scatter estimation for a digital radiographic system using convolution filtering. Med. Phys. 14(2), 178–185 (1987). https://doi.org/10.1118/1.59612610.1118/1.596126 ↗
16. Maier, J., Leinweber, C., Sawall, S., Stoschus, H., Ballach, F., Müller, T., Hammer, M., Christoph, R., Kachelrieß, M.: Simulation-based artifact correction (SBAC) for metrological computed tomography. Meas. Sci. Technol. 28(6), 065,011 (2017). https://doi.org/10.1088/1361-6501/aa666e10.1088/1361-6501/aa666e ↗
17. Meyer, E., Maaß, C., Baer, M., Raupach, R., Schmidt, B., Kachelrieß, M.: Empirical scatter correction (esc): a new CT scatter correction method and its application to metal artifact reduction. In: IEEE Nuclear Science Symposuim & Medical Imaging Conference, pp. 2036–2041. IEEE (2010). https://doi.org/10.1109/NSSMIC.2010.587413410.1109/NSSMIC.2010.5874134 ↗
18. Meyer, M., Kalender, W.A., Kyriakou, Y.: A fast and pragmatic approach for scatter correction in flat-detector CT using elliptic modeling and iterative optimization. Phys. Med. Biol. 55(1), 99–120 (2010). https://doi.org/10.1088/0031-9155/55/1/00710.1088/0031-9155/55/1/007 ↗
19. Nelder, J.A., Mead, R.: A simplex method for function minimization. Comput. J. 7(4), 308–313 (1965). https://doi.org/10.1093/comjnl/7.4.30810.1093/comjnl/7.4.308 ↗
20. Ning, R., Tang, X., Conover, D.: X-ray scatter correction algorithm for cone beam CT imaging. Med. Phys. 31(5), 1195–1202 (2004). https://doi.org/10.1118/1.171147510.1118/1.1711475 ↗
21. Ohnesorge, B., Flohr, T., Klingenbeck-Regn, K.: Efficient object scatter correction algorithm for third and fourth generation CT scanners. Eur. Radiol. 9(3), 563–569 (1999). https://doi.org/10.1007/s00330005071010.1007/s003300050710 ↗
22. Poludniowski, G., Evans, P.M., Hansen, V.N., Webb, S.: An efficient Monte Carlo-based algorithm for scatter correction in keV cone-beam CT. Phys. Med. Biol. 54(12), 3847–3864 (2009). https://doi.org/10.1088/0031-9155/54/12/01610.1088/0031-9155/54/12/016 ↗
23. Rinkel, J., Gerfault, L., Estève, F., Dinten, J.M.: A new method for x-ray scatter correction: first assessment on a cone-beam CT experimental setup. Phys. Med. Biol. 52(15), 4633–4652 (2007). https://doi.org/10.1088/0031-9155/52/15/01810.1088/0031-9155/52/15/018 ↗
24. Ritschl, L., Fahrig, R., Knaup, M., Maier, J., Kachelrieß, M.: Robust primary modulation-based scatter estimation for cone-beam CT. Med. Phys. 42(1), 469–478 (2015). https://doi.org/10.1118/1.490326110.1118/1.4903261 ↗
25. Ronneberger, O., Fischer, P., Brox, T.: U-Net: Convolutional networks for biomedical image segmentation. In: miccai, pp. 234–241 (2015)10.1007/978-3-319-24574-4_28 ↗
26. Rührnschopf, E.P., Klingenbeck, K.: A general framework and review of scatter correction methods in X-ray cone-beam computerized tomography. Part 1: scatter compensation approaches. Med. Phys. 38(7), 4296–4311 (2011). https://doi.org/10.1118/1.359903310.1118/1.3599033 ↗
