journal article·2001-08

Secondary electron detector with a micro-porous plate for environmental SEM

Witold Slówko

Vacuum· Volume 63· Issue 4
Publisher page

Abstract

Abstract Simple equipment for a standard scanning electron microscope (SEM) that could extend its operating range toward higher pressures in a sample chamber is presented. For a rotary pump of the speed 2.2 dm 3 /s and apertures φ 0.3 mm in diameter, the level 50 hPa was achieved, which is typical for the environmental SEM (ESEM) technique. To improve efficiency of secondary electron detection a micro-porous plate is proposed as a semi-permeable wall separating zones of high and intermediate pressures. Experiments made for the micro sphere plate (MSP) proved that it can work in low vacuum (up to 0.1 hPa) and demonstrates a relatively low gas conductivity (below 30 cm 3 /s for MSP 1.4 mm thick and 12 mm in diameter).
Source: openaire · Quality: candidate unverified · Source record ↗

Research context

OpenAlex topics: Photocathodes and Microchannel Plates · Particle Detector Development and Performance · Plasma Diagnostics and Applications

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (7)

Show source-provided references
1. Cameron · European Microsc Anal · 1994 · 29 · 9
2. Shah JS. British Patent 1477458, 1977.
3. Hollis KJ, Shah JS. EMAG97, Institute of Physics Conference Series No. 153. Bristol: IOP Publishing Ltd, 1997. p. 69.
4. Danilatos · 1988
5. Danilatos · J Microsc · 1990 · 160 · 910.1111/j.1365-2818.1990.tb03043.x ↗
6. Slówko W. Patent application. 1998. p. 329, 339.
7. Naaman · Rev Sci Instrum · 1996 · 67 · 333210.1063/1.1147416 ↗