Experimental verification of a Monte-Carlo simulation of the gas density in a vacuum chamber
Source abstract
Abstract
Abstract An investigation was performed to check the results of a Monte-Carlo simulation of the gas density variations inside a vacuum chamber used for calibration purposes. The simulations revealed variations of up to 0.33% between gas densities at different calibration gauge ports. In total seven spinning rotor gauges were used to detect the differences experimentally. The mean values of the differences that could be determined with an uncertainty of 0.1% of the mean gas density only are consistent with the results of the Monte-Carlo simulation and give confidence in the simulation overall.
Source: openaire · Quality: candidate unverified · Source record ↗
Research context
OpenAlex topics: Scientific Measurement and Uncertainty Evaluation · Advanced Sensor Technologies Research · Experimental and Theoretical Physics Studies
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (4)
Show source-provided references
1. New, fully automated, primary standard for generating vacuum pressures between 10-10Pa and 3×10-2Pa with respect to residual pressure10.1088/0026-1394/36/6/2 ↗
2. A new fully automated gas flowmeter at the PTB for flow rates between 10-13mol/s and 10-6mol/s10.1088/0026-1394/39/6/2 ↗
3. Monte carlo study of a new PTB primary standard for very low pressures10.1088/0026-1394/36/6/15 ↗
4. Cohen R, Clifford P, Giacomo P, Mathiesen O, Peters C, Taylor B, et al. Guide of the expression of uncertainty in measurement. International Organisation for Standardization; 1995. ISO publication ISBN-92-67-10118-9.