Offset scatter reduction of spinning rotor gauges by vibration isolation
Source abstract
Abstract
Abstract The spinning rotor gauge is an accurate vacuum gauge of good long-term stability. Many efforts have been aimed at the reduction of the uncertainties of this device. Here we examine the possibility of reducing the uncertainty in the low-pressure range caused by the scatter of the pressure-independent residual drag (offset). Random vibrations exposed to the system produce random eddy currents and lead to an additional scatter of the offset value. We tested different configurations, where the suspension head only was isolated from vibration or where both suspension head and stainless steel thimble were isolated. Data sets containing time, frequency of the rotor, relative deceleration rate and temperature were taken at residual pressure conditions. It was found that vibration isolation of the suspension head alone does not cause a significant reduction of the offset scatter, while vibration isolation of suspension head and thimble reduces the scatter by a factor of down to 1 30 . Surprisingly, it was found that the size of the reduction depends on the controller suspension head unit used for the measurement.
Source: openaire · Quality: candidate unverified · Source record ↗
Research context
OpenAlex topics: Scientific Measurement and Uncertainty Evaluation · Advanced Sensor Technologies Research · Non-Destructive Testing Techniques
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (9)
Show source-provided references
1. The spinning rotor gauge10.1116/1.573007 ↗
2. The effect of rotor acceleration on zero stability of spinning rotor gauges10.1016/0042-207X(88)90487-3 ↗
3. Stabilisation techniques for spinning rotor gauge residual drag10.1116/1.576443 ↗
4. An analysis of the effects of eddy currents on the accuracy of a spinning rotor gauge10.1016/S0042-207X(96)00297-7 ↗
5. Mechanik · 1973
6. MKS Instruments, Gasreibungsmanometer SRG-1, Teil 2, Schaltungen, Lizensiert durch Kernforschungsanlage Jülich GmbH, Deutschland.
7. Ponslet E. Isolation stacks preliminary design methodology. LIGO project, HYTEC-TN-LIGO-02. HYTEC Inc., Los Alamos, NM; 1996.
8. New fully automated primary standard for generating vacuum pressures between 10-10Pa and 3×10-2Pa with respect to residual pressure10.1088/0026-1394/36/6/2 ↗
9. Guide of the expression of uncertainty in measurement. International Organisation for Standardization; 1995.