Some considerations on handling the calibration results of capacitance membrane gauges
Source abstract
Abstract
Abstract Capacitance diaphragm gauges (CDG) are widely used also as transfer instruments; consequently, they need to be referred frequently to primary systems mostly in the (1000–0.1) Pa range where no other accurate gauges are available. They must be characterized individually since a general rule on their metrological behaviour is not possible. Various CDGs are calibrated at IMGC for the Italian Accreditation Service. The aim of the present paper is to demonstrate that, for each gauge and after the first calibrations, it is possible to arrive at an equation which represents the calibration coefficient as a function of pressure and is based on a reduced number of experimental points; as a consequence the time for a complete calibration procedure is cut down. One of the four gauges we examined, for which many calibration curves were available, was considered for a complete evaluation; then the evaluation model was applied to other gauges.
Source: openaire · Quality: candidate unverified · Source record ↗
Research context
OpenAlex topics: Sensor Technology and Measurement Systems · Advanced Electrical Measurement Techniques · Scientific Measurement and Uncertainty Evaluation
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (12)
Show source-provided references
1. Hyland · J Vac Sci Technol · 1991 · A9 · 284310.1116/1.577209 ↗
2. Bergoglio · Automazione e Strumentazione · 1997 · 129
3. Messer · Metrologia · 1989 · 26 · 18310.1088/0026-1394/26/3/004 ↗
4. Hirata · Metrologia · 1997 · 34 · 42110.1088/0026-1394/34/5/6 ↗
5. Poulter · J Vac Sci Technol · 1980 · 17 · 67910.1116/1.570541 ↗
6. Jousten · J Vac Sci Technol · 1997 · A15 · 239510.1116/1.580754 ↗
7. Sullivan · J Vac Sci Technol · 1985 · A3 · 172110.1116/1.573008 ↗
8. Jitschin · Vac Sci Technol · 1990 · A8 · 94810.1116/1.576901 ↗
9. Poulter · Vacuum · 1983 · 33 · 631110.1016/0042-207X(83)90098-2 ↗
10. Siu · J Vac Sci Technol · 1973 · A3 · 1721
11. Bergoglio M, Calcatelli A. Proceedings of the Imeko, vol. IXA, 1997. p. 140.
12. Guide to the expression of uncertainty in measurement. International Organization for Standardization, Genéve, 1993, ISBN 92-67-10188-9.