Dynamic pressure measurements during pulsed gas injection in a quadrupole ion trap

D. Papanastasiou, L. Ding, E. Raptakis, I. Brookhouse, J. Cunningham, M. Robinson

Vacuum· Volume 81· Issue 4
Publisher page

Abstract

Abstract A quadrupole ion trap has been modified to perform dynamic pressure measurements during pulsed introduction of gases. A continuous electron beam is directed through the ion trap where the gas is ionized via electron impact. Ion and electron currents are monitored on the ring and end-cap electrodes, respectively. Dynamic pressure measurements in a region not accessible to a standard gauge are performed using a static quadrupole field. Characteristic current–voltage curves of the ion-trap gauge are presented and optimum operating conditions of the electron–ion optical system are identified in steady state conditions. The sensitivity of the ion–trap gauge is calibrated at these optimum conditions. In the pulsed gas mode ion and electron signals are measured simultaneously on a fast oscilloscope. The time constant of the circuit for the dynamic measurements is ∼129 μs and pressure variations of ∼10−4–1 Pa occur within 60–130 ms. The exponential decay of the ion signals is used to calculate pumping speeds for helium and argon gases. The distinctive advantages of pulsed gas injections over the use of static pressures in quadrupole ion-trap mass spectrometry are emphasized.
Source: openaire · Quality: candidate unverified · Source record ↗

Research context

OpenAlex topics: Mass Spectrometry Techniques and Applications · Spectroscopy and Laser Applications · Atomic and Molecular Physics

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (19)

Show source-provided references
1. Dehmelt · Adv At Mol Phys · 1967 · 3 · 5310.1016/S0065-2199(08)60170-0 ↗
2. Stafford · Int J Mass Spectrom Ion Process · 1984 · 60 · 8510.1016/0168-1176(84)80077-4 ↗
3. Douglas · J Am Soc Mass Spectrom · 1992 · 3 · 39810.1016/1044-0305(92)87067-9 ↗
4. Xu · Nucl Instr Meth Phys Res A · 1993 · 333 · 27410.1016/0168-9002(93)91166-K ↗
5. Jennings · Int J Mass Spectrom Ion Phys · 1968 · 1 · 22710.1016/0020-7381(68)85002-8 ↗
6. Haddon · J Am Chem Soc · 1968 · 90 · 474510.1021/ja01019a053 ↗
7. Danell · J Am Soc Mass Spectrom · 2003 · 14 · 109910.1016/S1044-0305(03)00404-5 ↗
8. Doroshenko · Anal Chem · 1996 · 68 · 46310.1021/ac950954l ↗
9. Vachet · J Am Soc Mass Spectrom · 1996 · 7 · 119410.1016/S1044-0305(96)00109-2 ↗
10. Ding · Proc SPIE, Int Soc Opt Eng · 1999 · 3777 · 144
11. Arad · Rev Sci Instrum · 1998 · 69 · 152910.1063/1.1148791 ↗
12. Kuznetsov · Instr Exp Technol · 2001 · 44 · 14010.1023/A:1017591725773 ↗
13. Dahl DA. Idaho National Engineering & Environmental Laboratory, Idaho Falls, ID, Version 7.0.1999.
14. Bayard · Rev Sci Instrum · 1950 · 21 · 57110.1063/1.1745653 ↗
15. Chambers A, Fitch RK, Halliday BS. Basic vacuum technology, 2nd ed., IoP, 1998, p. 93.10.1201/NOE0750304955 ↗
16. Tilford · J Vac Sci Technol · 1983 · 1 · 15210.1116/1.572063 ↗
17. Dushman · 1962
18. Active Ion Gauge Instruction Manual, BOC Edwards, West Sussex, UK.
19. Tilford · J Vac Sci Technol · 1985 · 3 · 54610.1116/1.572991 ↗