27. Rührnschopf, E.P., Klingenbeck, K.: A general framework and review of scatter correction methods in cone beam CT. Part 2: scatter estimation approaches. Med. Phys. 38(9), 5186–5199 (2011). https://doi.org/10.1118/1.358914010.1118/1.3589140 ↗
28. Schorner, K., Goldammer, M., Stierstorfer, K., Stephan, J., Boni, P.: Scatter correction method by temporal primary modulation in X-ray CT. IEEE Trans. Nucl. Sci. 59(6), 3278–3285 (2012). https://doi.org/10.1109/TNS.2012.221812710.1109/TNS.2012.2218127 ↗
29. Siewerdsen, J.H., Jaffray, D.A.: Cone-beam computed tomography with a flat-panel imager: magnitude and effects of X-ray scatter. Med. Phys. 28(2), 220–231 (2001). https://doi.org/10.1118/1.133987910.1118/1.1339879 ↗
30. Sossin, A., Rebuffel, V., Tabary, J., Létang, J.M., Freud, N., Verger, L.: A novel scatter separation method for multi-energy X-ray imaging. Phys. Med. Biol. 61(12), 4711–4728 (2016). https://doi.org/10.1088/0031-9155/61/12/471110.1088/0031-9155/61/12/4711 ↗
31. Spies, L., Ebert, M., Groh, B.A., Hesse, B.M., Bortfeld, T.: Correction of scatter in megavoltage cone-beam CT. Phys. Med. Biol. 46(3), 821–833 (2001). https://doi.org/10.1088/0031-9155/46/3/31610.1088/0031-9155/46/3/316 ↗
32. Star-Lack, J., Sun, M., Kaestner, A., Hassanein, R., Virshup, G., Berkus, T., Oelhafen, M.: Efficient scatter correction using asymmetric kernels. In: Samei, E., Hsieh, J. (eds.) SPIE Medical Imaging Conference, vol. 7258, p. 72581Z (2009). https://doi.org/10.1117/12.81157810.1117/12.811578 ↗
33. Sun, M., Star-Lack, J.M.: Improved scatter correction using adaptive scatter kernel superposition. Phys. Med. Biol. 55(22), 6695–6720 (2010). https://doi.org/10.1088/0031-9155/55/22/00710.1088/0031-9155/55/22/007 ↗
34. Swindell, W., Evans, P.M.: Scattered radiation in portal images: a Monte Carlo simulation and a simple physical model. Med. Phys. 23(1), 63–73 (1996). https://doi.org/10.1118/1.59779210.1118/1.597792 ↗
35. Tucker, D.M., Barnes, G.T., Chakraborty, D.P.: Semiempirical model for generating tungsten target X-ray spectra. Med. Phys. 18(2), 211–218 (1991). https://doi.org/10.1118/1.59670910.1118/1.596709 ↗
36. Yao, W., Leszczynski, K.W.: An analytical approach to estimating the first order scatter in heterogeneous medium. II. A practical application. Med. Phys. 36(7), 3157–3167 (2009). https://doi.org/10.1118/1.315211510.1118/1.3152115 ↗
37. Zbijewski, W., Beekman, F.: Fast scatter estimation for cone-beam X-ray CT by combined Monte Carlo tracking and Richardson-Lucy fitting. In: IEEE MIC Symposium Conference., vol. 5, pp. 2774–2777. IEEE (2004). https://doi.org/10.1109/NSSMIC.2004.146626410.1109/NSSMIC.2004.1466264 ↗
38. Zhu, L., Bennett, N.R., Fahrig, R.: Scatter correction method for X-ray CT using primary modulation: theory and preliminary results. IEEE Trans. Med. Imaging 25(12), 1573–1587 (2006). https://doi.org/10.1109/TMI.2006.88463610.1109/TMI.2006.884636 ↗
39. Zhu, L., Strobel, N., Fahrig, R.: X-ray scatter correction for cone-beam CT using moving blocker array. In: Flynn, M.J. (ed.) SPIE Medical Imaging Conference, vol. 5745, p. 251 (2005). https://doi.org/10.1117/12.59469910.1117/12.594699 ↗