Mass spectrometry with a time resolution of a few milliseconds using a helium flushed probe system
Source abstract
Abstract
At pressures higher than approximately 5 mbar it is nowadays common practice to use a two-stage pressure reduction set-up, consisting of a capillary tube as a first pressure reduction stage and an orifice as a second pressure reduction stage. We have recently demonstrated, that by using a specially constructed helium flushed probe system, a quick transport of species extracted from a process gas to a quadrupole mass spectrometer can be achieved. In order to achieve an improved time resolution a careful redesign of the probe and mass spectrometer system has been performed. An experimental investigation using a very fast chemical decomposition process revealed a time resolution of a few milliseconds.