Abstract

Abstract A variable reference leak rate system is developed for real time calibration of ultrasensitive leak detection. It is based on the molecular flow characteristics of the platinum wire-glass leak element combined with a helium pressure adjustable device. Helium reference leak rate range from 10−14 to 10−10 Pa m3/s is provided and has been used for real time calibration of ultrasensitive mass spectrometer leak detection system.