journal article·2022-03Open access

Stability of bakeable capacitance diaphragm gauges

Julia Scherschligt, Daniel Barker, Stephen Eckel, James Fedchak, Emmanuel Newsome

Vacuum· Volume 197

Abstract

Here, we study the stability of capacitance diaphragm gauges. We focus on their stability before and after a controlled series of bakes. We find that baking results in appreciable shifts of the zero offset, but note that these can easily be corrected at time of use. Linearity however cannor be corrected at time of use, so it is essential to understand its effect if one wished to use similar gauges in a system that requires baking. For the gauges in this study, we find that baking introduces a minimal additional uncertainty, and that the total uncertainty can be kept to below about 0.3% at the k=2 confidence level (95%).
Source: openaire · Quality: candidate unverified · Source record ↗

Research context

OpenAlex topics: Advanced Electrical Measurement Techniques · Scientific Measurement and Uncertainty Evaluation

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (8)

Show source-provided references
1. Final report on the key comparison CCM.P-K4.2012 in absolute pressure from 1 Pa to 10 kPa10.1088/0026-1394/54/1A/07002 ↗
2. Measurement performance of high-accuracy low-pressure transducers10.1088/0026-1394/36/6/26 ↗
3. On the stability of capacitance-diaphragm gauges with ceramic membranes10.1116/1.3529023 ↗
4. Outgassing rate comparison of seven geometrically similar vacuum chambers of different materials and heat treatments10.1116/6.0000657 ↗
5. Vacuum furnace for degassing stainless-steel vacuum components
6. Investigations of medium-temperature heat treatments to achieve low outgassing rates in stainless steel ultrahigh vacuum chambers10.1116/1.4983211 ↗
7. Effect of heat treatments and coatings on the outgassing rate of stainless steel chambers10.1116/1.4853795 ↗
8. Development and characterisation of a low pressure transfer standard in the range 1 Pa to 10 kPa10.21014/acta_imeko.v7i1.496 ↗