journal article·2004-03

Partial-pressure measurement of atmospheric-pressure binary gas using two pressure gauges

Akira Kurokawa, Kenji Odaka, Shingo Ichimura

Vacuum· Volume 73· Issue 2
Publisher page

Abstract

Abstract We investigated a new method to measure the partial pressure of a binary gas system at near-atmospheric pressure conditions. The method utilizes two types of vacuum gauges, a capacitance manometer and a quartz friction pressure gauge. The partial pressure of the binary gas can be estimated by measuring the impedance change with a quartz friction pressure gauge, which depends on both viscosity and total pressure of a binary gas, and considering the total pressure change as measured by a capacitance manometer. We applied this method to measure the partial pressure of ozone in an ozone–oxygen gas mixture. The results suggested that the new method allows a partial-pressure measurement of ozone with an accuracy of Δ p =0.2 kPa , close to atmospheric pressure. The advantages of the new method include compact size, high safety standards when measuring highly reactive gases, and the allowance for in-line monitoring.
Source: openaire · Quality: candidate unverified · Source record ↗

Research context

OpenAlex topics: Analytical Chemistry and Sensors · Gas Sensing Nanomaterials and Sensors · Atomic and Subatomic Physics Research

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (2)

Show source-provided references
1. Nishiguchi · Appl. Phys. Lett. · 2002 · 81 · 219010.1063/1.1507829 ↗
2. Kokubun · J Vac Sci Technol · 1987 · A5 · 245010.1116/1.574869 ↗