Preparation of cholesterol oxidase nanoparticles and their application in amperometric determination of cholesterol
Source abstract
Abstract
The nanoparticle (NP) aggregates of commercial cholesterol oxidase (ChOx) were prepared by desolvation method. The formation and characterization of ChOxNP aggregates were studied by transmission electron microscopy and scanning electron microscopy. NP aggregates were more stable, active and had a higher shelf life than that of free enzyme. An amperometric cholesterol biosensor was constructed by immobilizing ChOxNPs onto Au electrode. The biosensor showed optimum response within 8 s at pH 6.0 and 35 °C, when polarized at +0.27 V versus Ag/AgCl. The biosensor possesses high sensitivity and measures cholesterol concentrations as low as 1.56 mg/dl. The working linear range was 12.5–700 mg/dl for cholesterol. The biosensor was evaluated and employed for measurement of total cholesterol in human serum. The enzyme electrode lost 50 % of its initial activity during its regular use for 180 times over a period of 90 days when stored in 0.1 M sodium phosphate buffer, pH 7.0 at 4 °C.
Source: openaire · Quality: candidate unverified · Source record ↗
Research context
OpenAlex topics: Electrochemical sensors and biosensors · Analytical Chemistry and Sensors · Electrochemical Analysis and Applications
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (20)
Show source-provided references
1. Allain CC, Poon LS, Chan CSG, Richmond W, Fu PC (1974) Enzymatic determination of serum total cholesterol. Chin Chem 20(4):470–475
2. Arvind SSJ, Baby TT, Arockiadoss T, Rakhi RB, Ramprabhu S (2011) Cholesterol biosensor based on gold nanoparticles decorated functionalized graphene nanoplates. Thin solid films10.1016/j.tsf.2011.03.032 ↗
3. Bhatia D, Suman, Pundir CS (2005) Preparation of a reusable enzyme strip for determination of serum cholesterol. Ind J Biotechnol 4:471–475
4. Boller T, Meier C, Menzler S (2002) EUPERGIT oxirane acrylic beads: how to make enzymes fit for biocatalysis. Org Process Res Dev 6:509–51910.1021/op015506w ↗
5. Brahim S, Narinesingh D, Guiseppi-Elie A (2001) Amperometric determination of cholesterol in serum using a biosensor of cholesterol oxidase contained within a polypyrrole-hydrogel membrane. Anal Chim Acta 448:27–3610.1016/S0003-2670(01)01321-6 ↗
6. Cao L, Van-Rantwijk F, Sheldon RA (2000) Cross-linked enzymes aggregates: a simple and effective method for the immobilization of penicillin acylase. Org Lett 2:1361–136410.1021/ol005593x ↗
7. Cao L, Van-Langen L, Sheldon RA (2003) Immobilized enzymes: carrier bound or carrier free. Curr Opin Biotechnol 14:387–39410.1016/S0958-1669(03)00096-X ↗
8. Chauhan N, Narang J, Pundir CS (2010) Amperometric determination of serum cholesterol with pencil graphite rod. Am J Anal Chem 2:41–4610.4236/ajac.2010.12006 ↗
9. Jianping L, Peng T, Peng Y (2006) A cholesterol biosensors based on entrapment of cholesterol oxidase in a silicic sol–gel matrix at a prussian blue modified. Electroanalysis 15:1031–1033
10. Kundu N, Yadav S, Pundir CS (2012) Preparation and characterization of glucose oxidase nanoparticles and their application in dissolve oxygen metric determination of serum glucose. J Nanosci Nanotechnol 12:1–710.1166/jnn.2012.5193 ↗
11. Li YF, Liu ZM, Liu YL, Yang YH, Shen GL, Yu RQ (2006) A mediator-free phenol biosensor based on immobilizing tyrosinase to ZnO nanoparticles. Anal Biochem 349:33–4010.1016/j.ab.2005.11.017 ↗
12. Liu G, Lin Y, Ostatna V, Wang J (2005) Enzyme nanoparticles-based electronic biosensor. Chem Commun 27:3481–348310.1039/b504943a ↗
13. Nien PC, Chen PY, Ho KC (2009) Fabricating an amperometric cholesterol biosensor by a covalent linkage between poly(3-thiopheneacetic acid) and cholesterol oxidase. Sensors 9:1794–180610.3390/s90301794 ↗
14. Norouzi P, Rafiei-Sarmazdeh Z, Faridbod F, Adibi M, Ganjali MR (2010) Er3+ carbon paste electrode based on new nano-composite. Int J Electrochem Soc 5:367–37610.1016/S1452-3981(23)15291-6 ↗
15. Schoevaart R, Wolbers MW, Golubovic M, Ottens M, Kieboom APG, Van-Rantwijk F (2004) Preparation, optimization, and structures of cross-linked enzyme aggregates (CLEAs). Biotechnol Bioeng 87:754–76210.1002/bit.20184 ↗
16. Shah S, Sharma A, Gupta MN (2006) Preparation of cross-linked enzyme aggregates by using bovine serum albumin as a proteic feeder. Anal Biochem 351:207–21310.1016/j.ab.2006.01.028 ↗
17. Sheldon RA (2011) Cross-linked enzyme aggregates as industrial biocatalysts. Org Process Res Dev 15:213–22310.1021/op100289f ↗
18. Suman, Pundir CS (2003) Co-immobilization of cholesterol esterase, cholesterol oxidase and peroxidase onto alkylamine glass beads for measurement of total cholesterol in serum. Curr Appl Phys 3:129–13310.1016/S1567-1739(02)00178-5 ↗
19. Umar A, Rahman MM, Vaseem M, Hahn YB (2009) Ultra-sensitive cholesterol biosensor based on low-temperature grown ZnO nanoparticles. Electrochem Commun 11:118–12110.1016/j.elecom.2008.10.046 ↗
20. Van-Langon LM, Selassa RP, Van-Rantwijk F, Sheldon RA (2005) Cross-linked aggregates of R-oxynitrilase: a stable, recyclable biocatalyst for enantioselective hydrocyanation. Org Lett 7:327–32910.1021/ol047647z ↗