Method to test linearity of quadrupole mass spectrometers by use of a flowmeter and a standard leak
Source abstract
Abstract
Abstract A linearity test of the quadrupole mass spectrometer signal can be performed by using the pressure generated by the flow rate of a secondary leak as a fix point and varying the flow rate by a primary gas flowmeter around this value. Applying this method, we have investigated three different quadrupole mass spectrometer in a range of helium partial pressures between 10 −9 Pa and 10 −4 Pa, corresponding to flow rates of 10 −7 Pa L s −1 to 10 −2 Pa L s −1 in our system. Our preliminary results indicate significant non-linearities for even modest partial pressure changes.
Source: openaire · Quality: candidate unverified · Source record ↗
Research context
OpenAlex topics: Mass Spectrometry Techniques and Applications · Analytical Chemistry and Chromatography · Quantum, superfluid, helium dynamics
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (12)
Show source-provided references
1. Ehrlich · J Vac Sci Technol A · 1992 · 10 · 110.1116/1.578137 ↗
2. Hidalgo · Vacuum · 2008 · 82 · 115110.1016/j.vacuum.2008.01.050 ↗
3. ISO 14291:2012. Vacuum gauges – definitions and specifications for quadrupole mass spectrometers.
4. Lieszkovszky · J Vac Sci Technol A · 1990 · 8 · 383810.1116/1.576458 ↗
5. Bley · Vacuum · 1988 · 38 · 10310.1016/0042-207X(88)90606-9 ↗
6. Nakashima · J Vac Sci Technol A · 1995 · 13 · 247010.1116/1.579490 ↗
7. Jitschin · Vacuum · 1988 · 38 · 88310.1016/0042-207X(88)90485-X ↗
8. Solomon · J Vac Sci Technol A · 1986 · 4 · 32710.1116/1.573921 ↗
9. Bley · Vacuum · 1990 · 41 · 186310.1016/0042-207X(90)94115-7 ↗
10. Bergoglio · Flow Meas Instrum · 2006 · 17 · 12910.1016/j.flowmeasinst.2005.07.002 ↗
11. Jousten · Vacuum · 1993 · 44 · 13510.1016/0042-207X(93)90362-E ↗
12. Jousten · Metrologia · 2002 · 39 · 51910.1088/0026-1394/39/6/2 ↗