Antibacterial Efficacy and Biocompatibility of HiPIMS-Ag Films for Prosthetic Application
Source abstract
Abstract
Implant-associated infections (IAIs) remain a major cause of orthopedic implant failure, motivating the development of surface coatings that deliver durable antibacterial activity without compromising host compatibility. Here, we deposit silver (Ag) thin films onto commercially pure titanium (Ti) using high power impulse magnetron sputtering (HiPIMS) and assess their antibacterial performance and osteoblast cytocompatibility. Film formation, morphology and crystallinity were characterized by electron microscopy and X-ray diffractometry, while interfacial integrity was probed using ASTM D3359 cross-cut and VDI 3198 Rockwell-C indentation. Antibacterial activity against Escherichia coli and Staphylococcus aureus was quantified by culture-based enumeration, and Ag+ release was measured by ICP-MS. HiPIMS enabled rapid formation of dense, continuous and crystalline Ag films with excellent adhesion. Even ultrathin coatings (~7 nm) produced strong antibacterial effects (activity value > 2.0) while releasing controllable trace Ag+ (ultimately 0.43 ppb/day), and osteoblast assays indicated no cytotoxicity under the tested conditions. The results show that HiPIMS-Ag achieves a favorable balance between antimicrobial efficacy and biocompatibility at low thickness, supporting its use as a robust antibacterial surface for Ti implants and providing a foundation for translation to device level and in vivo studies.
Source: crossref · Quality: baseline
Research context
OpenAlex topics: Bone Tissue Engineering Materials · Polymer Surface Interaction Studies · Metal and Thin Film Mechanics
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
Affiliations
Department of Materials Science and Engineering, Feng Chia University, Taichung 407102, Taiwan
Institute of Plasma, Feng Chia University, Taichung 407102, Taiwan
Functional Neurosurgery Division, Neurological Institute, Taichung Veterans General Hospital, Taichung 407219, Taiwan
Department of Post-Baccalaureate Medicine, National Chung Hsing University, Taichung 402202, Taiwan
Otolaryngology, Kuang Tien General Hospital, Taichung 433401, Taiwan
Funding
Taichung Veterans General Hospital
References (21)
Show source-provided references
1. Prosthetic-Joint Infections10.1056/NEJMra040181 ↗
2. Prosthetic Joint Infection10.1128/CMR.00111-13 ↗
3. Antibacterial Coatings on Titanium Implants10.1002/jbm.b.31463 ↗
4. Silver Nanoparticles as a New Generation of Antimicrobials10.1016/j.biotechadv.2008.09.002 ↗
5. Sui, J., Liu, S., Chen, M., and Zhang, H. (2022). Surface Bio-Functionalization of Anti-Bacterial Titanium Implants: A Review. Coatings, 12.10.3390/coatings12081125 ↗
6. High Power Impulse Magnetron Sputtering Discharge10.1116/1.3691832 ↗
7. An Introduction to Thin Film Processing Using High Power Impulse Magnetron Sputtering10.1557/jmr.2012.8 ↗
8. Microstructural Evolution during Film Growth10.1116/1.1601610 ↗
9. High Power Pulsed Magnetron Sputtering: A Review on Scientific and Engineering State of the Art10.1016/j.surfcoat.2009.11.013 ↗
10. Li, S.T. (2023). Preclinical Evaluation of High Power Impulse Magnetron Sputter Deposited Antibacterial Silver Film on Titanium Implant. [Master’s Thesis, Feng Chia University].
11. (2010). Standard Test Methods for Rating Adhesion by Tape Test (Standard No. ASTM D3359-02).
12. (1992). Characterisation of Adhesion/Cohesion of Coatings by Indentation Test (Standard No. VDI 3198).
13. (2010). Antibacterial Products—Test for Antibacterial Activity and Efficacy (Standard No. JIS Z 2801:2010).
14. Yang, X., Gao, P., Yang, Z., Zhu, J., Huang, F., and Ye, J. (2017). Optimizing Ultrathin Ag Films for High Performance Oxide-Metal-Oxide Flexible Transparent Electrodes through Surface Energy Modulation and Template-Stripping Procedures. Sci. Rep., 7.10.1038/srep44576 ↗
15. Ultrathin Sputter-Deposited Plasmonic Silver Nanostructures10.1039/C9NA00762H ↗
16. Quantification of silver ion release, in vitro cytotoxicity and antibacterial properties of nanostructured Ag-doped TiO2 coatings on stainless steel deposited by RF magnetron sputtering10.1016/j.vacuum.2011.06.011 ↗
17. Understanding long-term silver release from surface modified porous titanium implants10.1016/j.actbio.2017.05.048 ↗
18. Gudkov, S.V., Serov, D.A., Astashev, M.E., Semenova, A.A., and Lisitsyn, A.B. (2022). Ag2O nanoparticles as a candidate for antimicrobial compounds of the new generation. Pharmaceuticals, 15.10.3390/ph15080968 ↗
19. A pharmacological and toxicological profile of silver as an antimicrobial agent in medical devices
20. Barberi, J., and Spriano, S. (2021). Titanium and protein adsorption: An overview of mechanisms and effects of surface features. Materials, 14.10.3390/ma14071590 ↗
21. Antimicrobial particulate silver coatings on stainless steel implants for fracture management10.1016/j.msec.2012.02.020 ↗