journal article·2007-12

Inverted magnetron manometer with enhanced operating time

St. Wilfert, Chr. Edelmann

Vacuum· Volume 82· Issue 4
Publisher page

Abstract

Abstract If a cold-cathode ionization gauge is operated in a gas atmosphere with a high content of hydrocarbons at high pressures, the pressure reading of the gauge is disturbed by the contamination after a certain time of operation. Since in the most cases it is not possible to eliminate these disturbing residual gas components, one has to demand that the measuring process may not be substantially influenced by these gas components. To enhance the time of reliable pressure reading of a cold-cathode gauge we have equipped the entrance of an inverted magnetron gauge head with an ion baffle. By this measure the time of a reliable operation of the gauge is increased at least by a factor 3. The construction of this ion baffle is rather simple. It does not require any additional voltage supply or additional magnets. An additional feedthrough is only necessary for the measurement of the discharge current of the measuring system. The gauge has a range of pressure measurement from 10−9 to 3×10−3 mbar.
Source: openaire · Quality: candidate unverified · Source record ↗

Research context

OpenAlex topics: Plasma Diagnostics and Applications · Sensor Technology and Measurement Systems · Gas Sensing Nanomaterials and Sensors

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (20)

Show source-provided references
1. Wilfert · J Vac Soc Japan · 2003 · 46 · 3110.3131/jvsj.46.31 ↗
2. Klemperer OEH. Patent GB 555 134, 1943.
3. Conn · J Sci Instrum · 1954 · 31 · 43310.1088/0950-7671/31/11/119 ↗
4. O’Neal CD. Patent US 4.000.457, 1976.
5. Kageyama K, Utsunomija S. Patent US 3.872.377, 1975.10.1042/bst0030872a ↗
6. Grigorév AM. Proceedings of the fourth International Vacuum Congress 1968. p. 671.
7. Pakulin · Izmer Tekh · 1984 · 1 · 36
8. Tallon JAF, Gontero GM, Annency HS. Patent DE 164 8 689 A1, 1971.
9. Wutz M, Adam H, Walcher W. Theorie und Praxis der Vakuumtechnik 5th ed. Vieweg, Braunschweig: 1986. p. 447 (in German).
10. Haefer · Vak-Tech · 1962 · 11 · 41
11. Haefer · Vak-Tech · 1961 · 2 · 41
12. Haefer RA. Transactions of the eighth National Vacuum Symposium & second International Vacuum Congress 1961. p. 1346.
13. Haefer · Vak-Tech · 1961 · 4 · 95
14. Pátỳ L. Transactions of the third Czechoslovak Conference on Electronics and Vacuum Physics, Prague, 1965. p. 539–50.
15. Edelmann C, Wilfert S. Patent DE 102 43 634 A1, WO 2004/4029572 A1, 2004.10.1053/j.ajkd.2003.07.025 ↗
16. Wilfert · Vak Forsch Prax · 2006 · 18 · 710.1002/vipr.200600304 ↗
17. Kendall · J Vac Sci Technol · 1996 · A14 · 1296
18. Wilfert S, Schindler N. Technical report 2001, Otto von Guericke University Magdeburg, Germany (unpublished).
19. Wilfert · Appl Phys · 2004 · A78 · 66310.1007/s00339-003-2277-3 ↗
20. Schindler · Appl Phys · 2004 · A78 · 69110.1007/s00339-003-2283-5 ↗