Surface analysis of Fe–Co–Nb–Cu–B metallic glasses using a miniaturised Mössbauer spectrometer ‘MIMOS’
Source abstract
Abstract
Abstract A miniaturised Mossbauer spectrometer ‘MIMOS’ operating in a backscattering geometry using 6.4 keV X-ray Mossbauer radiation and 14.4 keV backscattered γ -ray Mossbauer radiation has been used to obtain information about the depth-selective surface crystallization of thermally annealed Fe 58 Co 25 Nb 7 Cu 1 B 9 and Fe 60 Co 25 Nb 6 Cu 1 B 8 alloys. Crystallization of the specimens starts after annealing at 390 °C. An increase of the annealing temperature increases the hyperfine field showing the increase of Co-content in the crystalline phase from 2% to 10%. The relative abundance of Fe in the crystalline phase ranges between 8% and 42%. An increase in the annealing temperature enhances the order in the crystalline phase and randomises the atomic spins, which are preferentially aligned within the ribbon plane in the as-cast state.
Source: openaire · Quality: candidate unverified · Source record ↗
Research context
OpenAlex topics: Metallic Glasses and Amorphous Alloys · Magnetic Properties and Applications · Magnetic Properties of Alloys
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (7)
Show source-provided references
1. Klingelhöfer · 1999 · 413
2. Kane · Hyper. Inter. · 2002 · 144&145 · 27310.1023/A:1025496225898 ↗
3. Grabias · Nanostruct. Mater. · 1999 · 12 · 89910.1016/S0965-9773(99)00262-7 ↗
4. Herzer · J. Magn. Magn. Mater. · 1986 · 62 · 14310.1016/0304-8853(86)90136-8 ↗
5. www.//mars.jpl.nasa.gov/mer/.
6. Brand · Nucl. Instrum. Methods. B · 1987 · 28 · 39810.1016/0168-583X(87)90182-0 ↗
7. Johnson · Phys. Rev. Lett. · 1961 · 6 · 45010.1103/PhysRevLett.6.450 ↗