Abstract
Abstract A new two-stage flow-dividing system has been developed for the calibration of ultrahigh vacuum gauges from 10−9 Pa to 10−5 Pa for N2, Ar, and H2. This system is designed based on the techniques for our previously developed calibration system in the range from 10−7 Pa to 10−2 Pa. Three modifications were performed to extend the calibration pressure to a lower range. The relative standard uncertainty of the generated pressure (k = 1) is in the range from 2.3% to 2.6%, from 10−9 Pa to 10−5 Pa. The characteristics of ultrahigh vacuum gauges were also examined by using this system. The stabilities of the pressure reading, the linearity, the temperature dependence, and the long-term stability were examined. These results show that the calibration of ultrahigh vacuum gauges is possible in the range from 10−9 Pa to 10−5 Pa for N2, Ar, and H2 with the uncertainty of about 6.0% (k = 2) by this new system.
Source: openaire · Quality: candidate unverified · Source record ↗
Research context
OpenAlex topics: Scientific Measurement and Uncertainty Evaluation · Advanced Sensor Technologies Research · Advanced Frequency and Time Standards
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (24)
Show source-provided references
1. Yoshida · J Vac Sci Technol A · 2008 · 26 · 12810.1116/1.2823481 ↗
2. Yoshida · Vacuum · 2010 · 84 · 70510.1016/j.vacuum.2009.09.011 ↗
3. Yoshida · Vacuum · 2010 · 85 · 7810.1016/j.vacuum.2010.04.005 ↗
4. Bergoglio · Vacuum · 1988 · 38 · 88710.1016/0042-207X(88)90486-1 ↗
5. Tilford · J Vac Sci Technol A · 1988 · 6 · 285310.1116/1.575612 ↗
6. Jousten · Metrologia · 1999 · 36 · 49310.1088/0026-1394/36/6/2 ↗
7. Hong · J Vac Sci Technol A · 2006 · 24 · 183110.1116/1.2244534 ↗
8. Bergoglio · Vacuum · 2010 · 84 · 27010.1016/j.vacuum.2009.04.002 ↗
9. Li · J Vac Sci Technol A · 2010 · 28 · 109910.1116/1.3457934 ↗
10. Ichimura · Vacuum · 1999 · 53 · 29110.1016/S0042-207X(98)00373-X ↗
11. Akimichi · Metrologia · 2005 · 42 · S18410.1088/0026-1394/42/6/S08 ↗
12. Operation manual of IE 514, Oerlykon Leybolt vacuum GmbH.
13. Operation manual of Axtran, ULVAC, Inc.
14. Akimichi H, Komatsu E, Arai K, Hirata M. Proceedings of the 44th international conference on instrumentation, control and information technology (SICE 2005), Okayama, Japan, 8–10 August 2005, pp. 2145–2148, unpublished.
15. The BIPM key comparison database, calibration and measurement capabilities (Appendix C), mass and related quantities .
16. ISO 27894:2009 Vacuum technology – Vacuum gauges – Specifications for hot cathode ionization gauges.
17. Jousten · Vacuum · 1998 · 49 · 8110.1016/S0042-207X(97)00133-4 ↗
18. Hirata · Sinku · 1994 · 37 · 224
19. Poulter · Vacuum · 1981 · 31 · 14710.1016/0042-207X(81)90005-1 ↗
20. Wood · J Vac Sci Technol A · 1985 · 3 · 54210.1116/1.572990 ↗
21. Filippelli · J Vac Sci Technol A · 1995 · 13 · 258210.1116/1.579454 ↗
22. Li · J Vac Sci Technol A · 2003 · 21 · 93710.1116/1.1578654 ↗
23. Takahashi · J Vac Soc Jpn · 2008 · 51 · 37710.3131/jvsj2.51.377 ↗
24. The joint BIPM/ILAC working group, Working document CIPM 2008–11, Calibration and measurement capabilities.