journal article·2019-05

Uncertainty analysis of the LIP vacuum standard for XHV range

Zhenhua Xi, Yongjun Cheng, Huzhong Zhang, Yali Li, Detian Li

Vacuum· Volume 163
Publisher page

Abstract

Abstract This work analyzes the uncertainty sources of the vacuum standard for extremely high vacuum range constructed at Lanzhou Institute of Physics. To establish XHV calibration with extended lower limit, meanwhile, decreasing the uncertainty at room temperature, the selective pumping technique and a new type of leak element have been adopted. For XHV range (10−10 Pa), the expanded uncertainty of the generated pressure is 7.2% (k = 2), and the total uncertainty of the generated pressure is mainly determined by the base pressure and the measurement process of the conductance of the leak element. Results show that, a relatively lower uncertainty can be obtained for the same throughput by use of the nanofluidic channel leak element which can generate molecular flow with atmospheric gas input.
Source: openaire · Quality: candidate unverified · Source record ↗

Research context

OpenAlex topics: Scientific Measurement and Uncertainty Evaluation · Advanced Sensor Technologies Research · Advanced Electrical Measurement Techniques

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (39)

Show source-provided references
1. Traceability to SI units for vacuum measurement in industrial applications10.1016/j.measurement.2011.10.036 ↗
2. Discharge-produced plasma extreme ultraviolet (EUV) source and ultra-high vacuum chamber for studying EUV-induced processes10.1088/0963-0252/24/3/035003 ↗
3. A. Krämer, M.C. Bellachioma, H. Kollmus, et al., The vacuum system of fair accelerator facility, Proceedings of EPAC 2006, Edinburgh, Scotland 1429-1431.
4. Berg
5. New fully automated primary standard for generating vacuum pressures between 10-10 Pa and 3×10-2 Pa with respect to residual pressure10.1088/0026-1394/36/6/2 ↗
6. Calibration of ultrahigh vacuum gauge from 10-9 Pa to 10-5 Pa by the two-stage flow-dividing system10.1016/j.vacuum.2011.06.012 ↗
7. INRIM continuous expansion system as high vacuum primary standard or gas pressure measurements below 9×10-2 Pa10.1016/j.vacuum.2009.04.002 ↗
8. Dynamic extension e A new principle for a vacuum primary standard for UHV-range (10-6-10-10 Pa) calibration of hot cathode ionisation gauges10.1016/j.vacuum.2011.03.019 ↗
9. Calibration of total pressure gauges in the UHV and XHV Regions10.3131/jvsj.37.678 ↗
10. Vacuum-calibration apparatus with pressure down to 10-10 Pa10.1116/1.3457934 ↗
11. Very low gas flow measurements for UHV/XHV and leak calibration10.1088/0026-1394/50/1/15 ↗
12. Residual gas survey of stainless steel 304 extreme high vacuum chamber with hot cathode ionization gauge10.1016/j.measurement.2008.02.006 ↗
13. Microfabricated silicon leak for sampling planetary atmospheres with a mass spectrometer10.1063/1.2748360 ↗
14. Focused-ion beam fabrication of nanometer orifices for leak detection10.1116/1.3243229 ↗
15. New leak element using sintered stainless steel filter for in-situ calibration of ionization gauges and quadrupole mass spectrometers10.1016/j.vacuum.2011.02.013 ↗
16. Nano-holes as standard leak elements10.1016/j.measurement.2014.09.017 ↗
17. Experimental and computational study of gas flow delivered by a rectangular microchannels leak10.1016/j.measurement.2015.06.011 ↗
18. Realization, characterization and measurements of standard leak artefacts10.1016/j.measurement.2014.10.045 ↗
19. New leak element using 1D nanofluidic channels and indium sealing10.1016/j.vacuum.2018.07.015 ↗
20. Comparison of some metrological characteristics of hot and cold cathode ionisation gauges10.1016/S0042-207X(02)00781-9 ↗
21. The uncertainties of calibration pressure at PTB10.1016/0042-207X(93)90097-T ↗
22. ISO 3567:2011 Vacuum Gauges - Calibration by Direct Comparison with a Reference Gauge.
23. Nonevaporable getter coating chambers for extreme high vacuum10.1116/1.5010154 ↗
24. ISO/IEC Guide 98-3, Uncertainty of Measurement – Part 3: Guide to the Expression of Uncertainty in Measurement (GUM: 1995).
25. Calibration apparatus of vacuum leak using constant conductance method
26. K. Jousten, Hanbook of Vacuum Technology, WILEY: 707-708.
27. Free molecule transmission probabilities10.1063/1.1702984 ↗
28. Erratum: free molecule transmission probabilities10.1063/1.1782171 ↗
29. Effusion of gases through conical orifices10.1021/j100796a006 ↗
30. Influence of orifice geometry on the simplification of transmission probability calculations10.1016/S0042-207X(98)00351-0 ↗
31. CODATA recommended values of the fundamental physical constants: 201410.1103/RevModPhys.88.035009 ↗
32. Development of a new UHV/XHV pressure standard (cold atom vacuum standard)10.1088/1681-7575/aa8a7b ↗
33. Review article: quantum-based vacuum metrology at the national Institute of standards and technology10.1116/1.5033568 ↗
34. Peksa · 2012
35. Dependence of the outgassing rate of “vacuum fired” 316LN stainless steel chamber on bake-out temperature10.1016/S0042-207X(98)00002-5 ↗
36. Karl Joustenb, K. Szyma!nski, the gas flux distribution in the XHV chamber of the vacuum primary standard developed by PTB10.1016/j.vacuum.2003.12.010 ↗
37. Monte Carlo study of a new PTB primary standard for very low pressures10.1088/0026-1394/36/6/15 ↗
38. Uncertainty analysis of the high vacuum part of the orifice-flow-type pressure standard10.1016/j.vacuum.2004.05.009 ↗
39. Report of pilot study CCM.P-P1 for international comparison of absolute pressure measurements in gas from 3 × 10-9 Pa to 9 × 10-4 Pa10.1088/0026-1394/52/1A/07012 ↗