journal article·2006-11

The Reference Gauge technique for static expansion ratios—Applied to NPL medium vacuum standard SEA3

John Greenwood

Vacuum· Volume 81· Issue 4
Publisher page

Abstract

Static expansion pressure standards are typically characterised in terms of the ratio between the density of an expanded sample of gas and that of an initial sample. For standards that perform a series of expansions through several system vessels most of the existing techniques to measure this expansion ratio involve disturbance to the system. They usually also involve operation with higher pressures and with valve sequences that are not encountered under normal operational conditions, thereby introducing the risk of systematic errors. A modification of the Reference Gauge technique for expansion ratios is described which has none of these disadvantages. The technique is described with reference to the medium vacuum standard SEA3 at the UK National Physical Laboratory.
Source: openaire · Quality: candidate unverified · Source record ↗

Research context

OpenAlex topics: Scientific Measurement and Uncertainty Evaluation · Radioactive Decay and Measurement Techniques · Advanced Sensor Technologies Research

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (18)

Show source-provided references
1. Jousten K. In: Lafferty JM, editor. Foundations of vacuum science and technology. New York: Wiley; 1998 (Chapter 12).
2. Messer · Metrologia · 1989 · 26 · 18310.1088/0026-1394/26/3/004 ↗
3. Schumann S. Transactions of the ninth national vacuum symposium of the American Vacuum Society. The Macmillan Company; 1962. p. 463–7.
4. Jitschin · Vacuum · 1990 · 40 · 29310.1016/0042-207X(90)90047-3 ↗
5. Elliott · Vacuum · 1967 · 17 · 43910.1016/0042-207X(67)91784-8 ↗
6. Berman · J Vac Sci Technol A · 1987 · 5 · 243610.1116/1.574865 ↗
7. Bergoglio · Vacuum · 1988 · 38 · 88710.1016/0042-207X(88)90486-1 ↗
8. Elliott KWT, Clapham PB. The accurate measurement of the volume ratios of vacuum vessels. NPL report MOM 28, January 1978.
9. Jousten · Vacuum · 1999 · 52 · 49110.1016/S0042-207X(98)00337-6 ↗
10. Redgrave · Vacuum · 1999 · 53 · 15910.1016/S0042-207X(98)00350-9 ↗
11. Sharma · J Vac Sci Technol A · 1988 · 6 · 314810.1116/1.575051 ↗
12. Greenwood · Vacuum · 2006 · 80 · 54810.1016/j.vacuum.2005.09.003 ↗
13. Miiller · Metrologia · 1999 · 36 · 61710.1088/0026-1394/36/6/26 ↗
14. Barker RM, Cox MG, Forbes AB and Harris PM. Software support for metrology best practice guide no. 4—discrete modelling and experimental data analysis. London: HMSO; April 2004, ISSN 1471-4124, version 2.
15. BIPM, IEC, IFCC, ISO, IUPAC, IUPAP, OIML. Guide to the expression of uncertainty in measurement, 1st ed. Geneva, Switzerland: International Organisation for Standardization; 1993. ISBN 92-67-10188-9.
16. Cox MG, Forbes AB, Harris PM, Smith IM. The classification and solution of regression problems for calibration. NPL report CMSC 24/03, August 2004.
17. Greenwood JC. NPL report DEPC-EM 005, London: HMSO; March 2006 ISSN 1744-0254.
18. Jitschin · Metrologia · 2002 · 39 · 24910.1088/0026-1394/39/3/2 ↗