Use of an electrical aerosol detector (EAD) for nanoparticle size distribution measurement
Source abstract
Abstract
Recently, Nanoparticle Surface Area Monitor (NSAM, TSI model 3550) and EAD (EAD, TSI Model 3070A) have been commercially available to measure the integral parameters (i.e., total particle surface area and total particle length) of nanoparticles. By comparison, the configuration of the EAD or NSAM is similar to that of electrical mobility analyzer of the early generation for particle size distribution measurement. It is therefore possible to use the EAD or NSAM as a particle sizer. To realize the objective of using the EAD as a sizer, we characterized the average electrical charges of monodisperse particles passing through the EAD particle charger and ion trap set at voltages ranging from 20 to 2500 V. The average charge data collected at different ion-trap voltages were then summarized by the empirical correlation using the parameter of Zp*V, where Zp is the particle electrical mobility and V is the ion-trap voltage. A data-reduction scheme was further proposed to retrieve the size distribution of sampled particles from the EAD readout at different ion-trap voltages. In the scheme, the functional format of each mode in a number size distribution of particles was assumed as log-normal, but the number of modes in an entire size distribution is not limited. A criterion was used to best fit the simulated EAD readouts with experimental ones by varying the count median diameter (CMD), geometric standard deviation (σg), and total particle number (Nt) of each mode in a particle size distribution. Experiments were performed to verify the proposed scheme.
Source: openaire · Quality: candidate unverified · Source record ↗
Research context
OpenAlex topics: Coagulation and Flocculation Studies · Air Quality and Health Impacts · Air Quality Monitoring and Forecasting
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (16)
Show source-provided references
1. Donaldson K, Li XY, MacNee W (1998) Ultrafine (nanometer) particle mediated lung injury. J Aerosol Sci 29(5–6):553–56010.1016/S0021-8502(97)00464-3 ↗
2. Fissan H, Trampe A, Neunman S, Pui DYH, Shin WG (2007) Rationale and principle of an instrument measuring lung deposition area. J Nanoparticle Res 9:53–5910.1007/s11051-006-9156-8 ↗
3. Gleiter H (1989) Nanocrystalline materials. Prog Mater Sci 33:223–31510.1016/0079-6425(89)90001-7 ↗
4. Hinds W (1999) Aerosol technology: properties, behavior, and measurement of airborne particles. Wiley-Interscience, New York
5. Keskinen J, Pietarinen K, Lehtimaki M (1992) Electrical Low Pressure Impactor. J Aerosol Sci 23:35310.1016/0021-8502(92)90004-F ↗
6. Knutson EO, Whitby KT (1975) Aerosol classification by electric mobility: apparatus, theory, and applications. J Aerosol Sci 6:443–45110.1016/0021-8502(75)90060-9 ↗
7. Martin SJ, Granstaff VE, Frye GC (1991) Characterisation of a quartz crystal microbalance with simultaneous mass and liquid loading. Anal Chem 63:2272–228110.1021/ac00020a015 ↗
8. Oberdörster G (1996) Significance of particle parameters in the evaluation of exposure-dose–response relationships of inhaled particles. Part Sci Technol 14(2):135–15110.1080/02726359608906690 ↗
9. Oberdörster G, Oberdörster E, Oberdörster J (2005) Invited review: nanotechnology: an emerging discipline evolving from studies of ultrafine particles. Environ Health Perspect 113(7):823–83910.1289/ehp.7339 ↗
10. Park D, Kim S, An M, Hwang J (2007a) Real-time measurement of submicron aerosol particles having a log-normal size distribution by simultaneously using unipolar diffusion charger and unipolar field charger. J Aerosol Sci 38:1240–124510.1016/j.jaerosci.2007.09.002 ↗
11. Park D, An M, Hwang J (2007b) Development and performance test of a unipolar diffusion charger for real-time measurements of submicron aerosol particles having a log-normal size distribution. J Aerosol Sci 38(4):420–43010.1016/j.jaerosci.2007.01.003 ↗
12. Patashnick H, Rupprecht EG (1991) Continuous PM–10 measurements using the tapered element oscillating microbalance. J Air Waste Manage Assoc 41:1079–108310.1080/10473289.1991.10466903 ↗
13. Scheibel HG, Porstendorfer J (1983) Generation of monodisperse Ag and NaCl aerosol with particle diameters between 2 and 300 nm. J Aerosol Sci 14:113–12610.1016/0021-8502(83)90035-6 ↗
14. Stoltzenburg MR, McMurry PH (1991) An ultrafine aerosol condensation nucleus counter. Aerosol Sci Technol 14:48–6510.1080/02786829108959470 ↗
15. Wilson WE, Han HS, Stanek J, Turner J, Chen DR, Pui DYH (2004) Use of electrical aerosol detector as an indicator for the total particle surface area deposited in the lung. Symposium on air quality measurement methods and technology sponsored by the Air and Waste Management Association. Research Triangle Park, NC. Paper #37
16. Woo KS, Chen DR, Pui DYH, Wilson WE (2001) Use of continuous measurements of integral aerosol parameters to estimate particle surface area. Aerosol Sci Technol 34:57–6510.1080/02786820117549 ↗